![background image](http://html.mh-extra.com/html/ge/eagle-1000/eagle-1000_servicing-instructions_128587065.webp)
GE Medical Systems
EAGLE 1000
Page 65
Information Technologies
Servicing Instructions
227 468 01 Rev. G
5
1
9
6
2.3.5 Socket on the rear panel of the Eagle 1000
RS 232 Connector
1
+12.5V ±3V <100mA short circuit-proof
2
RxD
3
TxD
4
(-12V V24)
5
GND
6
-
7
Connected to Pin 1
8
Nurse Call*
to 10V ±3.5V <50mA short circuit-proof
9
-
Nurse Call Connector
1
to 12.5V ±3V <100mA short circuit-proof
2
ECG Trigger (Delay max. 25ms,
Pulse +12.5V ±3V <15mA 1 ms)
3
Alarm (ALARM_open Collector
max. 20V, max. 20mA load)
4
-
5
-
6
TxD_20mA Current Loop*
7
GND
8
Nurse Call
to +10V ±3.5V <50mA short circuit-proof
9
Pause Alarm (KUE_open Collector)
* for V2.1 or higher