Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
60
Document no
.
M1137263-04
4.5 Interfacing printer
It is possible to interface a laser printer (either serial or parallel) to the S/5 Anesthesia Monitor
and S/5 Critical Care Monitor via the UPINET Board, B-UPI4NET. The printer must be PCL5 or
PCL6 compatible and it should contain at least 2 MB of memory.
4.5.1 Interconnection
Use a commercial serial printer interface cable or the UPI4 - Printer parallel interface cable
713701 or a standard parallel port printer cable for connecting the printer to the monitor.
1.
Make sure that the power on both units is turned off.
2.
Connect the cable to the corresponding connector on the B-UPI4NET.
3.
Connect the cable to the corresponding connector on the printer.
WARNING
Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
WARNING
Connecting the power supply cord of the printer to the wall power outlet
may cause the printer leakage current to exceed the limit specified for
medical equipment. A printer must be supplied from an additional
transformer providing at least basic isolation (isolating or separating
transformer).
4.5.2 Setting interfacing parameters on the printer
NOTE: For more information about printer settings, please refer to the documentation provided
with the printer.
4.5.3 Setting interfacing parameters on the S/5 Anesthesia Monitor or
S/5 Critical Care Monitor
Select the appropriate printer connection in the
Record/Print
menu.
4.6 Interfacing computer
It is possible to interface a computer to the S/5 Anesthesia Monitor and to the Critical Care
Monitor.
Contact your authorized GE Healthcare distributor for further advice on computer interface.
WARNING
Always make sure that the combination complies with the international
safety standard IEC 60601-1-1 for medical electrical systems and with the
requirements of local authorities.
WARNING
Connecting the power supply cord of the computer to the wall power outlet
may cause the computer leakage current to exceed the limit specified for
medical equipment. A computer must be supplied from an additional
transformer providing at least basic isolation (isolating or separating
transformer).
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...