Chapter 6 - Table
Page 211
17.) Refer to
TABLE TYPE
Autobase
.
Figure 6-140:
Hardware Configuration Panel
18.) Click
ACCEPT
to close the panel with the current selections.
19.) Restart applications.
20.) Flash download:
a.) Click the
SERVICE
icon to open the Service Desktop.
b.) Click the
TOOLBOX/UTILITY
icon to open the tab.
c.) Click
INSTALL
to open the tab.
d.) Click
FLASHDOWNLOAD
to open the GUI.
e.) Click
UPDATE
.
NOTICE
FLASH download may require being run more than once. Continue to update FLASH
download until the
successful completion
message displays.
f.) Upon successful completion of the download, click
DISMISS
to close the GUI.
21.) Return to the Service Desktop and characterize the secondary base.
a.) Refer to
. Press the cradle release button on the gantry keypad to disengage
the secondary drive clutch motor.
b.) Push the table base to a location
between
the PET and CT detents.
c.) Click the
SERVICE
icon to open the Service Desktop.
d.) Select
CALIBRATION
.