Page 248
Chapter 8 - System QC, Calibration and Functional Tests
8.1.1.5 VQC Processing and Software Realignment
1.) Set the system into PET mode by clicking the [
PET
] icon on the Image Monitor desktop area.
The Scan Monitor displays the PET main screen.
2.) Click the [
Full QA
] button on the PET main screen.
3.) Select
Volumetric QC
from the drop-down menu. The Volumetric QC panel shown in
displays.
Figure 8-6:
Initial Volumetric QC Panel
. Click [
Load
] to load the image sets that represent the CT and PET planes.
The
Images Database Navigator
displays showing the Patient List.
Figure 8-7:
Images Database Navigator