12.4.1 Clutch Fails to Engage/Disengage
If the clutch fails to engage or disengage on an extended cradle table, perform the following checks:
Figure 12-10:
Troubleshooting for Clutch Fails to Engage/Disengage
Clutch fails to engage/
disengage
Is E-STOP circuitry or Cradle
Release botton activated ?
Reset the E-STOP
circuitry or press the
Cradle Release botton
yes
no
Does the SSR LED (which is found on the SSR in the
Upper Connectors Bracket) toggle on and off when you
press the Cradle Release button on the Control Panel
several times?
Perform a voltage measurement on
J221 between pins 1 & 4 (on the
Upper Connectors Bracket).The
voltage should change from 0Vdc to
130Vdc when you press the Cradle
Release button.
Check the cables in the path from the
ETC I/F J7 to the Upper Connectors
Bracket including the ETC Card
PNs:
1. CT Signal Harness J7&J9 to J18
yes
no
Check all the cables in the path
including the Clutch.
PNs:
1. 2323725
2. 2309059
3. 2323726
4. 2323727
Does the voltage
change between 0 to
130 Vdc?
The malfanction is in the SSR or
the Bridge or the connection
between the SSR and the J221
connector
check PNs:
1. 2326953
2. 2350128
no
yes
Measure the voltage on cable CAP J18
side, between pins 1 & 4. The voltage
should change from 0Vdc to 24Vdc when
you press the Cradle Release button.
Does the voltage
change between 0 to
24 Vdc?
no
The malfanction is in the
SSR or the connection
between the SSR and
the J18 Connector
yes
End of test