Chapter 6 - Table
Page 213
Figure 6-142:
Adjust PET Flag
8. Return to
. Characterize the base for the second time.
9. Run another hardware reset.
10. Set up and run a CT-PET scan to test the new characterization values.
6.2.10 Manual Table Wiring
Refer to
and
for manual table wiring diagrams.
Figure 6-143:
2328900-2SCH
Loosen hardware with 2.5 mm Allen wrench.
Adjusting Screw