GE H
EALTHCARE
D
IRECTION
2307224-100, R
EVISION
6
D
ISCOVERY
LS S
YSTEM
S
ERVICE
M
ANUAL
Page 234
Chapter 7 - Gantry
. Replace and lay the cables inside the retractor. The length of each cable
that protrudes from the side to which the bracket is assembled is given in
.
Figure 7-13:
Placing Cable Inside Retractor
Table 7-1:
AEDI Cable Connections
Note:
Connect the cables so that their minimum bending radius is 12.5 mm.
6.) Tie-wrap the bundle of cables on the bracket side to prevent the cables from moving.
7.) Close the retractor by fastening the clips.
Cable
Part No.
(Old
Gantries)
Part No.
(New
Gantries)
Connection
Length of Cable
Protruding from
Retractor
AEDI
46-325061
2121185
J1
207 cm.
AEDI
46-325061
2121185
J2
207 cm
AEDI
46-325061
2121185
J3
207 cm
AEDI
46-325061
2121185
J4
207 cm
AEDI
46-325061
2121185
J5
207 cm
AEDI
46-325061
2121185
J6
207 cm
AEDI
46-325061
2121185
J7
207 cm
AEDI
46-325061
2121185
J17
69 cm
AEDI
46-325061
2121185
J18
69 cm
AEDI
46-325061
2121185
J19
69 cm
AEDI
46-325061
2121185
J20
69 cm
AEDI
46-325061
2121185
J21
69 cm
AEDI
46-325061
2121185
J22
69 cm
AEDI
46-325061
2121185
J23
69 cm
Connector
Length Protruding
Retractor
Cable
Bracket
J#
AEDI Cable
46-325061
Clips Open