12 Service Application
M1046983 04/08
12-27
12.9.2 Vent Subsystem
—
Vent Flow and Pressure
Vent Flow and Pressure
Menu Item
Value
Units
Range
Counts (0-4095)
Inspiratory Flow
XXX.X
l/min
-120.0 to 120.0
XXXX
Expiratory Flow
XXX.X
l/min
-120.0 to 120.0
XXXX
<*> Airway Pressure
XX X
cmH
2
O
-20.0 to 120.0
XXXX
<*> Manifold Pressure
XXX
cmH
2
O
-20.0 to 120.0
XXXX
O2 Cell
XXX
%
5 to 110 %
XXXX
ADC Ref Voltage
X.XXX
Vdc
3.179 to 3.221
Flow Valve Setting
XXX.X
l/min
0.00 to 140.00
XXXX
Flow Valve Feedback
xxxx
mV
0 to 4095
XXXX
Flow Valve Current
xxxx
mA
0.0 to 102.4
XXXX
<*> kPA
-2.0 to 11.8
hPA
-19.6 to 117.7
mmHg
-14.7 to 88.3
mBar
-19.6 to 117.7