7.10 Electronic Vaporizer (eVap) Troubleshooting
As a general rule, proceed with electronic vaporizer fault isolation as follows:
1. Review the system error logs for vaporizer specific entries.
• Refer to Section 7.9 for procedures related to the system error log entries.
2. Within the PC service application, perform the Vaporizer Test (Section 12.10.2).
• Follow procedures below as directed by the vaporizer test results.
3. After servicing the vaporizer, rerun the Vaporizer Test to confirm any repairs made
were successful.
7.10.1 Vaporizer
Test Results
The Vaporizer Test will display extended diagnostic information upon completion.
If the test passes, the message
PASSED: Electronic Vaporizer Subsystem Check
is displayed along with the data used during the check. This indicates that the vaporizer
is ready for use.
If the check encounters a problem, it will perform an automated fault isolation
procedure. With most single fault failures the automated procedure will identify the
failed component or a small set of possibilities.
In the event of a failure, a variety of messages are possible:
• Refer to the table in Section 7.10.2 for corrective actions.
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......