GE
HEALTHCARE
D
IRECTION
2307224-100, R
EVISION
6
D
ISCOVERY
LS S
YSTEM
S
ERVICE
M
ANUAL
Chapter 6 - Table
Page 121
6.2.4.3 Install the Mechanical Stop Assembly
At least four versions of the Secondary Base currently exist in the field. Some versions contain both
a front and rear mechanical stop, some contain a single front
or
rear stop, and at least one version
has no mechanical stops at all.
1.) Refer to
. Determine the table type.
The mechanical stops are located on the right side of the secondary base, next to the linear
rail. The front stop is located about midpoint, and the rear stop is located near the clutch motor
assembly. If the base has:
• Front and Rear Stops, remove the Front stop.
• Front Stop, but no Rear Stop, move the Front stop to the rear location.
• No Stops, install a new Rear Stop from the kit.
Figure 6-12:
Front and Rear Mechanical Stops
2.) To remove the Front stop, use a 6 mm Allen wrench to remove the two mounting bolts, flats
and lock washers that fasten the bracket into place.
3.) If the table has no Rear stop, use the same M8 hardware to fasten the Front stop to the rear
location. (The table base should already have pre-drilled holes in the rear location.)
Front Mechanical Stop (Remove)
Rear Mechanical Stop (Install)