Pressure calibrator operation (DPI 620) 5-7
Issue 1
5.3.6 Example procedure: Gauge/indicator calibration
1.
Assemble the pressure calibrator with the correct PM 620
module and PRV (if applicable); see Section 5.2.3.
2.
Set the applicable software options; see Section 5.3.1
(Procedure overview). This example shows one function:
•
Pressure (P1) is set to measure the pressure supplied by
the pressure station.
3.
To attach the device under test, use the specified
procedures for your PV 62x pressure station*.
2
1
3
4
Select Zero
Accept the adjustment value
TAP the window
(Maximise)
Select Settings
See the new reading and go back to the Home
window.
5
2
1
3
4