Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
46
Document no
.
M1137266-02
Pneuma error
Module pneumatics error status. See
Table 6 Module pneumatics error status
for a detailed
description of the message.
Table 6
Module pneumatics error status
Bit
Error label
Description
What to do
0
pumping error
Stepper motor is not able to
generate negative pressure at the
pump pressure sensor during
zeroing.
Check the function of the stepper
motor.
1
eq. to amb. press failed Before filling the catheter, the
pressure in the module tubing is
equilibrated to the ambient pressure
by switching the pump valve on and
off. If ambient pressure is not
reached, this error is shown.
Calibrate the pressure sensors. Check
the pump valve function.
2
tubing/zero block occl Pressure at the pump pressure
sensor gets too low, when air is
aspirated though the zeroing valve.
Check tubing/valves for occlusion.
8
press sens values
differ
Pump pressure and catheter
pressure readings differ more than
50 mbars.
Calibrate the pressure sensors.
9
pneuma fatal error
This error is shown, if the pneumatic
error occurred is considered fatal.
Check for other pneumatics errors.
10
cath press too low
Catheter pressure reading is below
reasonable pressure values.
Make sure that you are not applying
too low pressure to the catheter
connector. Calibrate the pressure
sensors.
11
cath press too high
Catheter pressure reading is above
reasonable pressure values.
Make sure that you are not applying
too high pressure to the catheter
connector. Calibrate the pressure
sensors.
12
pump press too low
Pump pressure reading is below
reasonable pressure values.
Make sure that you are not applying
too low pressure to the catheter
connector. Calibrate the pressure
sensors.
13
pump press too high
Pump pressure reading is above
reasonable values.
Make sure that you are not applying
too high pressure to the catheter
connector. Calibrate the pressure
sensors.
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...