Section 7-5 - Monitoring Tests
7-5-1-5
Battery Parameters Test
NOTE:
For a description of this test, refer to the information in
Battery Parameters Test on page 7 - 14
.
1.) In the
Diagnostic
Test
window, trackball to
Battery Parameters
in the list of
Monitoring
tests
available
for selection from the Diagnostic Test Tree (see
) and press
Set
. The
Battery Parameters Test name is displayed under
Execute;
the relevant system components to be
tested are highlighted light blue in the Schematic View (AC/DC, Battery, PS), as shown below:
2.) Under the
Special Instructions
area, activate the
Loops
checkbox
and select the required number of
Loops
from the combo box.
3.) Adjacent to
On Fail,
select either
Stop
or
Continue
, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
4.) Trackball to the
Start
button and press
Set
.
The Battery Parameters Test commences. While the test proceeds, the name of the test currently
in progress (Battery Parameters) is displayed in the
Execute
field and the current
Loop #
and
specific test are shown below. The messages
Init Done
and
Please
Wait
are displayed beside the
progress bar in the
Status
area.
NOTICE
IMPORTANT
In order to access an individual board test, it is necessary to insert the service dongle from
system boot-up and enter the Service Password.
Figure 7-70 Battery Parameters Test