12.8.6 Gas Delivery Subsystem
—
Perform Mixer Tests
Selecting
Perform Mixer Tests
brings up the following screen. This screen includes and
automatic and a manual leak test of the Mixer’s balance gas inlet check valves.
Automatic
The
BalGas Check Valve Leak Test Auto
does not require disassembly of the system. It
is a sensitive test that will Pass check valve that have leak rates within specifications;
however, it may Fail some Mixers with marginal but acceptable leak rates.
• Mixers that Pass the auto leak test do not require further testing and can be left in
service.
• Mixers that Fail the auto leak test should be further tested using the manual leak test.
Manual
The
BalGas Check Valve Leak Test Manual
requires disassembly of the system.
• Mixers that Pass the manual leak test have acceptable leak rates and can be left in
service.
• Mixers that Fail the manual leak test should be replaced.
Automatic leak test
1. Connect an O
2
supply.
2. Select
Start Test
to perform the
BalGas Check Valve Leak Test Auto
.
• If Pass, balance gas check valve leak rate is acceptable.
• If Fail, verify leak rate using the manual leak test.
Summary of Contents for Datex-Ohmeda Aisys Carestation
Page 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Page 16: ...Notes xiv 04 08 M1046983 ...
Page 32: ...Notes 1 16 04 08 M1046983 ...
Page 158: ...Notes 5 20 04 08 M1046983 ...
Page 288: ...Notes 7 106 04 08 M1046983 ...
Page 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Page 356: ...Notes 9 60 04 08 M1046983 ...
Page 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Page 452: ...Notes 10 96 04 08 M1046983 ...
Page 466: ...11 14 04 08 M1046983 Notes ...
Page 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Page 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Page 497: ......