GE
HEALTHCARE
D
IRECTION
2307224-100, R
EVISION
6
D
ISCOVERY
LS S
YSTEM
S
ERVICE
M
ANUAL
Chapter 6 - Table
Page 115
3.) Refer to
. Remove the table top side covers. (Each side cover is held in place with
two quarter-turn captive screws, located on the under the table top.)
4.) Refer to
. Move the interlock jumper to the cover off position.
Figure 6-3:
Remove the Table Side Covers
5.) Remove the right table side covers from the elevator cover (scissor section) of the table.
a.) Unfasten the two screws and ground connection from two covers on the right side of the
table and one cover on the left side of the table.
Note:
Wait until
Section
, after
the pedal assembly is removed, to remove the cover from the left
side of the table.
Move jumper to "cover off" position.
Captive Quarter-Turn Screw