6.2.5.10 Remove the Motor Clutch Assembly (P/N 2309067)
For best results, check the table and secondary base with a level before removing the motor clutch
assembly. Check level again, before fastening the covers into place. If necessary, adjust the table
levelers to re-level the table and secondary base.
To disassemble the secondary base drive belt:
1.) Refer to
. Use two strips of masking tape, or some white correction fluid, to mark
the edges of the belt connector before disassembling the belt.
Figure 6-22:
Mark the Secondary Base Drive Belt
2.) Refer to
. Use a 4 mm Allen wrench to remove the four socket-head cap screws
from the fastener plate, and set aside.
3.) Remove the fastener plate and set aside.
4.) Remove the drive belt from the pulley on the motor clutch assembly.
Figure 6-23:
Disassemble Secondary Base Positioning Belt