Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
24
Document no
.
M1125636-05
33. Watchdog circuitry
Test the Central Unit watchdog circuitry:
Monitor Setup - Install/Service
(password 16-4-34)
- Service
(password 26-23-8)
-
Set/Test
Go through the
Watchdog
,
WD by Overload
and
WD by UPI
tests one by one and check
that the monitor performs a restart in all other cases except when performing
WD by UPI
.
NOTE: When selecting WD BY OVERLOAD, restarting should take place approximately
after 15 seconds. With the other tests, restarting takes place within a couple of seconds.
If restarting did not take place, try to locate the fault:
Watchdog --> CPU board /Power supply unit
WD by Overload --> CPU board
"
34. Trend retaining
Check that the monitor is capable of storing the trend information and temporary
settings in a short (max. 15 minutes) standby.
Turn the monitor to standby and disconnect the power cord. Wait for two minutes, then
reconnect the power cord and turn the monitor back on. The monitor should perform a
“Warm start” which means the trend information and temporary settings should still be
available.
If the monitor performed a “Cold start” instead, the battery needs to be replaced.
The information regarding the start-up is saved also in the Service log.
"
35. Service reset button
Check the service reset button. Turn the monitor to standby and press the service reset
button for at least five seconds. Turn the monitor back on and check that the monitor
performs a “Cold start”.
"
36. Recovering from power loss
Disconnect the power cord (during operation) for a moment, reconnect it and check that
the monitor recovers without problems. The monitor should perform a “Warm start”.
NOTE: F-CU5(P) may give an audible alarm during the power loss. The alarm is set by the
DIP switch on the F-CPU DC Power Board.
"
37. Electrical safety check
Perform an electrical safety check and a leakage current test.
"
Summary of Contents for Datex-Ohmeda S/5
Page 4: ......
Page 6: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 2 Document no M1137263 04 ...
Page 12: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors vi Document no M1137263 04 ...
Page 28: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 16 Document no M1137263 04 ...
Page 58: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 46 Document no M1137263 04 ...
Page 62: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 50 Document no M1137263 04 ...
Page 86: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 74 Document no M1137263 04 ...
Page 100: ......
Page 102: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1125633 06 ...
Page 108: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1125633 06 ...
Page 126: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 24 Document no M1125633 06 ...
Page 132: ......
Page 136: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1137266 02 ...
Page 216: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 80 Document no M1137266 02 ...
Page 220: ......
Page 258: ......
Page 262: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors iv Document no M1125636 05 ...
Page 302: ......
Page 326: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1125637 04 ...
Page 330: ......
Page 332: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137269 02 ...
Page 348: ......
Page 350: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137272 02 ...
Page 384: ......
Page 386: ...Datex Ohmeda S 5TM Anesthesia and Critical Care Monitors ii Document no M1125639 03 ...
Page 400: ......
Page 402: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor ii Document no M1125641 03 ...
Page 416: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitor 14 Document no M1125641 03 ...
Page 420: ......
Page 422: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors ii Document no M1137274 01 ...
Page 428: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 6 Document no M1137274 01 ...
Page 431: ...Spare parts 9 Document no M1137274 01 3 2 Central Processor Unit for F CU5 F CPU 02 ...
Page 444: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 22 Document no M1137274 01 ...
Page 448: ...Datex Ohmeda S 5 Anesthesia and Critical Care Monitors 26 Document no M1137274 01 ...