Anesthesia and Critical Care Monitors
6
Document no
.
M1125639-03
3.
ComWheel cover
Check that the ComWheel cover is intact and attached properly.
"
4.
ON/STBY switch
Check that the ON/STBY switch is intact and attached properly. Check that the ON/STBY
switch changes its state firmly when turning it back and forth. Leave the switch into STBY
position.
"
5.
Stand-by LED
Reconnect and lock the command bar cable to the B-CPU6 board.
Check that the green stand-by LED is lit up (the Central Unit power cord is connected to
the mains).
"
3.2.3 Functional inspection
•
Turn the monitor on and enter the service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
- Service
(password 26-23-8)
6.
Keyboard version
Record the information regarding keyboard software.
"
7.
Alarm LEDs
Select the menu
KEYBOARD
with the ComWheel.
Highlight the text
UPPER LED
. Check that the red alarm LED is turning on and off when
pressing the ComWheel. Check also the yellow alarm LED by selecting
LOWER LED
from
the menu.
"
8.
Membrane keys
Press the keys one by one. Check that each key generates a sound from the loudspeaker
and the corresponding text in the menu changes from yellow to red.
"
9.
ComWheel
Turn the ComWheel clockwise and counterclockwise and check that each step generates
a sound and the corresponding values at the bottom of the menu increase.
Select
DUMMY PRESS
. Press the ComWheel and check that the press generates a sound
and the corresponding value in the menu increases.
"
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...