Issue 1
3-4 Electrical and IDOS operations
K0449 - [EN] English
1.
Set the applicable software options; see
Section 3.2.1
(Procedure overview)
.
2.
Complete the electrical connections and continue with the
measure operation.
3.2.5 Example procedure: Measure AC voltage (CH1) with the AC probe
WARNING
To prevent electrical shocks, use only the GE specified
AC probe (Part: IO620-AC) to measure AC voltages that
are more than 20 Vrms (maximum: 300 Vrms). Attach it
to the specified connections only.
This example shows Channel 1 (CH1) set up to measure an
AC voltage with the AC probe (maximum: 300 Vrms).
1.
Set the applicable software options for the AC Probe; see
Section 3.2.1 (Procedure overview)
.
2.
Complete the electrical connections.
Red
: V/Hz connector;
Black
: COM connector
Then continue with the measure operation.
2
Measure AC millivolts on
channel 1 (CH1)
Range: 0 to 2000 mV(rms)
1
B
Measure AC volts on
channel 1 (CH1)
Range: 0 to 20 V(rms)
A
1
Maximum: 0 to 20 Vrms
AC
2
1
AC
Maximum: 300 Vrms
Summary of Contents for Druck DPI 620
Page 8: ...Issue 1 viii About this manual K0449 EN English ...
Page 18: ...Issue 1 1 6 Instrument parts accessories and options K0449 EN English ...
Page 38: ...Issue 1 2 20 Prepare the instrument K0449 EN English ...
Page 52: ...Issue 1 3 14 Electrical and IDOS operations K0449 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0449 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0449 EN English ...
Page 68: ...Issue 1 7 4 Documenting functions K0449 EN English ...
Page 80: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 82: ...Issue 1 10 2 General specification K0449 EN English ...
Page 83: ...Customer service Visit our web site www gesensinginspection com ...
Page 84: ......