Planned maintenance check list
19
Document no. M1125633-06
Invasive blood pressure measurement
2.
Check the function of the front panel membrane
Zero P4
key.
3.
Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
"
SvO
2
measurement
4.
Check that the SvO
2
values Meas. state, OM fail and OM temp. in the
COP
module service
menu all show NO OM.
5.
Checks with the SvO
2
simulator:
Turn the SvO
2
simulator pulsation switch to Medium and the range switch to Normal
pulse. Connect the simulator to the module and check that the following messages
appear in the digit field for SvO
2
:
Initializing, please wait > Warming up > Not calibrated
Check that ‘Meas. state’ has changed to
NORMAL
and ‘OM fail’ and ‘OM temp.’ show OK.
NOTE:
The ‘OM temp.’ may show
UNSTABLE
at first, but the message should change to OK
within a half a minute.
6.
Perform an In Vitro calibration with the SvO
2
simulator.
Check that the calibration date for In Vitro calibration was updated correctly and the SvO
2
reading on the screen is 81% (±2%).
"
Cardiac Output measurement
7.
Enter the
COP
module service menu.
Check the front panel
Start C.O.
membrane key.
8.
Enter the
C.O.
menu:
Others
- C.O.
Connect a catheter connecting cable to the module connector C.O.
If the module contains a C.O. Test connector, attach the catheter connector of the
connecting cable to the C.O. Test connector. Check that the message ‘Cable OK’ appears
in the menu after the self-test.
No Catheter > Self Test in Progress > Cable OK
9.
Check the C.O. measurement with a patient simulator.
"
2.2.15 NeuroMuscular Module, E-NMT
1.
Enter the
NMT
module service menu:
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check that the module’s memories have passed the internal
memory test, i.e. RAM, ROM and EEPROM all state OK.
2.
Check the function of the front panel membrane keys
Start-Up
and
Stop/Continue
.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...