3.) Use a 13 mm or 1/2 inch socket to reattach the new assembly to the secondary base.
a.) Fasten the motor clutch assembly to the right secondary base side rail with five bolts.
b.) Fasten the motor clutch assembly to the left secondary base side rail with four bolts.
4.) Slightly loosen the spreader hardware.
5.) Remove the C-clamps.
6.) Follow local guidelines to recycle or discard the spreader bar assembly.
6.2.6.11 Install the Table Bulkhead (If Necessary)
Early ME1 tables do not have a bulkhead, so the kit ships with a new table bulkhead. If the system
already has a bulkhead, do not replace it.
1.) If necessary, locate and remove the following hardware from the upgrade kit:
2.) Refer to
. Use a 5 mm Allen wrench to fasten the table bulkhead in place with two
M6 socket-head cap screws, flats and lock washers.
Note:
For best results, temporarily remove the bulkhead, attach the Cable P/N 2323724 to the back side
of the bulkhead, then fasten the bulkhead into place.
Figure 6-40:
Install Table Bulkhead Bracket, P/N 2309772
Part Number
Description
Quantity
2309772
Table Bulkhead Bracket
1
46-328417P13
M6 x 16 mm Socket-Head Cap Screw
2
2315373-4
M6 Lock Washer
2
2109878-3
M6 Flat Washer
2
Rear Side of Bulkhead