12 Service Application
M1046983 04/08
12-5
12.3.2 Gas Delivery Schematic
The Softkeys at the bottom of the Gas Delivery schematic bring up related diagnostic
screens that are also accessible from the Gas Delivery Subsystem menu.
•
Gas Supplies —
•
Mixer Output —
•
Mixer Press/Temp —
Notes
Settings
Remarks/Values
Flow/Concentration
Move slider to increase/decrease value. Use arrow keys to
fine tune setting.
O2 Concentration
(if O2 Concentration selected)
0%, 21%, 25% to 100%.
21% minimum for Air Balance Gas
0% minimum for N2O Balance Gas
Total Flow
(if O2 Concentration selected)
Off, 0.10 l/min to 0.50 l/min (in 10 ml/min increments)
0.50 l/min to 15.0 l/min (in 0.25 l/min increments)
* O2 Flow
(if Individual Flow selected)
0.00 l/min to 15.00 l/min
* Balance Gas Flow
(if Individual Flow selected)
Air = 0.00 l/min to 15.00 l/min
N2O = 0.00 l/min to 12.00 l/min
* Total Fresh Gas Flow limited to 15.00 l/min
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......