GE Medical Systems
EAGLE 1000
Page 108
Information Technologies
Servicing Instructions 227 468 01 Rev.G
5
1
9
6
4.7.5 QuickLink test: RS-232 connections
Female socket
Cable for testing internally QuickLink adapter leading to QuickLink Station
male
female
If necessary, use 9-pin male to 25-pin female adapter
Pin 1
Pin 1
Use the 4 connections indicated only
Pin 2------------- Pin 2
Pin 3------------- Pin 3
Pin 4
Pin 4
Pin 5------------- Pin 5
Pin 6
Pin 6
Pin 7------------- Pin 7
Pin 8
Pin 8
Pin 9
Pin 9
Cable for testing internally QuickLink adapter leading to Eagle 1000
male
female
If necessary, use 9-pin male to 25-pin female adapter
Pin 1
Pin 1
Use the 3 connections indicated only
Pin 2------------- Pin 3
(In zero modem Pin 2 and Pin 3 are swapped round)
Pin 3------------- Pin 2
Pin 4
Pin 4
Pin 5------------- Pin 5
Pin 6
Pin 6
Pin 7
Pin 7
Pin 8
Pin 8
Pin 9
Pin 9
Adapter 9-pin male > 25-pin female
Pin 2 > Pin 3, Pin 3 > Pin 2, Pin 5 > Pin 5