Chapter 7 - Gantry
Page 233
7.5.2.1 Replacing the External Retractor or its Cables
The 14 AEDI Cables, P/N 46-325061, are routed through the external retractor as illustrated in
Figure 7-12:
External Retractor Cables Cross Section
External Retractor Preparation
1.) Refer to
. Fold the retractor and disconnect the bracket to the legs
crossbeam using four M6 x 20 mm screws.
2.) Place the External Retractor, P/N HF000375200, on the floor and open it so that its clips face
toward to the floor.
Note:
Use a flat-blade screwdriver to unlock the retractor chains.
3.) If replacing the retractor, disconnect the Outer Retractor Bracket, P/N 2296530, and assemble
it on the new retractor. Secure the outer retractor bracket to the end of the external retractor
with four socket-head M6 x 10 screws.
4.) Position the retractor so that the Outer Retractor Bracket faces upward.
Top
Bottom
AEDI
AEDI
AEDI
AEDI
AEDI
AEDI
AEDI
AEDI
AEDI
AEDI
AEDI
AEDI
AEDI
AEDI