[EN] English - K0449
Electrical and IDOS operations 3-7
Issue 1
2.
Complete the electrical connections.
3.
If necessary change the
Source Settings
and continue with
source operation.
Values to set
:
Waveform (Square, Triangle, Sine);
Amplitude (Amplitude value)
4.
Automation
: Set the applicable output value; see
Section 2.11.2
.
3.2.9 Example procedure: Measure or simulate an RTD (or Resistance)
These examples (A and B) show Channel 1 (CH1) set up to
measure or simulate an RTD. A 4-Wire configuration gives the
best accuracy; a 2-Wire configuration has the lowest accuracy
(4-Wire RTD shown).
Note:
To measure or simulate resistance (
:
), set the
Resistance function
Example A
1.
Set the applicable software options; see
Section 3.2.1
(Procedure overview)
.
2.
Complete the electrical connections.
3.
If necessary change the
Settings
and continue with the
measure operation.
2
1
Measure an RTD on
channel 1 (CH1)
Range: 850 °C
RTD Type: PT100
Connection: 4 wire
A
3
Simulate an RTD on
channel 1 (CH1)
Range: 850 °C
RTD Type: PT100
Connection: 4 wire
Automation
:
Nudge
(
Section 2.11.2
)
2
B
4
1
3
Summary of Contents for Druck DPI 620
Page 8: ...Issue 1 viii About this manual K0449 EN English ...
Page 18: ...Issue 1 1 6 Instrument parts accessories and options K0449 EN English ...
Page 38: ...Issue 1 2 20 Prepare the instrument K0449 EN English ...
Page 52: ...Issue 1 3 14 Electrical and IDOS operations K0449 EN English ...
Page 60: ...Issue 1 5 2 Instrument communications K0449 EN English ...
Page 64: ...Issue 1 6 4 Datalog operation K0449 EN English ...
Page 68: ...Issue 1 7 4 Documenting functions K0449 EN English ...
Page 80: ...Issue 1 9 2 Maintenance procedures K0449 EN English ...
Page 82: ...Issue 1 10 2 General specification K0449 EN English ...
Page 83: ...Customer service Visit our web site www gesensinginspection com ...
Page 84: ......