10 Illustrated Parts
M1046983 04/08
10-27
10.18.2 Suction Control Module
Item
Description
Stock Number
1
1a
Gauge, 760 mmHg
O-ring, Gauge (included with gauge assy, 2ea. required)
1009-3227-000
6700-0133-500
2
Control panel assembly, with suction regulator knob and mode control knob
1009-3274-000
3
3a
3b
Regulator Module (plugs into manifold assembly)
O-ring, Regulator Module, Large (included with regulator module)
O-ring, Regulator Module, Stem (included with regulator module)
6700-1225-800
6700-0136-500
0210-0527-300
4
Manifold Assembly, without Gauge and Regulator Module
1009-3277-000
5
Screw, #6 - 2 inch
1009-3340-000
6
Filter
0206-5159-300
7
Pilot valve adapter assembly (includes plunger, jam nut, and valve assembly)
1009-3278-000
8
Cap, white
1009-3385-000
* Lubricate the regulator module o-rings and the mating bore of the manifold sparingly with Dow 111 lubricant.
** Drop the plunger (
7a
), round end first, into the manifold. Thread the pilot valve into the manifold body. Set the
mode switch to raise the plunger. Adjust the pilot valve (
7b
) so that the plunger actuates the pilot valve
approximately half of its travel. Tighten the jam nut (
7c
).
6
only with Venturi
7**
*
7c
8
1
1a
3a*
7a
7b
3b*
3
2
4
5
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......