Operating instructions
3
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
21
3
Operating instructions
The following sections will guide you through each stage of the isoelectric
focusing process using the Ettan IPGphor 3.
3.1
Power up
1
Turn on the mains power switch (
|
) located on the rear panel.
This activates a self diagnostic program that runs for approximately 10
seconds. A series of screens indicate progress. If any component fails, the
diagnostic program will stop and a message will indicate the source of the
fault; note the fault and press the
START
key to skip to the next test. If any
failure is detected, call GE Healthcare service.
3.1.1
Ettan IPGphor 3 control panel
Once the diagnostic program is successfully completed, the main screen
indicates that the instrument is ready, as shown below.
Fig 3-1.
Ettan IPGphor 3 control panel.
/
L
Mains power
switch
WARNING!
In a situation where there is a risk of injury, turn off the instrument
by switching the mains power switch (located on the rear panel) to position
0
.
Malfunctions must be rectified before the Ettan IPGphor 3 is restarted.
Ettan IPGphor 3
P r o t # 1 F i l e 1
R e h y d r a t e 0 : 0 0 H r @ 2 0
˚
C
I E F @ 2 0
˚
C 5 0
µ
A / S t r i p
1 S t r i p s 0 S t e p s
Summary of Contents for Ettan IPGphor 3
Page 1: ...GE Healthcare Ettan IPGphor 3 Instrument User Manual Ettan...
Page 2: ......
Page 4: ......
Page 54: ...4 Care and maintenance 54 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 62: ...6 Customer service information 62 Ettan IPGphor 3 Instrument User Manual 11 0034 58 Edition AB...
Page 63: ......