Page 155
6.2.6.21 Install Base CMD Adapter (P/N 2304737)
. The CMD adapter has three connectors with captive hardware. It is difficult to
attach the existing cable with the CMD Adapter fastened onto the ETC I/F Board. It is easier to
attach the existing cable to the back of the CMD Adaptor
before
attaching the adaptor to the ETC
I/F Board.
Figure 6-60:
CMD Connector, P/N 2304737
1.) Refer to
. Disconnect Signal Harness P/N 2271256, labeled
I/F J7
, from J7 on the
ETC I/F Board.
Figure 6-61:
ETC Board and ETC I/F
2. Connect this end, labeled
1. Connect existing
3. Connect new base CMD
Cable to J1.
Adapter Cable to J2.
ETC I/F J7
, to the ETC I/F Board.
ETC I/F J7