Chapter 4 - General Procedures and Functional Checks
4-17
4-2-9
Archiving and Loading Presets
NOTE:
Always save presets before any software reload. This action is necessary to ensure that the presets
loaded after the software reload are as up–to–date as possible.
All user presets except changes to
Summary
,
Anatomy
, and
Biometry
pages, can be saved on an
MO disk for reloading on the system.
4-2-9-1
Archiving Presets to an MO Disk
NOTICE
Presets should not be saved on the same MO disk as images, because the Archive Menu lists the
images but not the presets stored on a MO disk.
Table 4-3 Saving Presets
STEP
TASK
FUNCTION
1.
Insert an empty, formatted MO disk
labelled “PRESETS” and marked with
Vivid
i
n/ Vivid q N’s serial number and
the date, into the MO Drive.
For more information about
formatting, see
.
2.
Press
Config
(or
F2
); select
Admin
then select
Backup
.
The Backup window is displayed.
3.
Enable the
User Defined
Configuration
check box.
4.
Select the MOD as the Destination.
5.
Press
Backup Now
.
A copy of the Presets is saved on the MO disk.
6.
Press <
Alt> + E
on the alphanumeric
keyboard.
The MO disk is ejected.