DPI 740 User Manual
53
K0200 Issue No. 1
Command: CN - Get number of calibration
points required
Format:
CN?
Description:
Returns the minimum and maximum
calibration points required to carry out the
specified calibration procedure.
Example:
IU = 0
set-up input as mbar
PP = 123 PIN entered put instrument
into calibration mode
CT = 1
perform two point calibration
CN?
Reply: CN=1,2 requires between 1 and 2
calibration points
Query:
This is a query only command.
Command:
CA - Calibration Accept
Format:
CA
Description:
Accepts the calibrated data. If the correct
number of calibration points have been
entered with the CP command then the
calibration coefficients are calculated and
stored in the instrument's nonvolatile
memory. The instrument returns to
measurement mode.
Query:
There is no associated query command.
Command:
CX - Calibration abort
Format:
CX
Description:
Aborts the calibration procedure.
Example:
IU = 0
set-up input as mbar
PP= 123
PIN entered put instrument
into calibration mode
CT = 1
perform two point calibration
CP = 200
applied pressure 200 mbar
CX
abort calibration
Query: There is no associated query command.
Summary of Contents for Druck DPI 740
Page 1: ...GE Sensing Druck DPI 740 Precision Pressure Indicator User Manual K0200 ...
Page 2: ... The General Electric Company All rights reserved ...
Page 16: ...K0200 Issue No 1 8 DPI 740 UserManual Figure 2 2 Electrical connections ...
Page 32: ...K0200 Issue No 1 24 DPI 740 User Manual intentionally left blank ...
Page 46: ...K0200 Issue No 1 38 DPI740 UserManual ...
Page 72: ...K0200 Issue No 1 64 DPI 740 UserManual Approved Service Agents www gesensing com ...