[EN] English - K0457
Pressure calibrator operation (DPI 620) 5-9
Issue 1
* Reference
•
PV 621 models: see Chapter 2 (Pneumatic pressure
operation (PV 621))
•
PV 622 models: see Chapter 3 (Pneumatic pressure
operation (PV 622))
•
PV 623 models: see Chapter 4 (Hydraulic pressure
operation (PV 623))
4.
Complete the electrical connections. This example shows
Channel 2 (CH2) set up to measure current with 24V loop
power.
5.
To set the necessary pressure or vacuum for the calibration,
use the specified procedures for your PV 62x pressure
station*.
5.3.8 Example procedure: Switch test
1.
Assemble the pressure calibrator with the correct PM 620
module and PRV (if applicable); see Section 5.2.3.
2.
Set the applicable software options; see Section 5.3.1
(Procedure overview). This example shows one function:
•
Pressure (P1) is set to measure the pressure. The
Utility
is
set to
Switch Test
.
3.
To attach the device under test, use the specified
procedures for your PV 62x pressure station*.
2
1
3
5
4
6