12 Service Application
M1046983 04/08
12-19
12.8 Gas Delivery Subsystem menu
The Gas Delivery Subsystem menu includes the following menu items:
• Gas Supply Status (
• Mixer Pressure and Temperature (
• Gas Delivery Status (
• Mixer Post/Checkout Test Results (
• Perform Mixer Tests (
12.8.1 Gas Delivery Subsystem
—
Gas Supply Status
Gas Supply Status
Label
Value Format
Range (kPa)
Range (psi)
Range (bar)
O2 Cylinder 1
XXXXX.XX
0.00 to 27580.00
0.00 to 40000.14
0.00 to 275.80
O2 Cylinder 2
XXXXX.XX
0.00 to 27580.00
0.00 to 40000.14
0.00 to 275.80
Air Cylinder
XXXXX.XX
0.00 to 27580.00
0.00 to 40000.14
0.00 to 275.80
N2O Cylinder
XXXX.XX
0.00 to 9805.00
0.00 to 1422.10
0.00 to 98.05
O2 Pipeline
XXX.XX
0.00 to 697.00
0.00 to 101.09
0.00 to 6.97
Air Pipeline
XXX.XX
0.00 to 697.00
0.00 to 101.09
0.00 to 6.97
N2O Pipeline
XXX.XX
0.00 to 697.00
0.00 to 101.09
0.00 to 6.97
O2 Select Valve
Open, Closed
(Open = connected to Mixer)
Air Select Valve
Open, Closed
(Open = connected to Mixer)
N2O Select Valve
Open, Closed
(Open = connected to Mixer)
Alt O2 Valve
Open, Closed
(Open = O2 bypass)
Alt O2 Button
Not Pressed, Pressed
O2 Flush
Not Pressed, Pressed
Gas Outlet Config.
SCGO or ACGO
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......