Chapter 4 - Service Desktop, Tools and Diagnostics
Refer to
System Resets
function allows resetting and downloading the scanner
hardware as required to prepare the system for the scanning operation. Access
Systems Resets
as follows:
1.) Select
Systems Resets
from the Service Desktop. The reset applications selections include:
a.)
Scan
- Resets and downloads all controllers in the gantry and table.
b.)
Recon
- Resets the recon subsystem (IG, RIP, and SDC boards).
c.)
DAS/COLL
- Resets the DAS control and collimator control board.
d.)
Data Acquisition
- Resets the DAS, control, collimator control, and DIP boards.
2.) Select
Reset
on the displayed screen.
3.) Select
Run
.
The status box displays the status of the selected reset.
Figure 4-4:
Systems Reset
Click the appropriate button to reset an application.