[EN] English - K0394 Issue 3
Calibration
21
5.
Repeat steps
3.
&
4.
for each digit and for the decimal point.
The value is ignored if it is not in the permitted limits (Table 11,
page 19).
This value is then used as the set point (SP) for point 1 on the
subsequent displays.
6.
This sequence of displays will follow:
The
SP
value is followed by the measured pressure -
CP
. This
sequence continues until the point 1 value is accepted or rejected.
7.
When the pressure is stable:
The value is ignored if it is not in the permitted limits (5% FS) or if the
CP value is not stable.
Point 2 (P2)
- Use the same steps (
1
to
5
above) to set
C2 - Point 2
.
This is the FS value and it is adjustable for the absolute and gauge
type DPI 104 (Table 11).
6.
This sequence of displays will follow:
The
SP
value is followed by the measured pressure -
CP
. This
sequence continues until the point 2 value is accepted or rejected.
7.
When the pressure is stable:
The value is ignored if it is not in the permitted limits (5% FS) or if the
CP value is not stable.
P1
Example sequence: Absolute type
• Press this button to accept the new
P1
value.
• The display shows the calibration point C2 - point 2 (
C2
).
• Press this button to reject the new
P1
value and move to the next
calibration option (
V2
).
P2
Example sequence: Absolute type
• Press this button to accept the new
P2
value.
• The display shows “donE”, and does a two-point calibration.
• The instrument will restart.
• Press this button to reject the new
P2
value and move to the next
calibration option (
V2
).