GE Medical Systems
EAGLE 1000
Page 13
Information Technologies
Servicing Instructions
227 468 01 Rev. G
1.3.2.2 Data transfer between Eagle and Gateway
Eagle and Gateway will communicate while transferring trend data. Transfer is initiated by
the user request function.
The data are first stored in a file and, when data transfer is complete, arranged in proper
order. The transport monitor waits for the results of this processing procedure.
If an error occurs in the database during this procedure, all the data in the Gateway are
deleted and the entire data must be transferred once more.
When the data transfer is initiated, the time will be sent by the transport monitor real-time
clock. From this the Gateway can identify any possible time misalignment of the internal
clock of the transport monitor and account for this when the trend data are entered.
Sporadic trend data, e.g., non-invasive blood pressure values, are only transferred for the
period of time in which continuous trend data, e.g., ECG, are also available. Earlier or later
parameter readings will be ignored.
1.3.2.3 Connection to the LAN
The QuickLink Station is connected to a Multiport or an SMX using twisted-pair wiring.
The Multiport can lead to the LAN using BNC or AUI wiring.
The SMX 008 leads to the LAN using AUI wiring, the SMX 009 to the LAN using BNC
wiring.
Summary of Contents for Eagle 1000
Page 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 130: ......
Page 135: ......
Page 136: ......
Page 138: ......
Page 141: ......
Page 142: ......
Page 143: ......
Page 144: ......
Page 145: ......
Page 146: ......
Page 147: ......