5-Module Frame, F-CU5, F-CU5P
23
Document no. M1125636-05
The settings with a Dynatech Nevada medSim 300 Patient Simulator:
SENSITIVITY switch position: 5 µV/V/mmHg
BP - 1 - WAVE - ART
ECG - BASE - BPM - 160
PACE - WAVE - NSR
28. ECG out
Connect an oscilloscope between the signal out connector pins 8 (ECG out) and 6
(Ground). Check that an analog signal which corresponds with the ECG waveform on the
screen comes out. The output signal’s ratio to the actual ECG signal should be around
1V/1mV.
"
29. Pressure out
Connect the oscilloscope between the signal out connector pins 7 (Pressure out) and 6
(Ground). Check that an analog signal which corresponds with the InvBP waveform on the
screen comes out. The output signal’s ratio to the actual InvBP signal should be around
1V/100mmHg.
"
30. Monitor software
Enter the service menu.
Monitor Setup
- Install/Service
(password 16-4-34)
- Service
(password 26-23-8)
Take down the information regarding monitor software.
"
31. Content of service log
Select SERVICE LOG from the menu. Record the Service log onto the E-REC by selecting
RECORD DATA. Check the content of recording for possible problems, then empty the
Service log by selecting RESET LOG from the menu.
"
32. Voltages
Check the power supply unit output voltages through the service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
- Service
(password 26-23-8)
-
Frame - Power Supply
The output voltages should meet the following ranges:
+15V
14.20….15.60 V
-15V
-14.00...-15.50 V
+15VD
14.10.….15.60 V
+5V
4.70...…5.40 V
If any of the voltages is out of the tolerance, replace the F-CPU DC Power Board.
"
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...