12 Service Application
M1046983 04/08
12-21
12.8.3 Gas Delivery Subsystem
—
Mixer Pressure and Temperature
Mixer Pressure and Temperature
Label
Value
Units
Range (kpa/C)
Range (psi/F)
Range (bar/C)
O2 Pressure
XX.XX
62.05 to 220.63
9.00 to 32.00
0.62 to 2.21
O2 Pressure Cal
X.XXX
Vdc
0.55 to 2.90
0.55 to 2.90
0.55 to 2.90
Balance Pressure
XX.XX
62.05 to 220.63
9.00 to 32.00
0.62 to 2.21
Balance Pres Cal
X.XXX
Vdc
0.55 to 2.90
0.55 to 2.90
0.55 to 2.90
Mixer Output Pres
XX.XX
62.05 to 199.95
9.00 to 29.00
0.62 to 2.00
Mixer Output Pres Cal
X.XXX
Vdc
0.55 to 2.90
0.55 to 2.90
0.55 to 2.90
O2 Temp
XX.X
5.00 to 50.00
41.00 to 122.00
5.00 to 50.00
O2 Temp Volts
X.XXX
Vdc
0.25 to 3.00
0.25 to 3.00
0.25 to 3.00
Balance Temp
XX.X
5.0 to 50.0
41.0 to 122.0
5.0 to 50.0
Balance Temp Volts
X.XXX
Vdc
0.25 to 3.00
0.25 to 3.00
0.25 to 3.00
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......