[EN] English - K0449
Prepare the instrument 2-15
Issue 1
Scaling:
2.11.2 Set the Automation options (source
)
Note:
Section 2.9 (Display operation)
shows how to set and
change the values on the display.
Nudge
Span Check
Set a scale for
Point 1 and
Point 2
Set On or Off
Set a label name
Previous: Back one display
Home: Back to start
1
2
3
1
TAP
Description
Process
: Nudge;
Values to set:
Step Size
Use these buttons to increase or decrease the output
value. Increments = Step size.
Use this button (Setpoint) to set the output value.
1
TAP
Description
Process
: Span Check;
Values to set:
Low, High, Dwell
Use these buttons to change the value manually from
High
to
Low
.
Use the start and stop buttons to change the value
automatically from
High
to
Low
.
Dwell
sets the period between each change. The cycle
repeats automatically.
Summary of Contents for Druck DPI 620
Page 8: ...Issue 1 viii About this manual K0449 EN English ...
Page 18: ...Issue 1 1 6 Instrument parts accessories and options K0449 EN English ...
Page 38: ...Issue 1 2 20 Prepare the instrument K0449 EN English ...
Page 52: ...Issue 1 3 14 Electrical and IDOS operations K0449 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0449 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0449 EN English ...
Page 68: ...Issue 1 7 4 Documenting functions K0449 EN English ...
Page 80: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 82: ...Issue 1 10 2 General specification K0449 EN English ...
Page 83: ...Customer service Visit our web site www gesensinginspection com ...
Page 84: ......