Chapter 6 - Table
6.2.7.8 Attach the Drip Trough to the Roller Canister Assembly
The drip trough diverts fluids away from the roller canister assembly, and onto the floor.
. Rest the canister assembly across the secondary base rails, so the tips
of the roller cover hang toward the floor.
2.) Orient the drip tray at the head end of the table so its sides angle toward the foot end of the
table when fastening the canister assembly into place.
3.) Use a 3 mm Allen wrench to fasten the drip tray to the roller cover canister with two sets of the
following M5 hardware:
• Two M5 x 10 mm socket-head cap screws (SHCS), P/N 46-312577P4
• Two M5 lock washers, P/N 46-328432P8
• Two M5 flat washers, P/N 2236649
Figure 6-121:
Attach Drip Trough, P/N 2333438
Tip forward and attach to
front end of secondary base.
This surface faces the gantry.
Sides angle toward table when