Option Card RTD Inputs
Option cards with RTD inputs have a temperature range of –148
o
to 660
o
F (–100 ° to 350 °C). Complete the following
steps to set up the RTD inputs of an option card installed in Slot 1 (refer to Figure 55 on page 142):
1.
Double-click on the desired
Input
(A, B, C or D).
Note:
The setup of RTD input A is used here as an example. Identical procedures would be used to set up the
remaining RTD inputs.
2.
Enter a
Label
of up to eight characters for input A and click
[Next Item].
3.
Double-click on
Off
to disable Input A and return to step 1, or on
Temp
to enable input A as a live temperature
input.
4.
Enter a temperature value for the
Zero
(low) end of the analog input range and click
[Next Item].
5.
Enter a temperature value for the
Full Scale
(high) end of the analog input range and click
[Next Item].
After responding to the above prompt, the meter returns to step 1. Click
[Exit Page]
to proceed to another menu or
option.
Summary of Contents for DigitalFlow XGS868i
Page 2: ......
Page 4: ... no content intended for this page ...
Page 10: ...Contents viii DigitalFlow XGS868i Programming Manual ...
Page 52: ...Chapter 1 Programming Site Data 42 DigitalFlow XGS868i Programming Manual ...
Page 85: ...DigitalFlow XGS868i Programming Manual 75 Appendix A Menu Maps Appendix A Menu Maps ...
Page 86: ...76 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 96: ...86 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 146: ...Appendix C Programming the XGS868i Using PanaView 136 DigitalFlow XGS868i Programming Manual ...
Page 148: ...138 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 174: ...Warranty 164 DigitalFlow XGS868i Programming Manual no content intended for this page ...
Page 175: ......