8
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
Revision A
2018554-001
1.5 SERVICE MODE
The
Service Mode
provides the same features as the Clinician Mode
and adds these functions:
Check the BP calibration
Re-calibrate BP
Adjust the pneumatic system overpressure point
Configure communications with a host computer
Change the language of operation
1.5.1 Accessing the Service Mode
To enter Service Mode from the Main menu, use the rotor and LCD.
1. Select the
More…
button.
2. Select the
Service
button.
3. The service menu requires an access code. Turn the rotor to
highlight a number, and push to make a selection. Enter access code
2 2 1 3
.