GE
HEALTHCARE
D
IRECTION
2307224-100, R
EVISION
6
D
ISCOVERY
LS S
YSTEM
S
ERVICE
M
ANUAL
Chapter 2 - Discovery LS System Safety Information
Page 63
2.7.1.1 Hazard Awareness Indicators - CPDU
. With the CPDU’s top cover hinged open, a small power lamp is visible. When
illuminated, this lamp indicates power is present within the CPDU.
Figure 2-23:
CPDU Power Lamp (CPDU Top Cover Opened and Latched)
2.7.1.2 Protected Service Outlets - CPDU
Refer to
. The service outlet is protected by a circuit breaker. The outlet is located on
the A4 panel.
2.7.1.3 Circuit Breakers and Switches - CPDU
Circuit Breakers
Refer to
. There are three (3) groups of circuit breakers in the CPDU used to protect
various parts of the system.
CB1 - Console AC Power
CB3 - Table and Gantry AC Service Outlets
CB4 - Table and Gantry AC to Stationary Electronics
CB5 - Gantry Rotating Power, Tilt Power and Communications
CB6 - Main Axial Drive Power
CB7 - Master 120/208 VAC Power (CB1, 3, 4 and 5)
Figure 2-24:
CPDU Circuit Breakers
Latch
CPDU Power
Lamp