Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
8
Document no
.
M1125633-06
NOTE: Make sure that the tubes are not in contact with the sampling pump or the O
2
sensor, or
its springs.
NOTE: Check that tubes are not contaminated. Any contamination inside the tubing can
increase a risk of faulty operation in valves or sensors. If any contamination inside the tubing is
noticed then it is recommended to send the module to factory repair.
2.
Install the PM Kit:
−
Replace the special tubes (Nafion™).
−
Replace the Ref. filter assembly.
−
Replace the filters in the pneumatic unit (1 or 2 pcs).
−
Check the D-fend O-rings and replace them, if necessary.
−
Replace the D-fend and sampling line.
−
Clean or replace the fan filter.
NOTE: Use only approved sampling lines to ensure proper functioning.
Use a 2 m/7 ft. sampling line with Compact Airway Modules E-COVX and E-CAiOVX.
3.
The manufacturer recommendation is to replace the CO
2
absorber every 4 years. Replace
the absorber, if necessary.
"
4.
Connect the Compact Airway Module to a module slot in the module frame with a long
gas interface cable (the grounding plates of the cable should be removed for the
procedure).
2.1.4 Single-width Airway Module, E-miniC
1.
Detach the module box.
2.
Check that all cables and tubes are connected properly and that there are no loose
objects inside the module.
3.
Install the PM Kit:
−
Replace the special tube (Nafion™).
−
Replace the
every three years.
−
Check the D-fend body connector O-rings and replace them, if necessary.
−
Replace the
and the sampling line.
NOTE: Use only approved sampling lines to ensure proper functioning.
4.
Connect the module to a module slot with a long gas interface cable (the grounding
plates of the cable should be removed for the procedure).
"
2.1.5 Recorder module
1.
Clean the recorder.
−
Open the paper compartment hatch and remove the paper roll, if installed.
−
Remove any paper chaff from the paper compartment.
−
Clean the thermal printhead and the small glass window in front of the static brush
with a cotton swab dipped in isopropyl alcohol, if necessary.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...