Appendix C. Programming the XGS868i Using PanaView™
106
DigitalFlow™ XGS868i Programming Manual
C.5.5 Entering Setup Parameters
The signal limits and response times for the Model XGS868i are specified via the
SETUP
submenu. While following the
programming instructions, refer to Figure 53 on page 140 of Appendix D,
PanaView Menu Maps
. This submenu
includes four options:
•
Signal
- set the parameters related to the transducer signal
•
Default Setup
- initialize all parameters to default values
•
V Averaging
- specify the response of the meter to step changes
•
Advanced Features
- enable mass flow or activate K factors.
To enter the
Set Up
submenu, highlight and double-click on the
Set Up
option in the center pane of the
Channel
menu.
Remember to record all programmed data in Appendix B,
Data Records
.
C.5.5a The Signal Option
Use this option to set the limits for the incoming signal and other parameters affecting the transducer signal. For
example, the programmed signal strength low limit may be used to determine the trigger point for an alarm.
CAUTION!
The Signal default settings are suitable for most applications. Consult the factory
before changing any of these parameters.
1.
From the
Set Up
option menu, highlight and double-click on
Signal
in the center pane.
2.
For each parameter, click
[Next Item]
to accept the current value or enter a new value and click
[Next Item].
Table 20 on page 107 lists ranges and default parameters for each parameter.
Summary of Contents for DigitalFlow XGS868i
Page 2: ......
Page 4: ... no content intended for this page ...
Page 10: ...Contents viii DigitalFlow XGS868i Programming Manual ...
Page 52: ...Chapter 1 Programming Site Data 42 DigitalFlow XGS868i Programming Manual ...
Page 85: ...DigitalFlow XGS868i Programming Manual 75 Appendix A Menu Maps Appendix A Menu Maps ...
Page 86: ...76 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 96: ...86 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 146: ...Appendix C Programming the XGS868i Using PanaView 136 DigitalFlow XGS868i Programming Manual ...
Page 148: ...138 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 174: ...Warranty 164 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 175: ......