Chapter 7 - Diagnostics/Troubleshooting
7-99
7-5-1-4
Fans Speed Test
NOTE:
For a description of this test, refer to the information in
Fans Speed Test on page 7 - 14
1.) In the
Diagnostic
Test
window, trackball to
Fans Speed
in the list of
Monitoring
tests available
for
selection from the Diagnostic Test Tree (see
) and press
Set
. The Fans Speed
Test name is displayed under
Execute;
the relevant system components to be tested are
highlighted light blue in the Schematic View (AC/DC, Battery, PS), as shown below:
2.) Under the
Special Instructions
area, activate the
Loops
checkbox
and select the required number of
Loops
from the combo box.
3.) Adjacent to
On Fail,
select either
Stop
or
Continue
, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the
Start
button and press
Set
.
The Fans Speed Test commences. While the test proceeds, the name of the test currently in
progress (Fans Speed) is displayed in the
Execute
field and the current
Loop #
and specific test are
shown below. The messages
Init Done
and
Please
Wait
are displayed beside the progress bar in
the
Status
area.
As the testing sequence progresses, the progress bar will advance to reflect the test progress.
Status indicators in the Data Flow map corresponding to the various system components will be
highlighted in the appropriate color to indicate the current test status, as follows:
•
Green:
All recommended tests for these components have been completed and no problems
were identified.
•
Red:
Problems were identified in these components during the performance of the test.
NOTICE
IMPORTANT
In order to access an individual board test, it is necessary to insert the service dongle from
system boot-up and enter the Service Password.
Figure 7-68 Fans Speed Test