background image

GE H

EALTHCARE

D

IRECTION

 FQ091013, R

EVISION

 1 

 

V

IVID

 

I

 

N

 

AND

 V

IVID

 

Q

 N S

ERVICE

M

ANUAL

9-14

Section 9-10 - Optional Modo Cart

Section 9-10

Optional Modo Cart

Table 9-9

   Modo Cart - Replacement parts

ITEM

PART NAME

PART 

NUMBER

DESCRIPTION

QTY

FRU

1.

Modo Cart (without peripherals)

FL100100

BT’06 and BT’09
Modo Cart incl. isolation transformer, probe and gel holder, 
peripherals shelves, hand-rest and handles.

-

Y

2.

Cable Clip M-Cart

KTZ220885

BT’06 and BT’09
Cable Clip M-Cart

2

Y

3.

Isolation Transformer M-Cart

KTZ220714

BT’06 and BT’09
Isolation Transformer (Power Supply) IMED300WR M-Cart

1

Y

4.

Probe Holder Kit M-Cart

5183958

BT’06 and BT’09
Probe Holder Kit M-Cart

1

Y

5.

Front and Rear Wheels M-Cart

KTZ220887

BT’06 and BT’09
Front and Rear Wheels M-Cart

2

Y

6.

Fuses M-Cart

KTZ196333

BT’06 and BT’09
Fuses (10 pc. 4A/10 pc. 2A) for IMED300WR

-

Y

7.

Screw Kit

KTZ280045

Various screws

1

Y

Summary of Contents for FL000090

Page 1: ... VQN_SVC_FRNT_CVR FM P R E L I M I N A R Y Vivid i n and Vivid q N Service Manual Vivid q N with software version SW 11 x x BT11 Vivid i n with software version SW 9 x x BT 09 Vivid i n with software version SW 6 x x BT 06 Part Number FQ091013 Revision 1 ...

Page 2: ......

Page 3: ...MANUEL DANS UNE AUTRE LANGUE QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS ÉTÉ CONSULTÉ ET COMPRIS LE NON RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN L OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES MÉCANIQUES OU AUTRES DIESES KUNDENDIENST HANDBUCH EXISTIE...

Page 4: ... DO TÉCNICO OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS MECÂNICOS OU OUTROS ESTE MANUAL DE ASSISTÊNCIA ESTÁ DISPONÍVEL APENAS EM INGLÊS SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA QUE NÃO A GEHC SOLICITAR ESTES MANUAIS NOUTRO IDIOMA É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO NÃO TENTE EFECTUAR REPARAÇÕES NO EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO PREVIAMENTE E...

Page 5: ...TTEISTON KÄYTTÄJÄN TAI POTILAAN VAHINGOITTUMINEN SÄHKÖISKUN MEKAANISEN VIAN TAI MUUN VAARATILANTEEN VUOKSI ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ ΔΙΑΤΙΘΕΤΑΙ ΣΤΑ ΑΓΓΛΙΚΑ ΜΟΝΟ ΕΑΝ ΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ ΕΝΟΣ ΠΕΛΑΤΗ ΑΠΑΙΤΕΙ ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟ...

Page 6: ...RICKÁM PROUDEM PRACOVNÍKA POSKYTOVATELE SLUŽEB OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP PROUDU RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO JINÉMU RIZIKU DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE DENNE SERVICEMANUAL ER ...

Page 7: ...OPERATORIUS AR PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO MECHANINIŲ AR KITŲ PAVOJŲ DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT SERVICELEVERANDØREN OPERA...

Page 8: ...НИЙ МОЖЕТ ПРИВЕСТИ К ТОМУ ЧТО СПЕЦИАЛИСТ ПО ТЕХОБСЛУЖИВАНИЮ ОПЕРАТОР ИЛИ ПАЦИЕНТ ПОЛУЧАТ УДАР ЗЛЕКТРИЧЕСКИМ ТОКОМ МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ ДРУГОЕ ПОВРЕЖДЕНИЕ ТОВА СЕРВИЗНО РЪКОВОДСТВО Е НАЛИЧНО САМО НА АНГЛИЙСКИ ЕЗИК АКО ДОСТАВЧИКЪТ НА СЕРВИЗНИ УСЛУГИ НА КЛИЕНТ СЕ НУЖДАЕ ОТ ЕЗИК РАЗЛИЧЕН ОТ АНГЛИЙСКИ ЗАДЪЛЖЕНИЕ НА КЛИЕНТА Е ДА ПРЕДОСТАВИ ПРЕВОДАЧЕСКА УСЛУГА НЕ СЕ ОПИТВАЙТЕ ДА ИЗВЪРШВАТЕ СЕРВИЗНО ОБ...

Page 9: ...RA OPERATERA ILI PACIJENTA PROUZROČENE STRUJNIM UDAROM MEHANIČKIM I DRUGIM OPASNOSTIMA TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZÍCII LEN V ANGLIČTINE AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO ANGLIČTINU POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE ZODPOVEDNOSŤOU ZÁKAZNÍKA NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR AKO SI NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ ZANEDBANIE TOHTO U...

Page 10: ...ĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE ETMEYİNİZ BU UYARININ GÖZ ARDI EDİLMESİ ELEKTRİK ÇARPMASI YA DA MEKANİK VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN OPERATÖRÜN YA DA HASTANIN YARALANMASINA YOL AÇABİLİR DİKKAT TR JA Traditional Chinese ...

Page 11: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL ix 2 NOV 2011 FRONTMATTER FM P R E L I M I N A R Y ZH CN KO ...

Page 12: ...pecially trained field engineers All of GE s electrical work on these products will comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qualified personnel i e GE s field engineers personnel of third party service companies with equivalent training or licensed electricians to perform electrical servicing on the equipment OMISSIONS ERRORS...

Page 13: ...n may not be copied or duplicated in any form in whole or in part without prior written permission of GE Healthcare GE Healthcare may revise this publication from time to time without written notice TRADEMARKS All products and their name brands are trademarks of their respective holders COPYRIGHTS All Material Copyright 2011 by General Electric Company Inc All Rights Reserved ...

Page 14: ...n History List of Effected Pages LOEP REVISION DATE REASON FOR CHANGE 1 2011 NOV 02 Initial release Pages Revision Pages Revision Pages Revision Title Page 1 1 1 to 1 26 1 7 1 to 7 176 1 Warnings i to xii 1 2 1 to 2 12 1 8 1 to 8 154 1 TOC 1 3 1 to 3 180 1 9 1 to 9 20 1 4 1 to 4 40 1 10 1 to 10 36 1 5 1 to 5 48 1 Back Cover N A 6 1 to 6 16 1 ...

Page 15: ...rtant Conventions 1 5 Conventions Used in this Manual 1 5 Safety Considerations 1 7 Introduction 1 7 Human Safety 1 7 Mechanical Safety 1 7 Electrical Safety 1 9 Dangerous Procedure Warnings 1 11 Product Labels and Icons 1 12 Universal Product Labels 1 12 Label Descriptions 1 14 Vivid i n Vivid q N Battery Safety 1 17 Vivid i n and Vivid q N External Labels 1 19 SafeLock Cart Labels 1 20 EMC EMI a...

Page 16: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xiv Contact Information 1 25 System manufacturer 1 26 ...

Page 17: ... Requirements 2 2 Cooling Requirements 2 2 Lighting Requirements 2 2 Time and Manpower Requirements 2 2 Electrical Requirements 2 3 EMI Limitations 2 4 Probe Environmental Requirements 2 5 Facility Needs 2 6 Purchaser Responsibilities 2 6 Mandatory Site Requirements 2 7 Site Recommendations 2 7 Networking Pre Installation Requirements 2 9 Connectivity Installation Worksheet 2 10 Check List 2 12 ...

Page 18: ...tation box Contents 3 12 Physical Inspection 3 12 EMI Protection 3 12 Preparing for Installation 3 13 Confirming Customer Order 3 13 Verifying the Transportation Box Contents 3 13 Component Inspection 3 14 Confirming SafeLock Cart Voltage Configuration 3 19 System Voltage Confirmation 3 19 Ensuring Protection from EMI 3 20 Completing the Hardware Installation 3 21 Connecting Printers 3 21 Connecti...

Page 19: ...ity Configuration 3 98 Set Up Connection to a DICOM Server in a Network 3 110 Setup Connection to HL7 Server in a Network 3 117 Query Retrieve Q R Setup 3 120 MPEGVue Export Configuration 3 127 eVue Dataflow Configuration 3 132 Using MPEGVue eVue on a Remote PC 3 136 Storing and Transporting the Unit 3 177 Safety Precautions for Moving the Vivid i n Vivid q N Unit 3 177 Transportation Box and Pack...

Page 20: ...ew 4 3 Power ON Boot up 4 3 Power Shut Down 4 7 Log On to the System as ADM 4 10 Using Removable Media 4 12 Labeling Removable Media 4 15 Formatting Removable Media 4 15 Verifying Removable Media 4 16 Archiving and Loading Presets 4 17 Functional Checks 4 19 Basic Controls 4 19 Peripherals 4 28 SafeLock Cart Functional Tests 4 33 Back End Processor Tests 4 34 System Turnover Checklist 4 37 Softwar...

Page 21: ...eiver 5 15 Probe and MUX P M Board 5 18 Back End Processor 5 21 Introduction 5 21 ETX SBC Central Processing Unit CPU 5 23 ETX Base Board 5 23 Hard Disk 5 25 Fan Control 5 25 RTC BIOS Battery 5 25 Back End Processor Location of Components 5 26 Left Right Speakers 5 28 External Input Output I O 5 29 Power Supply System 5 30 Electrical Power 5 30 Introduction 5 30 Power Supply Unit Components 5 31 D...

Page 22: ... Q N SERVICE MANUAL xx ECG Patient I O Module Specifications 5 40 Isolation 5 40 ECG Respiratory Module 5 41 Internal ECG Respiratory Board 5 41 External ECG Respiratory Interface 5 43 Peripherals 5 46 Introduction 5 46 Cooling System 5 48 General Information 5 48 ...

Page 23: ...Calibration 6 2 Contrast Adjustment Procedure 6 2 Brightness Adjustment Procedure 6 3 Testing Your Contrast and Brightness Settings 6 3 Blue Tint Adjustment Procedure 6 3 Blue Tint 2 Adjustment Procedure 6 5 External Monitor Calibration 6 8 Keyboard Adjustments 6 10 Configuring Print Orientation 6 10 Video System Adjustments 6 10 Beamformer Calibration 6 10 Battery Calibration 6 11 ...

Page 24: ...sing the Front End Diagnostic Options 7 25 Calibration Tests 7 25 Radio Frequency Interface RFI Diagnostic Tests 7 31 TR 32 Diagnostic Tests 7 46 Probe and MUX Diagnostic Tests 7 61 Monitoring Tests 7 86 Accessing the Monitoring Diagnostic Test Options 7 86 Performing Back End Diagnostics on the System 7 108 Accessing the Back End Diagnostic Test Options 7 108 InSite ExC 7 137 Service Desktop 7 13...

Page 25: ... 8 38 BIOS Battery Replacement 8 45 HVPS Replacement Procedure 8 46 Fan Replacement Procedure 8 49 Speaker Assembly Replacement Procedure 8 52 LCD Display Replacement Procedure 8 53 LCD Flex Cable Replacement Procedure 8 58 SafeLock Cart Components Replacement 8 61 Overview of SafeLock Cart Replaceable Components 8 62 SafeLock Cart Replacement Procedures Quick Reference List 8 63 Upper Cover Rear ...

Page 26: ...Setting the BIOS 8 108 BIOS Firmware Update 8 118 Formatting and Partitioning the Hard Disk Automatic Procedure 8 121 Formatting and Partitioning the Hard Disk Manual Procedure 8 123 Full Re ghost and Software Installation 8 128 Installing Software only from DVD 8 130 Installing Software only or Patch 8 134 Installation of PMC Version 8 135 Software Roll back Procedure 8 139 Performing a Complete ...

Page 27: ...se of Chapter 9 9 1 List of Abbreviations 9 2 Mechanical Hardware Parts 9 3 Electronic Parts 9 4 Cables 9 7 Software 9 8 Probes 9 9 Peripherals 9 11 SafeLock Cart Parts 9 12 Optional Modo Cart 9 14 Vivid i n Spare Part Kits 9 15 Product Manuals 9 19 Product Manuals BT 11 9 19 Product Manuals BT 09 9 19 Product Manuals BT 06 9 20 ...

Page 28: ...tional Checks 10 10 Input Power Checks 10 11 Cleaning 10 12 Physical Inspection 10 15 Optional Diagnostic Checks 10 16 Probe Maintenance 10 17 Probe Checks 10 17 Probe Handling 10 17 Basic Probe Care 10 18 Probe Cleaning 10 18 Returning and Shipping of Defective Probes 10 20 Electrical Safety Tests 10 21 Safety Test Overview 10 21 GEMS Current Leakage Limits 10 22 Outlet Test Wiring Arrangement US...

Page 29: ...ARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxvii Excessive Current Leakage 10 33 Possible Causes of Excessive Current Leakage 10 33 Vivid i n Vivid q N Inspection Certificates 10 34 ...

Page 30: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxviii ...

Page 31: ... q N portable ultrasound scanner The service provider must read and understand all the information presented here before installing or servicing a unit Table 1 1 Contents in Chapter 1 Section Description Page Number 1 1 Overview 1 1 1 2 Important Conventions 1 5 1 3 Safety Considerations 1 7 1 4 Product Labels and Icons 1 12 1 5 EMC EMI and ESD 1 21 1 6 Lockout Tagout LOTO Requirements 1 23 1 7 Re...

Page 32: ...ock diagrams and functional explanations of the electronic circuits Chapter 6 Service Adjustments Contains instructions for performing service adjustments to the Vivid i n and Vivid q N portable ultrasound scanner Chapter 7 Diagnostics Troubleshooting Provides instructions for setting up and running diagnostic troubleshooting and other related routines for the Vivid i n and Vivid q N portable ultr...

Page 33: ... Covered in this Manual The Vivid q N models documented in this manual is shown in Table 1 3 below Table 1 2 Vivid i n Models Model H Cat No Part Number Description Comments Vivid i n Portable Ultrasound Scanner H45561AC FL000110 Laptop style system BT 09 H45551DE FL000090 Laptop style system BT 06 Table 1 3 Vivid q N Model Model H Cat No Part Number Description Comments Vivid q N Portable Ultraso...

Page 34: ... The Vivid q N ultrasound imaging scanner is similar to the Vivid i however newly designed to offer additional features including the ability to support use of M4S RS probes Also weighing only 5 Kgs 11 lbs this laptop style system is extremely versatile and depending upon the installed software can be used for a variety of applications Note Upgrade options may be available with future versions 1 1...

Page 35: ...ls of safety precautions are found on the equipment and throughout this service manual Different levels of severity are identified by one of the following icons which precede precautionary statements in the text DANGER Indicates the presence of a hazard that will cause severe personal injury or death if the instructions are ignored WARNING Indicates the presence of a hazard that can cause severe p...

Page 36: ...s chapter In addition to text several different graphical icons symbols may be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific procedures that should be followed Table 1 4 Standard Hazard Icons ELECTRICAL MECHANICAL RADIATION LASER HEAT PINCH Table 1 5 Standard Icons Indicating a Special Procedure Be Used AVOID STATIC ELECTRIC...

Page 37: ...T IS TO BE MOVED ALONG ANY INCLINE USE EXTREME CAUTION MAKE SURE THAT THE VIVID I n or Vivid q N SCANNER AND ALL PERIPHERALS ARE SECURELY MOUNTED ON THE CART BEFORE ATTEMPTING TO MOVE IT DANGER ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN USE DO NOT USE A DAMAGED OR DEFECTIVE PR...

Page 38: ...i n or Vivid q N system and SafeLock cart if applicable in a vehicle ALWAYS Eject any disks from the MOD if installed Ensure that the Vivid i n or Vivid q N system is well prepared and packed in its original packaging before transporting Special care must be taken to correctly position the packing material For further information refer to Chapter 3 System Setup Place the probes in their carrying c...

Page 39: ...be cables to become entangled or to be caught in the SafeLock Cart wheels NOTE For detailed information on handling endocavity probes and invasive intracardiac probes refer to the appropriate supplementary instructions for each probe In addition refer to the Vivid i n Vivid q N User Manual for detailed probe handling instructions 1 3 4 2 Peripherals 1 3 4 2 1 Safety and Environmental Guidelines WA...

Page 40: ...is carried out It is very difficult to attach unique dimensions to the PATIENT ENVIRONMENT In practice a distance of 2 5 m 8 2 ft above the floor on which the medical personnel stand and a horizontal distance of 1 5 m 4 9 ft have justified themselves as indicative of the dimensions of the Patient Environment The patient environment vicinity is depicted as a dashed line in this procedure see the ex...

Page 41: ...N THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING EXPLOSION WARNING DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A DEFINITE SAFETY HAZARD EQUIPMENT IS NOT SUITABLE FOR USE IN THE PRESENCE OF A FLAMMABLE ANAESTHETIC MIXTURE WITH AIR OR WITH OXYGEN OR NITROUS OXIDE DO NOT SUBSTITUTE PARTS OR MO...

Page 42: ...icons These represent pertinent information regarding the operation of the unit 1 4 1 Universal Product Labels The following diagrams illustrate the labels found on the Vivid i n and Vivid q N ultrasound units For an explanation of label icons and symbols refer to Table 1 8 on page 1 14 Table 1 6 Main Label for Vivid q N DESCRIPTION ILLUSTRATION First version of Main Label Table 1 7 Main Label for...

Page 43: ...ALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 1 Introduction 1 13 First version of Main Label Table 1 7 Main Label for Vivid i n cont d 2 of 2 DESCRIPTION ILLUSTRATION ...

Page 44: ...N Consult accompanying documents is intended to alert the user to refer to the User Manual or other instructions when complete information cannot be provided on the label Various CAUTION Dangerous voltage the lightning flash with arrowhead in equilateral triangle is used to indicate electric shock hazards Various Device Listing Certification Labels Laboratory logos or labels that denote conformity...

Page 45: ...injury that may be caused by the weight of the machine if one person attempts to move it considerable distances or on an incline Used in the Service and User Manual which should be adjacent to equipment at all times for quick reference DANGER Risk of explosion used in The system is not designed for use with flammable anesthetic gases See EXPLOSION WARNING on page 1 11 Protective Earth Indicates th...

Page 46: ...ISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 1 16 Section 1 4 Product Labels and Icons Date of manufacture On the base of the unit Table 1 8 Label Icons and Symbols Description and Location cont d 3 of 3 Label Name Description Location ...

Page 47: ...o not heat the battery or discard it in a fire Do not expose the battery to temperatures over 60o C 140o F Keep the battery away from fire and other heat sources Do not charge the battery near a heat source such as a fire or heater Do not leave the battery in direct sunlight Do not pierce the battery with a sharp object hit it or step on it Do not use a damaged battery Do not apply solder to a bat...

Page 48: ...ize Li lon smart packs Important For Vivid i n and Vivid q N portable ultrasound scanners supplied from the factory with a dummy battery when unpacking the unit do not discard the dummy battery When charging the Vivid i n or Vivid q N lithium ion battery on an external charger since the battery in fact forms one of the four legs of the scanner removing it will leave the unit standing unbalanced Fo...

Page 49: ... labels described in the previous section additional labels may be found on the Vivid i n and Vivid q N ultrasound unit as described in the following section Rating Labels section see below 1 4 4 1 Rating Labels Indicates the ultrasound unit s basic power compliance and regulatory compliance information For illustrations see Table 1 6 on page 1 12 and Table 1 7 on page 1 12 ...

Page 50: ...VID I N AND VIVID Q N SERVICE MANUAL 1 20 Section 1 4 Product Labels and Icons 1 4 5 SafeLock Cart Labels The labels shown in the example below are located on the rear side of the cart Figure 1 2 Vivid i n and Vivid q N SafeLock Cart Labels Example ...

Page 51: ...bances in the electrical power supply 1 5 2 CE Compliance The Vivid i n and Vivid q N portable ultrasound scanner conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge radiated and conducted RF fields magnetic fields and power line transient requirements For applicable standards refer to the Safety Chapter in the Vivid i n and Vivid q N User...

Page 52: ...s Used Scope Standard Directive Product safety requirements for Electrical Medical devices Class I Type B with BF and CF applied parts ICE catheter and ECG leads IEC 60601 1 EN 60601 1 UL 60601 1 C 22 2 601 1 M90 Product safety requirements for Electrical Medical devices safety requirements for medical electrical systems IEC 60601 1 1 EN 60601 1 1 Product safety requirements for Electrical Medical...

Page 53: ...agout Devices 5 Control all stored and residual energy 6 Verify isolation All potentially hazardous stored or residual energy is relieved NOTICE Energy Control and Power Lockout for Vivid i n and Vivid q N When servicing parts of the system where there is exposure to voltage greater than 30 Volts 1 Turn off the breaker 2 Unplug the system 3 Maintain control of the system power plug 4 Wait for at l...

Page 54: ...uids must be properly removed from any part or equipment prior to shipment GE Healthcare employees as well as customers are responsible for ensuring that parts equipment have been properly decontaminated prior to shipment Under no circumstance should a part or equipment with visible body fluids be taken or shipped from a clinic or site for example body coils or an ultrasound probe The purpose of t...

Page 55: ...here applicable save the appropriate log files by pressing Ctrl F or Alt D Remember to save the log files for each day on a separate floppy disk labelled accordingly Note Restart the application before resuming clinical scanning Table 1 10 Phone numbers for Customer Assistance LOCATION PHONE NUMBER USA GE Medical Systems Ultrasound Service Engineering 9900 Innovation Drive Wauwatosa WI 53226 Servi...

Page 56: ...IVID Q N SERVICE MANUAL 1 26 Section 1 8 Customer Assistance 1 8 2 System manufacturer Table 1 11 System manufacturer MANUFACTURER PHONE NUMBER FAX NUMBER GE VINGMED ULTRASOUND A S STRANDPROMENADEN 45 P O BOX 141 NO 3191 HORTEN NORWAY 47 3302 1100 47 3302 1350 ...

Page 57: ...id i n or Vivid q N ultrasound unit Included are descriptions of the electrical and facility requirements that must be met by the purchaser A worksheet is provided at the end of this chapter see Figure 2 2 on page 2 10 to help ensure that all the required network information is available prior to installation Table 2 1 Contents in Chapter 2 Section Description Page Number 2 1 Overview 2 1 2 2 Cons...

Page 58: ...bdued and indirect when a scan is being performed Therefore a combination lighting system dim bright is recommended Keep in mind that lighting controls and dimmers can be a source of EMI which could degrade image quality These controls should be selected to minimize possible interference 2 2 4 Time and Manpower Requirements Site preparation takes time Begin pre installation checks as soon as possi...

Page 59: ...ts Electrical specifications for the Vivid i n Vivid q N system are as follows 2 2 5 2 Inrush Current Inrush current is not a factor for consideration due to the inrush current limiting properties of the power supplies 2 2 5 3 Site Power Outlets A dedicated AC power outlet must be within reach of the unit without requiring the use of extension cords Other outlets adequate for the external peripher...

Page 60: ...quencies magnetic fields and transients in the air or wiring They also generate EMI The Vivid i n and Vivid q N ultrasound units comply with limits as stated on the EMC label However there is no guarantee that interference will not occur in a particular installation NOTE Possible EMI sources should be identified before the unit is installed and should not be on the same line as the ultrasound syst...

Page 61: ... external connection requires a magnet wrap at each end Install the shield over the front of the card cage Loose or missing covers or RF gaskets allow radio frequencies to interfere with the ultrasound signals Replace broken RF gaskets If more than 20 or a pair of the fingers on an RF gasket are broken replace the gasket Do not turn ON the unit until any loose metallic part is removed and replaced...

Page 62: ...and calibration and testing must also be performed by qualified personnel The products involved and the accompanying electrical installations are highly sophisticated and special engineering competence is required All electrical work on these products must comply with the requirements of applicable electrical codes The purchaser of GE equipment must utilize only qualified personnel to perform elec...

Page 63: ...r in their case or on a rack Material to safely clean probes performed using a plastic container never metal In the case of a network option An active network outlet in the vicinity of the ultrasound unit A network cable of appropriate length regular Pin to Pin network cable An IT administrator who will assist in configuring the unit to work with your local network A fixed IP address is required R...

Page 64: ...elow shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements Figure 2 1 Recommended Floor Plan 4 3m x 5 2m 14 ft x 17 ft 36 IN 92 CM Dedicated Power Outlets Hospital Network Dedicated Analog Telephone Line for Connection to InSite GE Cabinet for Software and Manuals optional Vivid i n or Vivid q N ...

Page 65: ...sferring images in this manner frees up the on board monitor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier and at a lower cost 2 3 4 4 DICOM Option Pre Installation Requirements To configure the Vivid i n Vivid q N ultrasound unit to work with other network connections the network administrator must provide...

Page 66: ...vices Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Dept Vivid i n SN Type Floor Room REV Comments CONTACT INFORMATION Name Title Phone E Mail Address Site Scanner IP Settings IP Address Subnet Mask Default Gateway Echo Server GEMNet Server EchoPac PC Server Name Remote DB User Name Name AE Title Remote Archive Setup IP Addres...

Page 67: ...M Network Information Vivid i n or Vivid q N Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP Addresses Default DICOM APPLICATION INFORMATION NAME Store 2 Store 3 Store 1 MAKE REVISION IP ADDRESSES PORT AE TITLE Store 5 Store 6 Store 4 Storage Commit MPPS Work list ...

Page 68: ...tence of any variances from the basic installation Make sure system and probes have been subject to acclimation period Environmental cooling is sufficient Lighting is adjustable to adapt to varying operational conditions of the scanner Electrical facilities meet system requirements EMI precautions have been taken and all possible sources of interference have been removed Mandatory site requirement...

Page 69: ...for checking and testing the unit probes and external peripherals for electrical safety are also provided NOTE Depending on the customer s specific requirements the Vivid i Vivid q N portable ultrasound scanner may have been supplied with or without the SafeLock Cart optional Where applicable the installation procedures for each scenario are clearly identified in this chapter Table 3 1 Contents in...

Page 70: ... Only qualified service personnel should carry out servicing and troubleshooting 3 2 2 1 System Acclimation Time Following transport the Vivid i n Vivid q N system may be very cold or hot Allow time for the system to acclimate before being switched ON Acclimation requires 1 hour for each 2 5oC increment when the temperature of the system is below 10o C or above 35o C Table 3 2 Average Installation...

Page 71: ...be lifted CAUTION If the unit is very cold or hot do NOT turn ON power to the unit until it has had sufficient time to acclimate to its operating environment CAUTION To prevent electrical shock connect the unit to a properly grounded power outlet Do NOT use a three prong to two prong adapter as this defeats safety grounding CAUTION To ensure proper grounding connect this equipment to a receptacle ...

Page 72: ...ackage even if the package itself is undamaged To make it easier to detection if the handeling during transportation has been unproper a set of Tilt Shock indicators have been attached to the transportation box 3 2 4 2 Position of the Tilt Shock Indicators The Tilt Shock indicators have been attached to the right side of the transportation box as illustrated in the figure below Figure 3 1 Tiltwatc...

Page 73: ...cator has been trigged Make a remark on the Post Delivery Checklist about the trigged indicator and then follow the rest of the instructions in subsection 3 3 2 Damage in Transportation If no Continue with the instructions in Section 3 3 Receiving the Vivid i n and Vivid q N WARNING WARNING For Vivid i n and Vivid q N portable ultrasound scanners supplied from the factory with a dummy battery when...

Page 74: ...ivery is accepted or signed for by a GE representative or hospital receiving agent 2 Report the damage to the carrier Whether noted or concealed damage MUST be reported to the carrier immediately upon discovery or in any event within 14 days after receipt and the contents and containers held for inspection by the carrier A transportation company will not pay a claim for damage if an inspection is ...

Page 75: ... q N Transportation Box Label details Figure 3 2 Label mounted Vivid i n label is sshown in this example Figure 3 3 Vivid q N Transportation Box Label details HANDLE WITH CARE FRAGILE HANDLE WITH CARE KEEP DRY PROTECT FROM MOISTURE TOP UPRIGHT TRANSPORTATION STORAGE RECYCLING RECYCLABLE WOOD RELATIVE HUMIDITY BETWEEN 30 and 95 KEEP TRANSPORTATION TEMPERATURE BETWEEN 10 C and 60 C KEEP AIR PRESSURE...

Page 76: ... N 3 3 2 3 The Vivid i n Transportation Box Label details Figure 3 4 Vivid i n Transportation Box Label details HANDLE WITH CARE FRAGILE HANDLE WITH CARE KEEP DRY PROTECT FROM MOISTURE TOP UPRIGHT TRANSPORTATION STORAGE RECYCLING RECYCLABLE WOOD RELATIVE HUMIDITY BETWEEN 30 and 95 KEEP TRANSPORTATION TEMPERATURE BETWEEN 20 C and 60 C ...

Page 77: ...AL Chapter 3 System Setup 3 9 Section 3 4 Unpacking the Equipment CAUTION Please read this section fully before unpacking the Vivid i n and Vivid q N ultrasound unit Figure 3 5 Shipping Box for the Vivid i n only Figure 3 6 To remove the Lid of the Box Break the seal Open the Lock ...

Page 78: ...ackages Examin the Transportation box closely at time of delivery as described in the procedure that follow Figure 3 7 Transportation box lid removed Table 3 6 Shipping Carton Dimensions and Weights Description Height Width Depth Weighta a Weight is approximate and will vary depending on supplied peripherals Scanner with probes peripherals and accessories 85 cm 87 cm 66cm 35 kgs Empty ...

Page 79: ...eady to mark the check list 3 Remove the surface packing material paper pad silica gel etc Continue to remove additional packing material as applicable during the following procedure steps 4 Carefully remove the box containing the probes 5 Take out the package containing the following accessories cables AD DC ECG Network Isolation Box and software CD 6 Verify the content against the Delivery Note ...

Page 80: ...t voltage The Voltage settings are 220 240V AC 3 4 5 EMI Protection The Vivid i n Vivid q N Ultrasound Unit has been designed to minimize the effects of Electro Magnetic Interference EMI Many of the covers shields and screws are provided primarily to protect the system from image artifacts caused by this interference For this reason it is imperative that all covers and hardware are installed and s...

Page 81: ...list the contents of the shipping carton that are additional to the Vivid i n and Vivid q N These include external accessories and probes as well as a regional language support kit and optional peripherals as ordered Ensure that all relevant components are present before completing the installation NOTE The transportation box contains an External Accessory Kit and Optional Peripherals Check List W...

Page 82: ...aged items Verify the proper illumination of all the control panel buttons 3 Probes Check all probes for wear and tear on the lens cable and connector Look for bent or damaged pins on the connector and in the connector socket on the unit Verify that the EMI fingers around the probe connector socket housing are intact Check the probe locking mechanism and probe switch 4 LCD Display Clean the LCD di...

Page 83: ...clean the peripherals in accordance with the manufacturer s directions To prevent EMI or system overheating dress the peripheral cables inside the peripheral cover 5 AC System Check the AC board connectors and the associated cabling for good connection and proper insulation Verify that the connections are secured 6 Power Cord Check the power cord for cuts loose hardware tire marks exposed insulati...

Page 84: ... are visible from the front and side of the ultrasound unit 1 Display Monitor tilts up and down 2 Speakers Two loudspeakers for Doppler sound 3 Control Panel Contains the alphanumeric keyboard and the buttons used to operate the ultrasound unit 4 Rear Handle 5 Air Vents 6 Probe Ports Two active probe connectors one for a pencil probe 7 ECG Connector 8 ON OFF Switch 9 Alphanumeric Keyboard and Oper...

Page 85: ... ultrasound unit rear panel and external peripheral accessory connectors 1 Two interchangeable USB ports digital printer CD RW and other peripherals 2 Docking connector currently not in use 3 Port for DC IN AC Adapter 4 SVGA Output VCR option or CRT monitor option 5 LAN 10 100 Base TX Ethernet network connector 6 PCMCIA port for PC card 7 Ejection lever for PCMCIA device Figure 3 10 View of the Vi...

Page 86: ...d i n Vivid q N SafeLock Cart Components Label Item Label Item 1 Top Shelf Assembly with handle 8 Bottom Cover Rear AC Assembly Cover 2 Probe Shelf with probe holders 9 Rear Lower Cover 3 LAN Isolation Box 10 SafeLock Cart Bottom Assembly 4 Peripheral Power Outlet 11 Front Wheels with locking lever 5 Rear Upper Cover 12 Rear Wheels 6 AC Distribution Assembly behind cover 13 AC Cable Hook 7 Bottom ...

Page 87: ...safety rating plate step 1 3 5 4 1 Connection and Usage of the optional Modo Cart Please refer to the Vivid i n Cart User Manual Supplement to Vivid i n User manual Direction Number FL092096 3 5 5 System Voltage Confirmation 3 5 5 1 System Voltage Settings Verify that the Vivid i n and Vivid q N is set to the correct voltage The Vivid i n and Vivid q N voltage settings are found on the base unders...

Page 88: ...nd unit will have noisy images Check the grounding of the power cord and power outlet Be aware of RF sources Keep the unit at least 5m 16 4 ft away from other EMI sources Special shielding may be required to eliminate interference problems caused by high frequency high powered radio or video broadcast signals Replace and or reassemble all screws RF gaskets covers and cores After you finish repairi...

Page 89: ... i n Vivid q N ultrasound scanner as follows Printer Categories and Outline of Procedure Instructions below Configuring Hot Keys to Activate Printing Direct from the Control Panel on page 3 22 Configuring Printing Orientation Portrait or Landscape and Paper Size on page 3 23 3 7 1 1 1 Printer Categories and Outline of Procedure Instructions NOTE The printers approved for use with the Vivid i n Viv...

Page 90: ...ure either the Print button or Alt Print to activate a variety of outputs 5 Use the left or right arrows below to add any of the available options listed on the left as Outputs listed on right 6 From the listed Outputs select highlight the required output right click and select Properties to configure the output device accordingly Printing may now be activated direct from the Vivid i n Vivid q N C...

Page 91: ...r first accesses the system at the Applications level from the Additional Outputs screen By selecting the Advanced option under Selected Devices the user accesses the Printer Properties dialog box and proceeds with the configuration procedure as explained below An additional procedure is required to configure the Sony UP D897 Printer This is described in Configuring the Sony UP D897 Printer on pag...

Page 92: ...leting the Hardware Installation 3 Click Configure in the Printer Properties dialog to enter Print Setup 4 In the Orientation area select Portrait or Landscape as needed 5 From the Paper Size drop down list select the required paper size Figure 3 15 Printer Properties Dialog Box Figure 3 16 Printer Setup ...

Page 93: ... 6 Click Properties The printer properties screen opens Figure 3 17 7 Verify that the selected orientation and paper size are correct 8 Click OK twice and then once more to return to the Advanced Outputs tab Figure 3 17 Printing Properties Selecting Orientation Orientation From scroll down list select Portrait ...

Page 94: ...nal non isolated power source WARNING Do not attempt to use a different type of DeskJet Color Printer brand or model other than the DeskJet Color Printer provided by GE Medical Systems The ultrasound system is an extremely sensitive and complex medical system Any unauthorized peripherals may cause system failure or damage NOTICE IMPORTANT Do not connect the DeskJet Color Printer to the Vivid i n V...

Page 95: ...t attempt to use a different type of DeskJet Color Printer brand or model other than the DeskJet Color Printer provided by GE Medical Systems The ultrasound system is an extremely sensitive and complex medical system Any unauthorized peripherals may cause system failure or damage No SafeLock Cart connection via USB hub As stated above Safety considerations must be taken into account The DeskJet Co...

Page 96: ...tion procedure for mounting the Vivid i q system on the SafeLock cart as described in Mounting the Vivid i n Vivid q N on the SafeLock Cart optional on page 3 65 NOTE When using the Vivid i q system mounted on the SafeLockcart the HP Officejet Pro 8000 printer must be connected to the AC outlet at the rear of the SafeLock cart see Figure 3 20 1 Connect the printer to the AC power outlet located at...

Page 97: ...ed with the specific B W Printer supplied by GE Medical Systems WARNING Do not attempt to use a different type of B W Printer brand or model other than the B W Printer provided by GE Medical Systems The ultrasound system is an extremely sensitive and complex medical system Any unauthorized peripherals may cause system failure or damage No SafeLock Cart connection via USB hub As stated above Safety...

Page 98: ...lick the Tcpip tab on the Connectivity screen see Figure 3 14 on page 3 23 2 Select Advanced Settings see Figure 3 22 3 The Network Connection screen opens Figure 3 23 Click the Up button to open the Windows Control Panel 4 Navigate to Printers and Faxes and double click Sony UP D897 to open the Sony UP D897 Printer Configuration dialog Figure 3 22 Tcpip screen Figure 3 23 Network Connections show...

Page 99: ...ID Q N SERVICE MANUAL Chapter 3 System Setup 3 31 5 From the menu select Printer Properties 6 The Sony UP D897 Properties screen opens Select Printing Preferences Figure 3 24 Sony UP D897 printer configuration dialog box Figure 3 25 Sony UP D897 Properties Screen ...

Page 100: ...g the Hardware Installation 7 From the Printing Preferences dialog under Orientation A Under Paper click the scroll down arrow and select the required paper size B Under Orientation choose the required printing orientation 8 Click OK and return to the Connectivity screen Figure 3 26 Printing Preferences screen ...

Page 101: ...ted to be used with the specific Digital Color Printer supplied by GE Medical Systems WARNING Do not attempt to use a different type of Digital Color Printer brand or model other than the Digital Color Printer provided by GE Medical Systems The ultrasound system is an extremely sensitive and complex medical system Any unauthorized peripherals may cause system failure or damage No SafeLock Cart con...

Page 102: ...eed for additional power The modem should be connected to the lower USB port on the rear connectors panel When connecting the modem there is no preference as to which cable to connect first One end of the telephone cable can be connected to the wall outlet and the other side to the modem followed by connecting one end of the USB cable to the modem and the other side to the lower USB port on the re...

Page 103: ...he Vivid i n Vivid q N system The installation procedure requires connecting the DVD CD drive The procedure includes installing the Compact Wireless G USB Adaptor software and then the configuration procedure which is performed via the Windows Network Connections support screens For procedure instructions see Wireless USB Network Configuration on page 3 90 WARNING Do not use any other type of wire...

Page 104: ...event the system will stop recognizing the CD DVD drive or any other device connected to this USB port A complete system re boot will then be required to make the USB port operational again The Samsung DVD R RW drive which is designed to operate with the Vivid i n Vivid q N system using the USB port combines the functions of a DVD or CD recorder and a DVD or CD player Table 3 1 Samsung DVD SE TO84...

Page 105: ...Vivid i n Vivid q N system To perform the Database Backup procedure in order to store back up copies of patient Exams data Archive and also to create free space on the local hard disk To export Exams in MPEGVue mode whereby the media can be viewed on any external PC To load or upgrade system software and for the use of GE Healthcare Field Engineers when performing various other functions Note For ...

Page 106: ...amsung DVD CD RW drive rear panel b Disconnect the 5V power supply output cable from the power port on the Samsung DVD R RW drive rear panel 2 Remove the Samsung DVD CD RW drive from the system Installation Procedure 1 Unpack the contents of the Samsung Cable Kit P N S2423956 2 Connect the mini USB connector Figure 3 29 into the mini USB port located at the rear of the replacement Samsung DVD CD R...

Page 107: ...following tests a Insert a CD R disk in the DVD CD RW Drive b Eject and remove the CD R disk to check the eject mechanism functions properly NOTE In the event that problems are experienced when exporting to DVD media the error message Failed to finalize media is displayed It is therefore recommended to use the Medical Grade 5V Power Supply for DVD Samsung Universal P N 2421169 and connect it to th...

Page 108: ...afeLock Cart main column 4 Disconnect both ends of the 5V power outlet one end from the SafeLock Cart main column power and the other from the Samsung DVD R RW drive rear panel 5 Release the locking arm retaining clamp lever located beneath the specially designed shelf DVD R RW drive compartment beneath the upper shelf of the SafeLock Cart see Figure 3 33 6 Slide the Samsung DVD R RW drive out of ...

Page 109: ...nto the USB ports located below the upper shelf of the SafeLock Cart see Figure 3 35 and Figure 3 36 Figure 3 34 Cable for Installation on Vivid i n Vivid q N Cart Figure 3 35 USB Cable Plugged Into the DVD CD Drive and the Cart s USB Connectors Figure 3 36 USB ports Located Below the Support Tray of the Vivid i n Vivid q N Cart USB connectors to be plugged into the USB ports on the cart Mini USB ...

Page 110: ... the USB port located at the rear of the Vivid i n Vivid q N system Figure 3 31 on page 3 39 6 Re connect power to the SafeLock Cart and to the Vivid i n Vivid q N system 7 Switch on the circuit breaker and turn ON power to the Vivid i n Vivid q N system 8 Perform the following test a Insert a CD R disk in the DVD CD RW Drive b Eject and remove the CD R disk to check the eject mechanism functions ...

Page 111: ... by GE Medical Systems WARNING Do not attempt to use a different type of mass storage device other than the MOD Drive supplied with the system The ultrasound system is an extremely sensitive and complex medical system Any unauthorized peripherals may cause system failure or damage No SafeLock Cart connection via USB hub As stated above Safety considerations must be taken into account The MOD Drive...

Page 112: ...rasound system is an extremely sensitive and complex medical system Any unauthorized peripherals may cause system failure or damage Refer to the External USB Hard Disk Installation Procedure section page 8 151 for instructions on initial configuration and connecting the External USB Hard Disk to the Vivid i n Vivid q N portable ultrasound scanner NOTICE IMPORTANT Do not connect the External USB Ha...

Page 113: ...uality No SafeLock Cart connection via USB hub As above If applicable the USB hub must be powered via the isolation transformer Vivid i n Vivid q N mounted on SafeLock Cart As above 3 7 2 7 Footswitch The Footswitch USB connection in the different scenarios is explained below No SafeLock Cart connection directly to the Vivid i n Vivid q N system The Footswitch may be connected to either of the USB...

Page 114: ...ware Installation Vivid i n Vivid q N mounted on SafeLock Cart As above NOTE After connecting the Footswitch to the system define the required functionality settings as follows 1 Press Config to access the Configuration screen 2 Select Imaging then activate the Application tab see Figure 3 41 Figure 3 41 Application Tab ...

Page 115: ...VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 47 3 Using the Footswitch Functionality drop down menus select the required functionality for the Left Middle and or Right Footswitch see the example in Figure 3 42 Figure 3 42 Footswitch Functionality Settings ...

Page 116: ...ocated on the Vivid i n Vivid q N rear connector panel see note below Note When performing step 1 it may be necessary to use the VGA adaptor P N 2420728 see Figure 3 44 to connect the end of the VGA cable to the Vivid i n Vivid q N rear connector panel Since some VGA plugs have thicker housing they do not fit properly into the Vivid i n Vivid q N VGA connector socket on the rear panel because the ...

Page 117: ...e correct positions as shown in Figure 3 47 6 When all power and signal cables are connected to the system VCR and video converter proceed as follows in the order stated a Press the power ON switch on the VCR b Press the power ON switch on the video converter c Turn ON power to the Vivid i n Vivid q N system 7 Press the Input Select button see Figure 3 45 on page 3 48 on the video converter three ...

Page 118: ...e Erase simultaneously The following screen will appear Figure 3 48 Intel Dual Display Clone 9 Select the Intel Dual Display Clone icon and press OK see Figure 3 48 NOTE If the VGA cable is not properly connected to the rear panel or in the event that the scan converter device is not switched ON you may see a single icon displayed as shown in Figure 3 49 Figure 3 49 Single Icon ...

Page 119: ...le on the cart It is possible to use a different type of VCR brand and model However it is highly recommended to use a medical graded VCR as this will ensure that both safety and image quality levels meet universal medical standards 1 Connect one end of the VGA cable to the VGA OUTPUT located on the Vivid i n Vivid q N unit refer to Figure 3 10 on page 3 17 2 Connect the other end of the VGA cable...

Page 120: ...ON 1 VIVID I N AND VIVID Q N SERVICE MANUAL 3 52 Section 3 7 Completing the Hardware Installation 3 7 2 9 Connecting the VGA Splitter Figure 3 51 VGA Splitter Figure 3 52 Cables and Bracket for Wall Mounting Power Cable VGA Cable DVI Cables ...

Page 121: ...ld Service Engineer must be present when connecting the VGA cable top right in Figure 3 52 between the Vivid i n Vivid q N system and the GE Video Splitter Isolation box 1 Connect the VGA cable Figure 3 52 to the VGA input port at the rear of the VGA splitter see Figure 3 53 and Figure 3 54 Connect the other end of the cable into the Vivid i n Vivid q N system s VGA output port 2 Plug the VGA Spli...

Page 122: ...d for further connection to other DVI devices Refer to Figure 3 54 above 4 When connecting the DVI output cable to an external monitor press V Out on the Vivid i n Vivid q N system console The screen shown in Figure 3 55 is displayed 5 Under Multiple Display activate the Intel R Dual Display Clone radio button Monitor is now selected as the Secondary Device as shown in Figure 3 55 Figure 3 55 Sele...

Page 123: ...following safety precautions Do not expose the charger or power supply to water or conductive liquids this is not a sealed case Do not open the charger or power supply case There are no user serviceable parts inside Do not cover the fan exhaust or obstruct the airflow as this will cause overheating Use only the manufacturer s 24V 2 5A power supply Place the charger in a cool location away from ext...

Page 124: ...be faster if the battery is fully discharged to begin with The recalibration time is approximately 9 hours for an empty battery and approximately 14 hours for a fully charged battery Note Calibration is initiated each time the button is pressed so it is not recommended to press the recalibration button part way through the recalibration cycle During the charging procedure the green LED will be see...

Page 125: ...f the battery charger unit No action is required if only a recharge is required as the charger will automatically begin to charge the battery The blue calibration LED will flash to indicate that the battery is undergoing the recalibration cycle There may be a short delay before the calibration begins During calibration the discharge resistors will heat up and the fan will operate to maintain tempe...

Page 126: ... side of the unit Probes can be connected or changed any time as described below regardless of whether the system is powered ON or OFF CAUTION Handle the probe gently while connecting and disconnecting Do NOT touch the patient and any of the connectors on the ultrasound unit simultaneously including ultrasound probe connectors 1 Place the probe s carrying case on a stable surface and open the case...

Page 127: ...e cord so that it is free to move and is not resting on the floor When the probe is connected it is automatically activated CAUTION Do not bend the probe cable acutely Fault conditions can result in electric shock hazard Do not touch the surface of probe connectors which are exposed when the probe is removed Do not touch the patient when connecting or disconnecting a probe NOTE It is not necessary...

Page 128: ...CTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 3 60 Section 3 7 Completing the Hardware Installation 3 7 4 1 Available Probes For a list of the available probes please see Section 9 7 Probes on page 9 9 ...

Page 129: ...external ECG is to be used See also information on Ext ECG Module Connector provided in ECG Module on page 5 39 Note For optimal ECG operation use only electrodes that meet universal standards see Table 3 5 Use of the External Respiratory option is enabled by removal of the ECG Board and installation of the Internal ECG Respiratory Board For instructions refer to the ECG Respiratory Module Assembl...

Page 130: ...wer plug should not under any circumstances be altered to a configuration rated less than that specified for the current DO NOT use an extension cord or adaptor plug Refer to 2 2 5 Electrical Requirements on page 2 3 for more details NOTE It is recommended to connect the AC DC converter to the ultrasound unit first before connecting it to the AC wall outlet 1 Connect the AC DC converter to the ult...

Page 131: ...ower cable connector from the AC input socket on the SafeLock Cart or the AC DC converter of the system 3 Unplug the mains power cable from the AC wall outlet socket 4 Unplug the AC DC power connector from the Vivid i n Vivid q N rear panel CAUTION Make sure the system is powered OFF or in standby mode before disconnecting the Vivid i n Vivid q Nunit from the electrical outlet and proceed as descr...

Page 132: ...ystem was shut down If the probe has been removed since the system was last used the currently connected probes and their available applications are displayed and selected by default 3 7 7 2 Switching the System OFF NOTE After turning OFF the system wait at least 10 seconds before turning it on again The system may not be able to boot up if power is recycled too quickly The system can be switched ...

Page 133: ...n the sliding guides on either side of the shelf as shown below NOTICE Only the SafeLock Cart supplied by GE Ultrasound and or any medical grade cart which has been approved as suitable for use with the system by GE Ultrasound should be used with the Vivid i n and Vivid q N portable ultrasound scanner NOTICE The Vivid i n and Vivid q N system and the SafeLock Cart should be operated at a distance ...

Page 134: ... of one finger between the SafeLock Cart arm rest and the front of the scanner 3 If necessary perform the alignment procedure as described in 3 8 2 Adjusting the SafeLock Cart Locking Mechanism on page 3 68 4 Connect the LAN Cable from the cart to the LAN connector on the Vivid i n and Vivid q N rear panel 5 Connect the Docking Connector to the rear of the scanner as shown below Figure 3 63 Vivid ...

Page 135: ...d i n and Vivid q N on the SafeLock Cart cont d 6 Connect the USB Cable from the cart to the USB port on the Vivid i n and Vivid q N rear panel Figure 3 64 Vivid i n and Vivid q N Rear Panel on page 3 66 Use the lower port the upper port is dedicated for the Bluetooth connector 7 If necessary connect the DVD to the USB hub on the cart ...

Page 136: ...et and the device see Figure 3 67 on page 3 68 2 Working from the underneath the main shelf use a Phillips screwdriver to loosen the two screws half turn counter clockwise holding the middle locking tooth in position on both the left and right sides of the main shelf as shown below 3 Check that the Vivid i n and Vivid q N portable ultrasound scanner is locked correctly onto the two rear locking te...

Page 137: ...vant sub sections Table 3 6 Configuration Procedures SUB SECTION DESCRIPTION PAGE NUMBER 3 9 2 Vivid i n Vivid q N Configuration 3 69 3 9 2 2 Enter Location 3 70 3 9 2 3 Date and Time Adjustments 3 71 3 9 2 4 Language Selection 3 72 3 9 2 5 Units of Measure 3 74 3 9 3 Service Screen Set up 3 75 3 9 4 Configuring Peripherals 3 81 3 9 4 1 Approved Peripherals 3 81 3 9 4 2 Network Peripherals Optiona...

Page 138: ...on Window see Preparations If needed select SYSTEM The System Settings Window is displayed 2 Select the Hospital field see a in Figure 3 68 and type the name of the hospital max 64 characters The 24 first characters of this name are displayed on the scanning screen s title bar after restart All 64 characters are displayed on the image properties on saved images after restart 3 Select the Departmen...

Page 139: ... New time is displayed 4 Select the preferred Time Format see c in Figure 3 69 24 the 24 hour format is used 12 the 12 AM PM hour format is used 5 Select the preferred Date Format see d in Figure 3 69 DD Date two digits MM Month two digits YYYY Year four digits EU the European DD MM YYYY format is used US the American MM DD YYYY format is used 6 Select Default Century 1900 2000 or None see e in Fi...

Page 140: ...n page 4 10 Select SYSTEM if needed The System Settings Window is displayed 2 Use the Language drop down dialog see Figure 3 70 to select your preferred language for the on screen interface The following languages are available ENG English NOR Norwegian DEU German ITA Italian FRA French ESP Spanish RUS Russian after selection continue to step 3 GRE Greek after selection continue to step 3 DAN Dani...

Page 141: ...egional Settings The Regional and Language Options dialog box opens 4 Under Standards and formats in the language field scroll down to the desired language e g Russian and click OK Reboot the system The interface and keyboard input will be in Russian Note For the Greek language only the user interface is supported It is not possible to input Greek from the keyboard Table 3 9 Language Adjustments c...

Page 142: ...EXPECTED RESULT S 1 Open the Configuration Window see Log On to the System as ADM on page 4 10 Select SYSTEM if needed The System Settings Window is displayed 2 Use the drop down Units dialog see Figure 3 71 to select Metric or US Units The selected units Metric or US will be used for measurements as soon as the unit has been restarted 3 Restart the scanner All the changed settings will be used af...

Page 143: ...ontrast and Backlight Intensity Keyboard Setups Printer Setup Disable TCPIP filtering NOTE The Disable TCPIP Filtering option should only be used for Troubleshooting procedures after which it it is recommended to return immediately to the TCPIP Filtering enabled state 3 9 3 2 Open Service Screen 1 Press CONFIG F2 and log on as Adm see Log On to the System as ADM on page 4 10 2 Select SERVICE to vi...

Page 144: ...ngs as required For instructions see LCD Display Adjustments and Calibration on page 6 2 3 9 3 5 Keyboard Setup 1 Press F2 select Service Tab The Select Keyboard Setup dialog is displayed 2 Click Keyboard Setup The Regional and Language Options dialog box is displayed which has controls for the following described below Regional Options affects regional formatting of numbers dates times etc Langua...

Page 145: ...ARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 77 3 9 3 5 1 Regional Options Select appropriate Format and Location then click Apply Figure 3 75 Regional Options ...

Page 146: ...VIVID Q N SERVICE MANUAL 3 78 Section 3 9 Configuration 3 9 3 5 2 Languages 1 Under the Languages tab click Details to select appropriate Language 2 If necessary activate applicable Supplemental Language Support check box then click Apply Figure 3 76 Languages ...

Page 147: ...13 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 79 3 9 3 5 3 Advanced 1 Under the Advanced tab select a language version of the non unicode programs you want to use then click Apply Figure 3 77 Advanced ...

Page 148: ...es where there is a need follow the instructions provided by GE Medical Systems or the instructions in the respective printer installation manual 1 Select ADD PRINTER to start the Add Printer Installation Wizard 2 Follow the instructions in the Wizard to install a new printer 3 When done perform a full system shutdown then boot up the system into Normal Scannng mode Restarting the system allows th...

Page 149: ...or Vivid q N system refer to the External USB Hard Disk Installation Procedure on page 8 151 DVD CD R Drive via USB Printer Deskjet HP 450 via USB Virtual printer any of the DeskJet printers can be used as virtual printers when the user is in a remote location 3 9 4 2 Network Peripherals Optional The printers detailed below are connected via Ethernet TCP IP Network Lexmark Color LaserWriter Networ...

Page 150: ...he spooled print jobs will be printed out automatically Sometimes this feature is device dependent Therefore in order to initiate the manual print of all spooled jobs process proceed as described below 1 Press ALT P The Printers and Faxes window is displayed 2 Right click on the destination virtual printer see Figure 3 80 and select Open The Virtual Printer Job Queue window is displayed 3 From the...

Page 151: ...ound scanner is mounted on the SafeLock Cart always use the lower USB port for the Hub 5 The Vivid i n and Vivid q N ultrasound scanner and all connected devices working with scanner should be powered and connected to the SafeLock Cart only For the SafeLock Cart there are two basic peripheral configurations that are recommended Configuration Recommendation 1 DVD always plugged into port 1 of the U...

Page 152: ...page 4 10 2 Select ADMIN lower part of window 3 Select the SYSTEM ADMIN tab 4 Select NEW to open the New Key dialog where you type the Software Option Key Serial String 5 Type the Password Software Option Key 6 Press OK to save the new setting 7 Close the Config window 3 9 5 3 Remote Check and Configuration Contact the OnLine Center for InSite Checkout Figure 3 83 System Admin screen Figure 3 84 T...

Page 153: ...eway A Vivid i n Vivid q N scanner configured for eVue Images or loops may be transmitted to a designated PC e Vue PC on the network in compact MPEG format and will appear at the review PC in almost real time 3 10 1 1 The Dataflow Concept Communication between the Vivid i n Vivid q N ultrasound unit and other information providers on the network takes the form of dataflows Each dataflow defines th...

Page 154: ... HD dataflow The local database is used for patient archiving Images are stored to internal hard drive LocalArchive MOD dataflow The local database is used for patient archiving Images are stored to a MOD Worklist Local Archive DICOMServer Int HD dataflow Search in the DICOM Modality Worklist the patient found is copied into local database The patient information and the examination results are st...

Page 155: ...om Vivid i n Vivid q N to an EchoPAC PC Workstation In this configuration the data is transferred from the Vivid i n Vivid q N to a dedicated EchoPAC PC Workstation over an Ethernet connection The connection may be done in one of three ways with a crossover cable as a Peer to Peer Network connection via a Local Area Network LAN set up for this special use only connection via the hospital network L...

Page 156: ...connect the Vivid i n Vivid q N system to an EchoPAC PC and a network printer Figure 3 86 Ethernet Switch Identifications Figure 3 1 Position Label Function X To PC To connect an end station or a Server to the Port To HUB To connect a HUB or another Switch to the Port Table 3 11 MDI Button Position Descriptions MDI Button Controls the fifth connector making it crossed or non crossed Informative LE...

Page 157: ...PAC PC Workstation Figure 3 88 Hospital Network Connections See the EchoPAC PC Service Manual Part Number EP091298 for details about the EchoPAC PC and the required set up procedures EchoPAC PC Vivid7 Do not use the fifth connector and the switch can be in any position Use only Standard non crossed Ethernet cabling Vivid3 Using standard non crossed cables and MDI switch set to position HUB connect...

Page 158: ...h the GE label in the DVD drive 3 Boot up the system in maintenance mode and navigate to Windows Explorer 4 Create a new folder on the system s D drive and name it WIFI DRIVER 5 Copy the contents of the disk to the WIFI DRIVER folder 6 Disconnect the DVD drive from the system 7 Plug the wireless USB device connected to the R adaptor into the USB port at the rear of the Vivid i n Vivid q Nsystem Re...

Page 159: ...nstall from a list or specific location Advanced see Figure 3 90 and click Next 10 Select Include this location in search Figure 3 91 and browse to D WIFI DRIVER Figure 3 92 11 Click Next The following screen opens Figure 3 92 Figure 3 90 Install from a List or Specific Location Figure 3 91 Include this Location in Search ...

Page 160: ...ivity Setup 12 Select the WIFI DRIVER folder then click OK The following screen opens 13 Click Next The installation process continues displaying wizard screens until installation is complete 14 Click Finish The installation process is complete Figure 3 92 WIFI DRIVER Folder Figure 3 93 Hardware Update Wizard ...

Page 161: ...d q N scanner screen press Config Select the Connectivity tab and from the Connectivity screen click TCP IP 2 Click Advanced Settings to open the Network Connections screen see Figure 3 94 3 From the Wireless Network Connection screen navigate to Wireless Network Connection 8 and double click to open the Wireless Network Connection 8 screen see Figure 3 95 Figure 3 94 Network Connections ...

Page 162: ...wireless network and click Connect In this example Figure 3 95 the wireless network connection is Vivid_S Testing The Wireless Network Connection dialog box opens see Figure 3 96 5 Type the required network key in the Network key field then click Connect Figure 3 95 Wireless Network Connection 8 Figure 3 96 Wireless Network Connection Dialog Box ...

Page 163: ...screen indicating that the wireless adaptor is connected see Figure 3 98 6 Return to the instructions for Installing the Compact Wireless G Adaptor Software on page 3 90 and repeat the entire procedure in order to configure the second USB port 7 When done proceed to perform the following functionality tests Wireless USB Network Test on page 4 30 Figure 3 97 Wireless Detection Process Figure 3 98 C...

Page 164: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 3 96 Section 3 10 Connectivity Setup Figure 3 99 CISCO_350 Sub directory ...

Page 165: ...d to the network connector to the EchoPAC PC Workstation see Chapter 3 in the EchoPAC PC Workstation Service Manual Connection via a Peer to Peer Network You will need a network hub and one network cable for each unit connected to the hub See 3 10 2 1 Ethernet Switch Connections on page 3 88 Connection via Hospital Network You will need one network cable to connect the Vivid i n Vivid q N to a wal...

Page 166: ...age 2 10 Typically the source for this information is the network administrator Follow the steps below to prepare the scanner for use on the network 3 10 3 2 Select TCP IP Set up Screen 1 Press CONFIG F2 and log on as Adm as described in Log On to the System as ADM on page 4 10 2 If not already selected select CONNECTIVITY from the bottom of the screen 3 Select the TCP IP TAB it is named Tcpip Fig...

Page 167: ... IP address 0 0 0 0 4 In addition the scanner s AE Title must be entered in the DICOM server s setup 3 10 3 4 Set the Remote Archive s Network Information In the Remote Archive Setup area of the screen see example in Figure 3 102 enter the following 1 Remote Archive IP address Default IP Address from factory 10 0 0 4 2 Remote Archive Name Default Remote Archive Name from factory ECHOPAC7 000001 Fi...

Page 168: ...emote Path used for Save As Export from Q Analysis and for exporting error logs with Alt D 1 From the system keyboard press Config and log on as an ADM user with the appropriate password 2 Click Connectivity from the bottom row of buttons on the screen and select the Tools tab The Tools screen opens see Figure 3 103 3 Under Save As PDF Network Path in the Remote Path field type the remote path For...

Page 169: ... log on the default user and password Under Configurable Remote Path User refer to Figure 3 103 type the desired user name and password as configured on the destination PC This is the password usually assigned by the local IT person 3 10 3 7 Save the New Settings 1 Press SAVE SETTINGS to save the new settings The new settings are saved to a common settings file After a restart the settings are als...

Page 170: ...e Vivid i n Vivid q N control panel press the Config button 2 From the System Configuration window click the Connectivity tab at the bottom of the window The Connectivity dialog box opens 3 Select the TCPIP Tab 4 Trackball to the Advanced tab and press the Set button on the control panel 5 Select the Local Area Connection as shown in Figure 3 105 6 Press the Menu button on the control panel 7 From...

Page 171: ...omputer 12 Type in the Subnet mask 255 255 255 0 Note The Subnet mask should be the same for both the network computer and the Vivid i n Vivid q N 13 Click OK 14 Press the Config button on the control panel to close all windows and return to the regular scanning mode NOTE The following steps are performed on the Network Computer 1 On the network computer access the Control Panel as follows a Doubl...

Page 172: ...d q N system 8 Type in the Subnet mask 255 255 255 0 Note The Subnet mask should be the same for both the network computer and the Vivid i n Vivid q N 9 Click OK close all windows and approve all changes Return to regular mode 10 Connect the Vivid i n Vivid q N and network computer LAN connection using a cross over Network cable Normally with RED connector covers on both sides cable not longer tha...

Page 173: ...VID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 105 a Select Start Run b Type CMD then press Enter c Type Ping and IP 10 0 0 1 for Vivid i n or Vivid q N then press Enter If return timing is OK you have an operating network connection ...

Page 174: ...ss the Set button on the control panel 1 On the Vivid i n Vivid q N control panel press the Config button 2 From the System Configuration window click the Connectivity tab at the bottom of the window The Connectivity dialog box opens 3 Select the TCPIP Tab 4 Trackball to the Advanced tab and press the Set button on the control panel 5 Select the Local Area Connection as shown in Figure 3 105 6 Pre...

Page 175: ...ivid q N IP address 10 0 0 1 or the one provided by the local network administrator Note The IP must be different from the IP address used for network computer 12 Type in the Subnet mask 255 255 255 0 or the one provided by the local network administrator Note The Subnet mask should be the same for both the network computer and the Vivid i n Vivid q N 13 Click OK 14 Press the Config button on the ...

Page 176: ...the network computer access the Control Panel as follows a Double click on My Computer icon on desktop b Double click on Control Panel icon 2 Double click on Network Connections icon 3 Double click on Local Area Connection icon The Local Area Connection Properties dialog box opens 4 Select Internet Protocol TCP IP 5 Click the Properties button Figure 3 113 Local Area Connection Properties ...

Page 177: ... 8 Connect the Vivid i n Vivid q N and network computer LAN connection using a regular Network cable The Vivid i n Vivid q N and the network computer are now ready for export import of files Note One can check the connection by pinging to the Vivid i n Vivid q N IP as follows a Select Start Run b Type CMD then press Enter c Type Ping and IP 10 0 0 1 for Vivid i n Vivid q N then press Enter If retu...

Page 178: ... work against the DICOM server the following information has to be entered in the scanner DICOM server IP address subnet mask and eventually the gateway DICOM server port number DICOM server AE title the server application s name Name of Device 3 10 4 2 DICOM Server IP Address Setting on the Scanner 1 Press CONFIG F2 and log on as Adm as described in Log On to the System as ADM on page 4 10 2 Sele...

Page 179: ... VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 111 The selected flow is shown as seen in the example in Figure 3 116 below Worklist Local Archive DICOM Server Int HD Figure 3 116 Worklist Local Archive DICOM Server Int HD Flow ...

Page 180: ...ect Properties to display the Properties dialog 7 Click on Search Criterias Figure 3 117 The Search Criterias window opens Figure 3 117 8 Highlight the modality filter in the list and click on Remove see Figure 3 117 Figure 3 117 Select Worklist Properties Dialog Box Figure 3 118 Search Criterias Window Select Worklist and then select Properties Worklist Properties Dialog ...

Page 181: ...Add to List The Search Criterias window is updated see Figure 3 118 11 Click OK to close the window 12 From the Worklist Properties Dialog select the IP Address down arrow to choose the Worklist Server from the pull down menu NOTE It is not possible to change the setting in the IP Address field by editing it See description below starting with step 13 to change the setting Figure 3 119 Search Crit...

Page 182: ...s below to change the IP Address settings a From the IP Address pull down menu select Modify to display the IPs dialog box b From the IPs dialog box select the server you want to modify then select Modify Figure 3 120 DICOM Server s Properties Dialog Box Figure 3 121 Select Modify Figure 3 122 Select the Unit to Modify Modify Select server ...

Page 183: ... dialog to return to the Properties dialog box f Enter the DICOM server AE Title This entry is case sensitive and must match exactly g Enter the DICOM server port h For some DICOM Servers the default Time out setting 30 is too low Change if necessary by typing a new value seconds i Click OK to close the Worklist Properties dialog box and save the changes Figure 3 123 Edit Name and or IP address Fi...

Page 184: ... the connection to the device verification process The verification process may require up to one minutes When done a sign prefix displayed in front of the listed device indicates whether the test passed green check mark or failed red X Figure 3 125 Verify Connection to a Device Figure 3 126 Verification Result Check Select device to be verified A green check mark indicates that the test passed A ...

Page 185: ...station and Vivid i n Vivid q N being configured to the Mitra support personnel 3 10 5 3 Configuration of Modality Worklist Services Identify Dataflow to be Used 1 Identify which dataflow is to be used at the site If the DICOM Modality Worklist will be used then the dataflow must include a Worklist service 2 On the Connectivity screen select the Dataflow tab 3 From the pull down menu select the na...

Page 186: ...git number for the Value 3 Click the Add to List button 4 Click the OK button Assign DICOM Worklist name to Dataflow 1 Select the Dataflow tab 2 From the Dataflow name pull down menu select the Dataflow which will be used 3 Remove the DICOM Worklist service currently used in the Dataflow 4 From the Select Service pull down menu select the Gateway Modality Worklist service 5 Click the Add button 6 ...

Page 187: ... POLL_DIR b Verify that Text is select for the Export file format Test Connectivity with the Vivid HL7 Gateway 1 Start a new study and acquire an image or use an existing study 2 In the Measurements mode make some of the measurements 3 In the Report screen create and store a report 4 In the Patient screen from the Examination List select the study in which you created the report 5 From the More pu...

Page 188: ...erver for further analysis on the Vivid i n Vivid q N 3 10 6 2 Query Retrieve Setup on the Vivid i n Vivid q N 1 Press CONFIG F2 and log on as Adm as described in Log On to the System as ADM on page 4 10 2 Select Connectivity 3 Select Dataflow 4 Select Query Retrieve from the Name pull down menu see Figure 3 129 on page 3 121 The selected dataflow Query Retrieve is displayed as shown in Figure 3 1...

Page 189: ... System Setup 3 121 5 Select QueryRetrieve so it is highlighted and then select Properties to display the Properties dialog box Figure 3 128 Query Retrieve Workflow Figure 3 129 Select Query Retrieve s Properties SELECT QUERY RETRIEVE AND THEN SELECT PROPERTIES QUERY RETRIEVE PROPERTIES DIALOG ...

Page 190: ...he pull down menu and edit the setup for one of the predefined servers 7 Enter the correct AE Title and Port Number for the DICOM Query Retrieve server in the respective fields in the Query Retrieve screen Change Search Criterias It is possible to set up special Search Criterias for DICOM Query Retrieve In most cases you may leave the Search Criterias as is and skip this adjustment Follow the step...

Page 191: ...MANUAL Chapter 3 System Setup 3 123 2 Select the correct tag from the Select Tag pull down menu 3 If needed type in the value 4 Select Add to List 5 Click OK to close the Search Criterias window Figure 3 131 Select Search Criterias SEARCH CRITERIAS SELECT TAG ADD TO LIST REMOVE ...

Page 192: ...Connection 1 Select the Smiley button to Ping the server 2 If the network connection to the server is OK it will be illustrated by a smiling Smiley A sad Smiley indicates that the network connection is failing Typical cause Network cable not connected 3 When ready click OK to close the DICOM Query Retrieve properties and save changes Figure 3 132 DICOM Query Retrieve Properties SMILEY OK VERIFICAT...

Page 193: ...th Ping and DICOM Ping 1 Select QueryRetrieve from the Selected devices list 2 Select Check If the test passes a pop up dialog is displayed see left illustration in Figure 3 134 Select OK to continue A green check mark to the left of the Selected devices indicates that the test passed successfully Figure 3 133 Check the DICOM Query Retrieve Workflow Figure 3 134 Test Check Passed QUERYRETRIEVE CHE...

Page 194: ...N SERVICE MANUAL 3 126 Section 3 10 Connectivity Setup If the test fails a pop up warning is displayed see left illustration in Figure 3 135 Select OK to continue A red X to the left of the Selected devices indicates that the test failed Figure 3 135 Test Check Failed ...

Page 195: ...TML MPEG and JPG formats for all of the images in the exam It also contains measurements and reports generated and saved during the exam the reports are all saved as files in CHM format Any media generated as MPEGvue also contains an embedded MPEGVue viewer allowing the user to view it on any PC without having to install any additional software The exported exam may be viewed on any PC operating o...

Page 196: ...EGVue Export 1 Press the Config or F2 button and log on as Adm as described in Log On to the System as ADM on page 4 10 2 Select Connectivity 3 Select Dataflow 4 From the drop down menu select Misc Export see Figure 3 136 5 From selected output devices highlight evue and then click the Properties button refer to Figure 3 137 on page 3 129 Figure 3 136 Dataflow Misc Export ...

Page 197: ...omputer name or the remote computer s ID address in the Remote Path field After the path has been typed it is available for selection from the Dataflow Name drop down menu or you can select an existing path from this menu 7 Activate the Copy Media Player Installation check box Note Step 7 is recommended for the first transfer of MPEGVue files This step is not needed for Windows XP After initial tr...

Page 198: ...ll open and select Patient List The Patient List will be displayed with the following header 4 Trackball to the Export softkey in the bottom menu shown below and press SET The Export dialog box opens 5 Under Export To select To MPEGvue in the combo box on the right as shown in Figure 3 139 6 Click OK The Patient List screen is displayed with the following header 7 Trackball to the required Patient...

Page 199: ...s displayed 9 If required click OK then select any additional Patient Files for export 10 When export complete click OK then remove the media containing the exported MPEGVue data from the Vivid i n Vivid q N system Note If any problems occurred with copying the Patient Files during the export process an error message such as the one below will be displayed showing the status of each of the transfe...

Page 200: ...where consultation with another doctor in the hospital in real time is required 3 10 8 2 Preparations Before configuring the Vivid i n Vivid q N for eVue dataflow it is necessary to decide upon a specific PC in the hospital as being the e Vue PC station The service technician then configures the Vivid i n Vivid q N to send images to that specific e Vue PC as described below 3 10 8 3 Configuring th...

Page 201: ...e drop down menu 7 From drop down menu select the Dataflow named Local Archive Int HD eVue 8 Trackball to the Selected devices window and select highlight eVue see Figure 3 140 9 Trackball to the Properties button and press SET The eVue Properties dialog box opens Figure 3 140 Connectivity Dataflow Tab Selected Figure 3 141 eVue Properties Dialog Box ...

Page 202: ... transfer of eVue files This step is not needed for Windows XP After initial transfer of eVue it is recommended to disable this check box The Vivid i n Vivid q N dataflow is set and ready for export import of eVue files into a network computer From now on whenever the Vivid i n Vivid q N is used for scanning and the Store button is pressed another person will be able to view the stored loop on the...

Page 203: ... select Local Archive Int HD eVue as shown in Figure 3 142 4 Using the soft key buttons select CREATE PATIENT 5 Now start scanning Each image you would like to store press the STORE button A few seconds after the Store button is pressed images will be transferred into the remote computer which was previously set as the destination refer to Figure 3 141 on page 3 133 6 To proceed follow to the inst...

Page 204: ...Figure 3 143 and Figure 3 144 requesting the user to download and install some drivers The requested drivers may be downloaded from the Internet or if the DVD CD was pre configured those drivers will be located on the MPEGvue DVD CD itself In addition to the above the installation program will test the configuration of the remote PC and will display a message if any component on the computer is no...

Page 205: ...10 9 1 1 Manually Setting the Network PC for eVue Import 1 On the network computer create a New User by accessing the Control Panel as follows a Double click on My Computer icon on desktop b Double click on Control Panel icon 2 Double click on User Accounts icon Figure 3 144 Download Installation Message Examples 2 ...

Page 206: ...0 Connectivity Setup The User Accounts dialog box opens 3 Select the Advanced Tab 4 Click on the Advanced button The Local Users and Groups dialog box opens 5 Select the Users folder and right click to select New User Figure 3 145 User Accounts Dialog Box Figure 3 146 Local Users and Groups Dialog Box ...

Page 207: ... Password never expires check boxes e Click Create to confirm setting and Close f Close all windows The new user account has been set and is ready for activation Create a new folder for import export and set sharing permissions by proceeding as follows 7 On the network computer double click on the My Computer icon on the desktop 8 Under My Computer window select any local drive for retrieval of da...

Page 208: ...con and select Rename Rename the folder as eVue _Import 9 Setting the Sharing properties for the new folder as follows a Right click on eVue_Import icon and select Properties The eVue_Import Properties dialog box opens as shown in Figure 3 148 on page 3 141 Important When more than one scanner is used to export with eVue it is recommended to configure each scanner with its own folder for exporting...

Page 209: ...ring Tab c Click on the Permissions button The Permissions for eVue Import dialog box opens d Click the Add button Note It is recommended to remove the User Everyone from Share Permissions Figure 3 148 eVue_Import Properties Dialog Box Figure 3 149 Permissions for eVue Import Dialog Box Note It is recommended to remove this user ...

Page 210: ...alog box select all to be active activate all the check boxes as shown below and press OK g From the Select Users Computers or Groups dialog box Figure 3 150 click on the Locations button h In the Locations dialog box select your computer from the list and press OK i In the Select Users or Groups dialog box type E1c2h3o4C5l6i7e8n9t in the empty window as shown in Figure 3 151 Figure 3 150 Select U...

Page 211: ...Setup 3 143 j Click the Check Names button The name and path will be detected by underlining as shown below k Now click OK l From the Permission for eVue_Import dialog box highlight and select E1c2h3o4C5l6i7e8n9t from the list see Figure 3 152 Figure 3 151 Select Users or Groups Dialog Box Check Names ...

Page 212: ...he network computer using the browsing option to select the appropriate path For an example of Vivid i n Vivid q N dataflow settings refer to Figure 3 140 on page 3 133 3 10 9 1 2 Manually Stopping the Firewall on the Network PC NOTE In some cases it is necessary to stop the firewall to allow functions such as Excel export and eVue to access and load files into the Network Computer The Firewall ca...

Page 213: ...VID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 145 c Select Control Panel from the Menu 2 Double click on Network Connections as shown above 3 Select your network connection and right click 4 From the drop down menu select Properties ...

Page 214: ... Uncheck the Internet Connection Firewall check box as shown in Figure 3 153 7 Click the OK button to confirm changes and close all windows Stopping an External Service Firewall 1 On the network computer access the Control Panel as follows a Press Start b Select Settings c Select Control Panel from the Menu 2 Double click on Administrative Tools Figure 3 153 Local Area Connection Properties Dialog...

Page 215: ...rvices dialog box select your firewall program for example Black Ice 5 Right click and select Stop from drop down menu The following message will be displayed indicating that the Firewall is being stopped 6 When finished click Close 7 Now you can try to reconnect the network computer to the Vivid i n Vivid q N system Figure 3 154 Administrative Tools Dialog Box Figure 3 155 Services Dialog Box ...

Page 216: ...he network computer access the Control Panel as follows a Press Start b Select Settings c Select Control Panel from the Menu 2 Double click on Windows Firewall The Windows Firewall dialog box opens 3 Select the General Tab 4 Enable the Off check box as shown in Figure 3 156 5 Press OK 6 Close all windows to return to the scanning screen Figure 3 156 Windows Firewall Dialog Box ...

Page 217: ... may need to obtain these from the local IT person 2 Close all Internet Explorer windows if open 3 From the C Program Files GEULS eStorePlayer Version1_0 folder double click on the UnRegisterAll bat icon as shown in Figure 3 157 below 4 If two messages are displayed one after another shown in Figure 3 158 and Figure 3 159 click the OK button on each of them to close the message Figure 3 157 Uninst...

Page 218: ...N AND VIVID Q N SERVICE MANUAL 3 150 Section 3 10 Connectivity Setup 5 From the C Program Files GEULS eStorePlayer Version1_0 folder delete all files as shown in Figure 3 160 on page 3 151 Figure 3 158 Uninstall Message 1 Figure 3 159 Uninstall Message 2 ...

Page 219: ...E DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 3 System Setup 3 151 6 Restart Windows then confirm that the MPEGVue application is no longer installed Figure 3 160 Deleting All Files ...

Page 220: ...play To read an MPEGVue exam stored on a DVD CD proceed as follows Insert the MPEGVue CD in the computer DVD CD drive The MPEGVue Patient List is displayed see Figure 3 161 below Note In the event that the MPEGvue DVD CD is being run for the first time the program will self install the MPEGvue viewer For more details refer to Installing MPEGVue on a Remote PC on page 3 136 To read MPEGVue exams st...

Page 221: ...splayed in full size in the MPEGVue Screen view pane There are several ways to select a thumbnail image from the clipboard Use the mouse pointer With the pointer on the required image click the left mouse button to select the image Use the up or down keyboard arrow to view the next or previous loop in the exam Click on the Next Image or Previous Image buttons 11 in Figure 3 162 as required 1 Clipb...

Page 222: ... As function an image may be saved to any of the following file formats BMP WMV MPEG4 JPEG To save an image click on the diskette icon 8 in Figure 3 162 A Save As dialog box will appear allowing you to select the desired file format in which to save the image 3 10 9 3 7 Performing Measurements on MPEGVue Images The following measurements can be performed on MPEGVue images Distance and area in 2D i...

Page 223: ...essage is displayed prompting the user to first logon to Outlook It is possible to send images loops or full exams from any eVue PC An eVue PC is a PC which contains the MPEGvue player and an e mail client application e g Microsoft Outlook The currently selected image or the entire examination can be sent by e mail as an attachment to one or several regular e mail messages Refer to Sending a Full ...

Page 224: ... 4 Select an existing recipient address from the drop down menu or type a new address in the Send To field Note Up to 10 addresses can be stored for future selection 5 Click Next The user is prompted to enter a password for encryption of the study in the window displayed Figure 3 163 E mail Dialog Box Sending Current Exam Figure 3 164 E mail Address Window Figure 3 165 Enter Password Window ...

Page 225: ...ail message has been sent from the Outlook Outbox Notes Examinations sent by E mail as attachments are compressed encrypted and password protected If the size of the exam is too large to be sent as a single attachment it is divided into several zip file attachments sent with several independent E mails each less than 1Mbyte in size The extension of the attached files is mpegvue and can only be ope...

Page 226: ...y to send the MPEGVue Player software to the recipient first complete this procedure then proceed to Sending the MPEGVue Player Software by E mail on page 3 159 The E mail address window is displayed 3 Select an existing recipient address from the drop down menu or type a new address in the Send To field Note Up to 10 addresses can be stored for future selection 4 Click Next Figure 3 168 E mail Di...

Page 227: ... 6 Check that the message has been sent from the Outlook Outbox When the message has been sent from the Outbox and received by the E mail recipient the image can now be opened in Windows Media Player Sending the MPEGVue Player Software by E mail Note This procedure is required once only the first time your recipient is receiving MPEGVue image s 1 From the E mail Dialog box activate the Send Player...

Page 228: ...has been sent from the Outbox and has arrived at the addressee the recipient should follow the instructions included in the body of the email in order to install the MPEGVue Player First Time Receipt of MPEGVue Player Software and Images by E mail When a group of messages arrive by e mail and the user wishes to review the exam by MPEGvue proceed as follows 1 From Outlook open the received email me...

Page 229: ...led on a remote computer as described in First Time Receipt of MPEGVue Player Software and Images by E mail on page 3 160 the user is able to manage and review images or exams received via email using the MPEGVue utilities described below 1 From the desktop double click on the MPEGVue Startup icon The MPEGVue Startup Menu opens Figure 3 174 MPEGVue Startup Menu 2 Proceed to use the required utilit...

Page 230: ... From the MPEGVue Startup Menu click on the Get new exams by email button The Start MPEGVue screen is displayed as shown in Figure 3 175 below 2 Click the Outlook E mail button shown in Figure 3 175 The E mail dialog box is displayed refer to Figure 3 168 on page 3 158 3 Activate the Receive Exam radio button then click Next Figure 3 175 Start MPEGVue Screen ...

Page 231: ...mail Inbox is being scanned for studies When complete all studies found in the Inbox are listed in the Found Studies dialog box as shown in Figure 3 176 below 4 Select the required Study from the list Note By default the Open Exams for Review and Delete Received Mail from Inbox check boxes are activated 5 Click Next Figure 3 176 List of Studies Found in Inbox ...

Page 232: ...ils Note The Password is case sensitive In the event that an incorrect password is typed the user is prompted with a warning message as shown here Click OK and enter the correct password 7 Click Next The message Receiving Selected Messages is displayed followed by another message see Figure 3 178 below showing progress of deletion of messages from the Inbox Figure 3 177 Password Window Figure 3 17...

Page 233: ...mail examination If you were previously viewing different MPEGvue Patient exams you will need to switch over to the previous instance of the MPEGvue Receiving and Viewing a Partial Examination In the event that the exam is large and a group of e mails is expected it is possible to view a partial exam even before all of the e mail messages have arrived In addition if an e mail message is lost or co...

Page 234: ...in the example above to select and display images of that exam The MPEGVue screen is displayed as shown in the example in Figure 3 162 on page 3 153 Activate e Vue 1 From the MPEGVue Startup Menu click on the Activate e Vue button The Patient List is displayed listing all e Vue exams received refer to Figure 3 180 above 2 Click on the required examination Exam Date labelled 1 in the example above ...

Page 235: ...from Email tab is activated 2 Manage Exams from Email as follows a Make sure the Manage Exams from Email tab is activated as shown in Figure 3 181 The Patient List displays details in table format of all patient exams received by email Patient ID Patient Name Exam Date etc The current path information is displayed below the Patient List as well as number of studies cumulative size and amount of sp...

Page 236: ...M Only when both these tasks have been performed will these two buttons become enabled c To copy or move the selected exams first click the Browse button and select the required destination folder then click the Copy Selected Patients or Move Selected Patients button as appropriate If necessary click the Review Destination folder button to review the destination folder Patient List During the copy...

Page 237: ...3 The Patient List displays details in table format of all patient exams received via export from E Vue Patient ID Patient Name Exam Date etc The current path information is displayed below the Patient List as well as number of studies cumulative size and amount of space available on the drive By default the destination path Other Folder radio button is enabled b Select the required exam or exams ...

Page 238: ...above The current path information is displayed below the Patient List together with details of the removable media e g CD DVD ROM By default the destination path Other Folder radio button is enabled additional information below shows the Path to destination folder b Select the appropriate Current Path Information if necessary click the adjacent Browse button and select the correct source director...

Page 239: ...AL Chapter 3 System Setup 3 171 e Click the Copy Selected Patients button During the copy operation the following message and progress bar is displayed When done the currently selected exam s are available in the destination folder for review Figure 3 185 Exams Selected for Copying ...

Page 240: ...headings Item Value Minimum Requirements Below this information the eVue Operation Utility buttons enable the user to enable or disable the eVue mode as required An additional button Configure eVue is used to perform the eVue configuration procedure Note In order to enable eVue operation it is necessary to first perform the eVue configuration procedure as described in the following step b Configur...

Page 241: ...uired eVue folder destination drive If applicable enable the peer to peer connection check box then type the correct current IP address and Subnet mask details Click Save to save the new configuration The Enable eVue button is now active as shown in Figure 3 187 below Figure 3 187 System Configuration Ready to Enable eVue Mode ...

Page 242: ...oid system configuration inconsistencies before using the Enable eVue or Disable eVue functions make sure that your PC is connected to the appropriate network and that all network cables are properly connected Otherwise configuration of your PC may become incorrect and the Configuration utility will be unable to execute the Enable eVue or Disable eVue operations In cases of a peer to peer connecti...

Page 243: ... Note On most computers when inserting the DVD CD media with MpegVue files the MpegVue Player will start automatically due to the autoplay option of DVD CD media In the event that any updates are required on the remote computer the program will prompt the user to install the appropriate components as necessary Follow the on screen instructions When done the Patient List is displayed 4 From the Pat...

Page 244: ...e The MpegVue Player can be used as a regular player using the Play Stop Forward and Rewind buttons In addition the user can forward images by e mail save images and perform measurements as previously described The MpegVue Player displays two rows of command keys at the bottom of the screen If the remote computer is set up to view many Tool Bars the lower row of the MpegVue command keys might not ...

Page 245: ...limation Time section on page 3 2 3 11 1 Safety Precautions for Moving the Vivid i n Vivid q N Unit CAUTION Whenever moving the Vivid i n Vivid q N ultrasound unit while mounted on the SafeLock Cart always observe the following safety precautions 1 Proceed cautiously when crossing door or elevator thresholds Grasp the front handle grips or the back handle bar and then push or pull DO NOT attempt t...

Page 246: ...art of the original system documentation This will ensure that all relevant safety and user information is available during the operation and service of the complete system 1 Verify that all details have been entered in the System Installation Details form as described in the System Installation Details section above If this form has not been completed enter the details now 2 Locate the Product Lo...

Page 247: ...ard NOTE The Product Locator Installation Card shown may not be the same as the provided Product Locator card From the factory a sheet with four Product Locator cards for transportation and one for Installation are included Mail the Product Locator Installation Card s to the appropriate product center Figure 3 190 Product Locator Installation Card Example ...

Page 248: ... the installation Specific language versions of the Vivid i n Vivid q N User Manual may also be available Check with your GE Sales Representative for availability During and after installation supplementary documentation such as User Manuals and Installation Manuals for the peripherals must be kept as part of the original system documentation This will ensure that all relevant safety and user info...

Page 249: ...st be performed either as a part of the installation procedure or as required during servicing and periodic maintenance of the Vivid i n Vivid q N ultrasound scanner Note Refer to Chapter 7 Diagnostics Troubleshooting for detailed descriptions of some of the functional checks described in this chapter Table 4 1 Contents in Chapter 4 Section Description Page Number 4 1 Overview 4 1 4 2 General Proc...

Page 250: ...ent Required An empty blank MO Disk P N 066E0490 1 3 GB P N 066E0519 5 4 GB P N 066E0514 8 6 GB ECG Pads ECG Cable IEC EUROPE ASIA ROW P N 2378938 2 At least one probe ideally you should check all the probes used on the system by the site See Section 9 7 Probes on page 9 9 for an overview of available probes ...

Page 251: ...current DO NOT use an extension cord or adaptor plug Refer to 2 2 5 Electrical Requirements on page 2 3 for more details DANGER ALWAYS CONNECT THE UNIT TO A FIXED POWER SOCKET WHICH HAS THE PROTECTIVE GROUNDING CONNECTOR DANGER NEVER USE A THREE TO TWO PRONG ADAPTER THIS DEFEATS THE SAFETY GROUND DANGER ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE POWER PLUG IS THE PROPER HOSPITAL G...

Page 252: ...ing sufficient slack so that the unit can be moved slightly 3 Verify that the power cable is without any visible scratches or any sign of damage 4 Verify that the on site mains voltage is within the limits indicated on the rating label on the base of the Vivid i n and Vivid q N portable ultrasound scanner 5 Connect the Mains Power Cable female plug via the AC Adapter to the Power Inlet DC IN refer...

Page 253: ...ring a normal boot you may observe that a The unit s ventilation fans starts on full speed but slow down after a few seconds listen to the fan noise b Power is distributed to the peripherals Operator Panel control panel Monitor Front End Processor and Back End Processor c The Back End Processor and rest of the scanner starts with the sequence listed in the following steps 1 Back End Processor is t...

Page 254: ... initiates and sets up the Front End electronics and the rest of the scanner 4 The Keyboard backlight is lit 5 As soon as the software has been loaded a 2D screen is displayed on the monitor as shown below NOTE Total time used for start up is typically three minutes Figure 4 3 Start up Progress Bar for Vivid i n Figure 4 4 2D Screen ...

Page 255: ... This is not recommended under normal circumstances Closing the lid will place the system in Standby Mode Note The system will remain in Standby Mode for as long as battery power will allow The System Exit dialog box Figure 4 5 used when switching OFF the unit gives you the choices described below Standby Use this button to select the Standby mode allowing a shorter reboot time Most of the system ...

Page 256: ...tem is unplugged for a longer period of time a full shutdown is automatically performed As described above the battery is automatically charged when the system is plugged to the wall outlet and the power is switched ON It takes approximately 10 hours to charge a flat battery If the system is left in the OFF condition system unplugged from the wall out let for an extended period of time 3 to 5 days...

Page 257: ... in the LCD display Then system shut down commences The time required to power down the unit including the Back End Processor may vary from 10 seconds up to approximately 1 minute The last component that shuts down is the light in the LCD display indicating that you may continue with the next step 3 Disconnect the mains power from the wall outlet only after the fans have stopped This will cut powe...

Page 258: ...adm usr It is possible for the administrator adm to establish unique passwords for each user but don t do it now 3 Select the user adm From the factory the password has been left blank so you may try to log on without any password If the log on fails you may try to use the password service As a last resort contact the hospital s network administrator or person responsible for the ultrasound unit a...

Page 259: ...CE MANUAL Chapter 4 General Procedures and Functional Checks 4 11 5 Select System to access the System Setup window The System Settings window is displayed 6 Proceed to select the required settings Table 4 2 Logging on to the System as ADM cont d Step Task Expected Result s ...

Page 260: ...other side to the rear of the MOD drive When connecting the MOD drive directly to the system it is necessary to use an additional power source As a safety precaution this must be done via an isolation transformer The MOD Drive may be connected once the system is powered ON or after shutdown All software drivers for the MOD Drive are pre configured and installed designated to be used with the speci...

Page 261: ...ce it in its cover box for storage in a safe place 4 2 5 4 Connecting External USB Hard Disk to USB Port The External USB Hard Disk can be connected to either of the USB ports upper or lower on the rear connectors panel One must use the additional power source supplied with the External USB Hard Disk to activate the External USB Hard Disk WARNING Do not attempt to use a different type of mass stor...

Page 262: ...only high grade media types NOTE For BT06 systems it is not possible to use DVD R media only CD R or DVD R 1 Press the Eject button on the right side of the DVD CD Drive once to open the disk tray 2 Put a DVD CD disk into the disk tray with the label side facing up 3 Press the Eject button once to close the disk tray After a few seconds the disk is available for reading formatting or writing depen...

Page 263: ...or the label name Remember that MO Disks have two sides Side A and Side B Each Side must have a unique name Example MO Disks for storing images may be labeled IMAGE_1A Side A on MO Disk 1 IMAGE_1B Side B on MO Disk 2 IMAGE_2A 4 2 7 Formatting Removable Media This procedure describes how to format an MO Disk or DVD CD media Before you continue be sure about what naming convention you will use for l...

Page 264: ...4 16 Section 4 2 General Procedures 4 2 8 Verifying Removable Media 1 Insert the removable media to be verified 2 Select the pre configured removable media from the Config Connectivity Tools pop up menu 3 Press Verify The unit starts to verify the DICOM directory on the selected media ...

Page 265: ...NOTICE Presets should not be saved on the same MO disk as images because the Archive Menu lists the images but not the presets stored on a MO disk Table 4 3 Saving Presets STEP TASK FUNCTION 1 Insert an empty formatted MO disk labelled PRESETS and marked with Vivid i n Vivid q N s serial number and the date into the MO Drive For more information about formatting see Formatting Removable Media on p...

Page 266: ...Presets STEP TASK FUNCTION 1 Insert the MO disk with the archived presets into the MOD 2 Press Config or F2 select Admin then select Restore The Restore window is displayed 3 Enable the User Defined Configuration check box 4 Select the MOD as the Source 5 Press Restore Now The Presets are restored to the System 6 Press Alt E on the alphanumeric keyboard The MO disk is ejected ...

Page 267: ...ve the handle from the maximum lowest position to the maximum upper position Verify that the handle is able to reach the maximum lowest and maximum highest positions smoothly 5 Pick up the system using the handle Ensure that the handle is firmly connected to the system Table 4 6 Alphanumeric Keyboard Test Recommended Step Task Expected Result s 1 Power up the System in Normal Scanning Mode 2 Press...

Page 268: ...ition The position value is 31 2 for all TGC Sliders 7 Perform the ROTARY TEST as follows Rotate the rotary in both directions and press the button at the center of the rotary The rotary is activated properly pressing on the button will display a row of data for pressing and another row of data for releasing Depress Release 8 Perform the ALPHANUMERIC KEYBOARD BACKLIGHT TEST as follows a Change the...

Page 269: ...check the Battery green check box Check the Battery yellow check box Press Set Battery yellow LED is On c Uncheck the Battery yellow check box Check the Extern DC green check box Press Set Extern DC green LED is On d Uncheck the Extern DC green check box Check the Extern DC yellow check box Press Set Extern DC yellow LED is On 12 Perform the LID BUTTON TEST as follows a Press the Diagnostic Info b...

Page 270: ...ve good control and that the x and y coordinate values displayed at the bottom of the screen move digit by digit up and down Movement is smooth no jumping and the trackball has no dead areas i e when the trackball is moved there are no areas in which no changes are seen Table 4 9 System ID Test Recommended Step Task Expected Result s 1 Boot up the System in Regular Scanning Mode 2 Press Config 3 U...

Page 271: ...ehind the unit 6 At the WAV window press the Right Channel button Make sure that the right internal speaker is working Verify that there is clear sound from both speakers at all frequencies Note that right and left are as seen when standing behind the unit RFI Path 1 From the diagnostic test tree select Front End Test RFI Then select RFI Audio Test and press Start During the test a high pitched so...

Page 272: ... are no missing pixels in the Full Screen Green or display in another color 8 Select Blue Make sure there are no missing pixels in the Full Screen Blue or display in another color 9 When done press Close The system will prompt Did the test pass Respond accordingly EXTERNAL MONITOR TEST 1 Connect the external monitor to the system 2 Boot up the System in Regular Scanning Mode 3 Press Config 4 Under...

Page 273: ...Maintenance Mode using the Service Dongle and load the application by double clicking the Echoloader icon on the desktop 2 Connect the ECG cable to the input on the ECG port on the right side of the Vivid i n Vivid q N scanner Connect a Cardiac probe for example 3S RS and press on Physio Set ECG as OFF and Int Resp as ON Connect the ECG leads to yourself as the simulated patient Refer to the instr...

Page 274: ...tory signal is displayed in the Data area as shown in the example below 9 Verify that a single clear Respiratory signal is displayed 10 When done trackball to the Exit button and press Set 11 When prompted by the following message select Yes or No as applicable 12 Press the Esc button on the keyboard to close the Diagnostic Test window 13 When done press Exit button Table 4 12 ECG Respiratory Test...

Page 275: ...t and press Start PASS 8 Select MUX Test then select 3S RS Test and press Start PASS 9 Select TR32 Test then select RX Test Select RSTest with 3S RS and press Start PASS 10 When done press the Esc button Table 4 14 CW Test Recommended Step Task Expected Result s 1 Boot up the System in Regular Scanning Mode 2 Press Alt Config to enter Diagnostics Mode 3 From the Diagnostic Test tree select Front E...

Page 276: ...w all the procedure instructions described in Connecting the VCR on page 3 48 2 Connect an external monitor to the VCR S Video or Video output 3 Connect a 3S RS probe Reset the VCR counter and press the VCR Record button Verify that the recording indicator red dot is displayed on the VCR and the tape counter is running on the VCR 4 Activate CW Mode and record 2 minutes of heart scanning or a simil...

Page 277: ...ozen on the screen 3 Press the Print button Evaluate the print quality Adjust the brightness and or contrast of the B W printer if required The image displayed on the screen is printed on the B W or color printer depending upon the key assignment configuration Table 4 18 Color Printer Test Recommended Step Task Expected Result s 1 Scan a blood vessel in 2D Mode or a phantom simulating a similar CF...

Page 278: ...or is installed in a system that is already configured proceed as described below 4 3 2 4 1 Wireless USB Network Functionality Test 1 With the system up and running press Config 2 Click Connectivity and select Dataflow see Figure 4 7 3 From the Dataflow Name field scroll down list Figure 4 7 select a dataflow that the user works with in day to day operations 4 Under Selected devices Figure 4 7 sel...

Page 279: ...k has been successfully performed as indicated by the green confirmation tick see Figure 4 9 7 Under Selected Devices Figure 4 9 proceed to the next device in this example Remote Database that is under Outputs click Check and proceed as before 8 Continue this procedure until all selected devices have been verified Figure 4 8 Information Check Status Box Figure 4 9 Selected Device Remote Database s...

Page 280: ...le on the Vivid i n or Vivid q N scanner Save several images in the patient file 3 Export the images to the External USB Hard Disk 4 Delete the dummy patient file including images from the Vivid i n Vivid q N database The deleted patient file and images are no longer listed 5 Import the dummy patient file and images from the External USB Hard Disk The patient file is now listed in the Vivid i n Vi...

Page 281: ...at the shelf movement is smooth 3 Press the release handle down while pushing the main shelf back downwards to return it to its lowest position The main shelf returns to its lowest position Table 4 21 Brakes Function Test SafeLock Cart Front Castor Wheels Recommended Step Task Expected Result s 1 Release the wheel lock upper lever on each front castor wheel by pulling it upwards Push and pull the ...

Page 282: ...rmatted successfully is displayed 8 Select the GetVolInfo button The message Media in drive G has following properties Media Label MY_LABEL Media File System Name FAT is displayed 9 Select the Get BTPsector button The message The Bytes Per Sector value of media in drive G is 512 is displayed 10 Select the SetVolLabel1 button The message Media Label set to TEST_LABEL1 is displayed 11 Select the Get...

Page 283: ...st Patient and press Export The Export pop up is displayed 8 Place a blank DVD media into the DVD CD device Note For information on types of media see DVD CD RW Drive on page 3 36 9 Select To CD DVD Archive and press OK A pop up is displayed with the message Current media is not formatted Do you wish to format it 10 Change the label to TEST 1 and press OK A progress bar is displayed 11 Press Selec...

Page 284: ... to confirm delete must confirm three times The Test Patient is deleted from patient list Table 4 25 Plug and Scan Test Recommended Task Expected Result s 1 Press the On Off button on the control console for no more than 3 seconds The system enters Standby mode 2 Turn OFF the system s circuit breaker The system remains in Standby mode indicated by the blinking On Off button LED 3 Turn ON the syste...

Page 285: ...o Date and Time Adjustments on page 3 71 2 Check the language settings Refer to Language Selection on page 3 72 3 Ensure that all of the software options are configured correctly Refer to Software Options Configuration on page 3 84 4 Ensure that the hospital details are configured correctly Refer to Table 3 7 on page 3 70 5 Ensure that the VCR is configured as required for the specific site and co...

Page 286: ... HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 4 38 Section 4 5 Site Log Section 4 5 Site Log Table 4 1 Vivid i n Vivid q N Site Log Date Service Person Problem Comments ...

Page 287: ...ARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 4 General Procedures and Functional Checks 4 39 Table 4 1 Vivid i n Vivid q N Site Log cont d Date Service Person Problem Comments ...

Page 288: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 4 40 Section 4 5 Site Log This page was intentionally left blank ...

Page 289: ...system functions for both the Vivid i n and Vivid q N systems It also describes the power distribution system and probes Table 5 1 Contents in Chapter 5 Section Description Page Number 5 1 Overview 5 1 5 2 General Information 5 2 5 3 Front End Unit 5 6 5 4 Back End Processor 5 21 5 5 External Input Output I O 5 29 5 6 Power Supply System 5 30 5 7 Monitor and Operator Panel 5 34 5 8 ECG Module 5 39...

Page 290: ...e Gray Scale imaging Color M Mode Doppler Different combinations of the above modes 5 2 1 Introduction The Vivid i n and Vivid q N systems are equipped with a highly efficient power supply resulting in low power consumption and minimal acoustical noise Among other significant features are the following Integrated low power Front End uses advanced ASIC and FGPA technologies Open connectivity using ...

Page 291: ...tor Panel to the Front End FE Electronics to the Back End Processor BEP and finally are displayed on the monitor and peripherals In addition the Vivid i n and Vivid q N systems enable acquisition of the following signals ECG and external ECG The ECG module which supports acquisition of three bipolar ECG channels leads I II III is incorporated in the cabinet cage assembly to provide ECG signals to ...

Page 292: ...rator Panel that has ultrasound specific buttons mode dependent softkey buttons and an alphanumeric keyboard 5 2 2 3 Electronics Vivid i n and Vivid q N internal electronics are divided into three main components Front End FE Processor Back End Processor BEP DC DC Power Supply DC DC PS For both Vivid i n and Vivid q N systems internal electronics comprise a single card cage that contains both the ...

Page 293: ... 5 6LJQDOV 56 3UREH 3RZHU 0RGXOH 7 9 6 9 YROWDJHV 9 9 9 9 9ROWDJHV 9 67 9 67 9 9 DWWHU 30 WHUQDO 0RQLWRU 2 73 FK 3UH PS 7 URVV 3RLQW 9 FK 7 URVV 3RLQW K 9 FRQYHUWHU QDORJ HDPIRUPLQJ 0L HU 3 4 3 5 63 7 6 5 0 0 17 3 02 3 4 5 N 86 5 DWD Q 0 86 7 7 67 6 LDJQRVWLF LUFXLW RSSOHU XGLR VZLWFK 6ZLWFKLQJ WR 3UH PS 3HQFLO 3XOVHU 3HQFLO 3UREH 5217 3 7R 75 2 URP 2 75 2 0 0 8 56 73 0 0 ORFN LUFXLW 0 0 6 VWHP 7 ...

Page 294: ... into two subsystems 1 Scanner Subsystem Consists of the RFI Board Scan Controller and tissue processing For a detailed description see RFI Board on page 5 11 2 Front End Subsystem Includes the following Two TR32 Boards each board includes a 32 channel digital Beamformer For a detailed description see TR32 Boards Transmitter Receiver on page 5 15 P M Probe MUX Board includes CW BF HVMUX For a deta...

Page 295: ...REH 3RZHU 0RGXOH 7 9 6 9 YROWDJHV 9 9 9 9 9ROWDJHV 9 67 9 67 9 9 DWWHU 30 WHUQDO 0RQLWRU 2 73 FK 3UH PS 7 URVV 3RLQW 9 FK 7 URVV 3RLQW K 9 FRQYHUWHU QDORJ HDPIRUPLQJ 0L HU 3 4 3 5 63 7 6 5 0 0 17 3 02 3 4 5 N 86 5 DWD Q 0 86 7 7 67 6 LDJQRVWLF LUFXLW RSSOHU XGLR VZLWFK 6ZLWFKLQJ WR 3UH PS 3HQFLO 3XOVHU 3HQFLO 3UREH 5217 3 7R 75 2 URP 2 75 2 0 0 8 56 73 0 0 ORFN LUFXLW 0 0 6 VWHP 7 7HPS RQWURO 73 R...

Page 296: ...IIHU 3XOVHU 3 86 6 9 08 3UREH 08 7 5 6LJQDOV 56 3UREH 2 73 FK 3UH PS 7 URVV 3RLQW 9 FK 7 URVV 3RLQW K 9 FRQYHUWHU QDORJ HDPIRUPLQJ 0L HU 3 4 3 5 63 7 6 5 0 0 17 3 02 3 4 5 N 86 5 DWD Q 0 86 7 7 67 6 LDJQRVWLF LUFXLW RSSOHU XGLR VZLWFK 6ZLWFKLQJ WR 3UH PS 3HQFLO 3XOVHU 3HQFLO 3UREH 5217 3 7R 75 2 URP 2 75 2 0 0 8 56 73 0 0 ORFN LUFXLW 0 0 6 VWHP 7 7HPS RQWURO 73 RQWURO 3 6ZLWFK RQQHF ...

Page 297: ...n the Bottom assembly on the right side as shown in Figure 5 5 Figure 5 5 Bottom Assembly showing Location of Front End WARNING When opening and closing the Keyboard Assembly take care not to damage the cables connected to the BEP indicated in Figure 5 6 above Figure 5 6 Keyboard Assembly in Opened Position BEP RS Probe Connector Power Supply Front End BEP Cables Keyboard Assembly ECG Cable ECG Mo...

Page 298: ...Q091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 5 10 Section 5 3 Front End Unit Figure 5 7 Bottom Assembly showing Position of Boards RFI TR32 P M Probe and MUX Board TR32 Boards located below the RFI Board RFI Board ...

Page 299: ...t Signal Generator Tx voltage Enable Disable Probe Sensing Control of TR32 and P M Boards Ultrasound data acquisition and processing Figure 5 8 RFI Block Diagram 5 3 3 2 Watchdog and Power ON Reset For safety reasons a HW watchdog stops ultrasound transmit and resets the DSP if the DSP stops toggling the watchdog inputs Once enabled it is not possible to disable the watchdog unless the board is re...

Page 300: ...PA Download proceeds as follows 1 The actual FPGA contents bit streams are written into DSP memory from the Host 2 The DSP configures serial port 0 to aid configuration 3 The DSP writes the bit streams to serial port 0 4 The DSP verifies that the FPGAs are configured correctly 5 3 3 5 PCI Connection The RFI Board interfaces directly to the PCI bus via the DSP PCI to the local bus bridge ...

Page 301: ...nction Theory 5 13 5 3 3 6 RFI Board Connectors Figure 5 9 RFI Board Connector Locations Table 5 2 RFI Board Connectors Connector Description P5 To P M P6 PCI Flex to RFI P14 To power supply J1 To TR32 Boards J2 To TR32 Boards P5 to P M P14 to Power Supply P6 PCI Flex to RFI J1 to TR32 Boards J2 to TR32 Boards ...

Page 302: ...reading falls below minimum temperature the RFI stops scanning and disables Tx voltage The minimum temperature references are set to 15 o C A temperature reading as low as 15 o C indicates a probe temperature sensing error Table 5 3 RFI Board Input DC Voltages Input Description Connector PIN Connected From 3 3 V DC voltage from PS P14 1 4 From PS HVP_IN Fixed DC high voltage 80 V P14 9 From PS HVN...

Page 303: ...2 Boards TR32A and TR32B see Figure 5 10 Each TR32 Board is a complete 32 channel ultrasound Front End that includes the following features and functions 32 Transmitters 32 receiving channels T R switch Pre Amp TGC 32 channel Beam Former First stage of Analog Beam Former for CW mode included in the receive input ASIC Clock distribution starting with input at PECL level Figure 5 10 TR32 Board Block...

Page 304: ...ges Input Description Connector PIN Connected From 5 VA DC voltage distributed via RFI Board A indicates analog circuit J2 80 81 82 RFI 1 8 V DC voltage distributed via RFI Board J2 85 96 RFI 3 3 V DC voltage distributed via RFI Board J2 97 110 RFI 5 VA DC voltage distributed via RFI Board A indicates analog circuit J2 78 79 RFI TXPSP High voltage positive Transmit high voltage varies from 2 5 to ...

Page 305: ...two TR32 Boards reside in the RFI Board The two TR32 Boards are linked together in parallel The way in which they operate depends upon the probe currently in use and also on the current type of Transmit Receive mode Transmit Receive modes can be divided into two types as shown in the following examples CW Mode half the channels transmit signals and half the channels receive signals simultaneously ...

Page 306: ...The P M enables up to 64 receive transmit channels and performs the following functions Receiving the channels and connecting them to the TR32 Boards Probe connectors High voltage MUX that selects 64 out of 128 inputs routed to the TR32 boards Pencil s probe Pulser and receive pre amplifier Analog Doppler processor a signal path for CW mode containing analog Beam Former Band Pass filter I Q base b...

Page 307: ...nts and Function Theory 5 19 Figure 5 15 Probe and MUX Board Connector Locations Table 5 6 Probe and MUX Board Connectors Connector Description J1 Pencil Probe Connector J2 RS Probe Connector J3 To Upper TR32 Board J5 To RFI Board J1 Pencil Probe Connector P3 Not connected J5 J3 J2 RS Probe Connector ...

Page 308: ... 12 V DC voltage distributed in P M Board J5 16 21 RFI Board 3 3 V DC voltage distributed in P M Board J5 5 8 RFI Board 5 VA DC voltage distributed in P M Board A indicates analog circuit J5 13 15 RFI Board 5 VA DC voltage A indicates analog circuit J5 9 12 RFI Board HVPSP Transmit high voltage varies from 2 5 to 50 V J5 2 RFI Board HVPSN Transmit high voltage varies from 2 5 to 50 V J5 1 RFI Boar...

Page 309: ... the main controller for the system A block diagram of the BEP is provided in Figure 5 17 on page 5 22 The various components that comprise the Vivid i n Vivid q N BEP are described in the following sections ETX SBC Central Processing Unit CPU on page 5 23 ETX Base Board on page 5 23 Hard Disk on page 5 25 Fan Control on page 5 25 RTC BIOS Battery on page 5 25 Back End Processor Location of Compon...

Page 310: ...CTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 5 22 Section 5 4 Back End Processor Figure 5 17 Vivid i n Vivid q N Back End Processor Block Diagram DFN QG 7 6 DVHERDUG 3 0RQLWRU 0RQLWRU 86 3 0 86 3RUWV ...

Page 311: ...X Base Board provides interfaces to the following components HDD Front End RFI Board USB ports Docking port speaker phones and speaker jacks Keyboard Battery VGA Fans Power supply The Power Management Controller PMC mounted on the ETX Base Board ensures automatic transition between the various power modes stable DC condition and provides power status LED display The Docking port located on the rea...

Page 312: ... the ETX Base Board Figure 5 18 ETX Base Board Block Diagram 5 4 3 2 VGA Interface The interface to the VGA is via the VGA connector on the rear side of the ETX Base Board The VGA port is connected to the DIB where VGA signals are converted to DVI and split into three channels to drive the system LCD display 15 or 17 as applicable and external monitor 5 4 3 3 CardBus Interface Not in use ...

Page 313: ...le space on the hard disk F Drive Swap file partition The partition size is 1GB The following additional partitions are hidden Grub 32MB Repository 10GB Note When connected to the system a DVD device or disk on key will not necessarily be shown as G drive 5 4 5 Fan Control The Fan board holds five fans and is mounted on the left side of the bottom assembly A special fan speed controller chip is us...

Page 314: ...4 Back End Processor 5 4 7 Back End Processor Location of Components Figure 5 20 Views of the Back End Processor BEP To Fan Board From DC DC Power Supply KBD LAN VGA PCI Cable to RFI Internal Battery Docking Connector RTC BIOS Battery PCMCIA Not in use HDD Upper View Lower View USB Port ETX SBC CPU ...

Page 315: ...on the BEP Table 5 9 BEP Connectors Connector Description J7 VGA J19 EXT DC INPUT J18 DOCKING J20 Speakerphones J16 RFI P1 P4a a All connections to P1 P2 P3 and P4 according to ETX specs ETX J5 LCD PANEL J6 KEYBOARD J17 PS UNIT J14 H D Adaptor Card J15 BATTERY J13 CARDBUS J10 RTC BIOS Battery J8 Protection Key J11 J12 FANS J9 Speakers J5 J16 J17 J11 J6 J44 J13 J7 J19 J18 J14 ...

Page 316: ...onnected to the Fan board see Figure 5 19 on page 5 25 Table 5 10 BEP Input DC Voltages Input Description Connector PIN Connected From EXT_DC 20 V DC from PS J17 1 2 3 4 5 PS Battery Varies from 14 V to 16 7 V DC depending on battery capacity Distributed via PS J17 11 12 13 14 15 PS 12 V_BE DC voltage from PS BE indicates Back End J17 19 PS 5 V_STB DC voltage from PS STB indicates Standby J17 20 P...

Page 317: ...nal monitor of VGA to S Video adaptor RJ 45 connector for LAN connection Two USB ports for external peripherals 20V DC inlet main DC power to the system Figure 5 22 Vivid i n Vivid q NSystem Rear Panel Connections Legend 1 Two interchangeable USB ports digital printer DVD RW CD RW and other peripherals 2 Docking connector 3 Port for DC IN AC Adapter 4 SVGA Output VCR option or CRT monitor option 5...

Page 318: ...Figure 5 22 DC DC Power Supply Unit located inside and mounted to the bottom assembly provides the system with both Low Voltage and High Voltage power and also enables recharging of the battery The Power Supply Unit controls the following main functions Input switching stage selects between the external DC source and the internal battery Back End Low Voltage power supply LVPS Standby power supply ...

Page 319: ... q N Power Supply Unit Components Block Diagram 5 6 3 1 Critical Temperature Sensor The PS Unit includes temperature sensors When the monitoring application measures the maximum temperature limit or above it will shut down the system After the temperature settles down the PS unit will enable the DC DC operation 5 6 4 DC Source Selector This circuitry selects the source of the power When the AC to ...

Page 320: ... battery operation to AC operation and vice versa When shutting down the system leave the main power cable connected to keep the battery fully charged NOTE The lithium ion technology used in the system s battery is significantly less hazardous to the environment than the lithium metal technology used in some other batteries NOTE The battery is designed to be replaced every 2 years CAUTION The batt...

Page 321: ...ot solder a battery Do not connect the battery to an electrical outlet CAUTION If the Vivid i n or Vivid q N portable ultrasound system is not being used for periods of one month or longer the battery should be removed during lengthy non use period CAUTION To avoid the battery bursting igniting or fumes from the battery causing equipment damage observe the following precautions Do not immerse the ...

Page 322: ...trates the layout of the keyboard and operator panel The buttons and controls are grouped together for ease of use the different areas of the keyboard and operator panel include the following Soft menu Keys part of the Extended Keyboard TGC GAIN rotary Alphanumeric KB Extended KB Trackball Trackball buttons Front panel Hand rest area of front panel Figure 5 25 Vivid i n Vivid q N System Keyboard a...

Page 323: ... Vivid i n Vivid q N Keyboard Block Diagram 5 7 2 Keyboard and Operator Panel Components 5 7 2 1 Power ON OFF Button The Power On Off button receives 3 3 V from the BEP Once shorted this voltage is routed to the PMC which will activate or perform the appropriate task as required 5 7 2 2 Alphanumeric Keyboard The Vivid i n Vivid q N alphanumeric keyboard with a key arrangement similar to that of a ...

Page 324: ...and Menu Buttons These are equivalent to the normal PC mouse left and right click Set left click Menu right click The operation of these buttons is supported via the PS 2 Board Zoom Trackball and Store Buttons These buttons even though located beside the trackball are part of the External Keyboard buttons and not part of the Alphanumerical or regular trackball buttons The operation of these button...

Page 325: ... standby for driving the power On Off and status LEDs 2 5V BE Back End is used for the remaining circuits The keyboard operates with 5 voltage deviation on each supply The keyboard has current limiting circuitry implemented on the BEP to protect it against excessive current consumption via the USB and or short circuits 5 7 2 10 Keyboard ID The keyboard PCB contains an EEPROM which stores a variety...

Page 326: ...RVICE MANUAL 5 38 Section 5 7 Monitor and Operator Panel 5 7 2 12 LCD Display Unit The display monitor is a 15 LCD and is connected via a flex cable to the BEP as shown in Figure 5 28 Figure 5 28 LCD Display Flex Cable Connection to BEP Flex Cable BEP Monitor LCD Display ...

Page 327: ...o bipolar ECG channels leads I II The External ECG module supports acquisition of a single bipolar ECG input channel ranging from 1 to 1 volt allowing amplification of 1000 The minimal allowed impedance is 10 M ohm The ECG module is located under the keyboard assembly see Figure 5 6 on page 5 9 All power and communication is supplied to the ECG module via the USB hub located in the keyboard module...

Page 328: ...leakage and auxiliary currents apply the values listed in Table 5 12 below taken from table IV in IEC601 1 section 19 3 The ECG signals have double reinforced isolation B a from live parts Hence creepage distance and air clearance in the module are 8 mm and 5 mm respectively to withstand 4000 volts isolation between applied part and grounded part The applied parts and enclosure have basic isolatio...

Page 329: ...w The Vivid i n and Vivid q N Internal ECG Respiratory module is designed to enable acquisition of the following signals ECG Respiratory and external ECG The Internal ECG Respiratory Board is located in the cabinet cage assembly refer to Figure 5 6 on page 5 9 All power and communication is supplied to the ECG module via the USB hub located in the keyboard module as shown in Figure 5 31 below Conn...

Page 330: ...width 3db 1 98 31 2 Hz for Monitor use only Lead Fault Indicator Yes A2D resolution 12 bit 2 complement Sample Rate 1000 Table 5 15 Internal ECG Respiratory Board Specifications External ECG Characteristic Specification Input Range ac 10 V peak to peak Bandwidth 3db 0 8 65 5 Hz 65 5 Hz in firmware Lead Fault Indicator Yes A2D resolution 12 bit 2 complement Sample Rate 1000 Table 5 16 Internal ECG ...

Page 331: ...iratory Monitor output type being used as follows Option 1 Monitor with single ended Positive output Connect the signal lead to pin 2 on the connector Connect the ground shielding to pins 5 and 7 on the connector Option 2 Monitor with single ended Negative output Connect the signal lead to pin 5 on the connector Connect the ground shielding to pins 2 and 7 on the connector Option 3 Monitor with Di...

Page 332: ...on Input Range ac 2v peak to peak A2D resolution 12 bit 2 complement Sample Rate 500 samples second Ground Isolation 4 kV rms Table 5 18 Connector Pin Out J5 ECG Respiration External ECG Pin Number Specification 1 External ECG Input 2 ECG Input Respiration Input Output LL 3 ECG Input Respiration Input Output RA 4 ECG Respiration Output LA 5 External ECG Select Signal Input Short to GND on Cable 6 ...

Page 333: ...ction Theory 5 45 Table 5 19 Connector Pins Out J3 J4 USB Pin Number Specification J3 1 J4 28 J4 30 VUSB 5V DC Input from USB Bus J3 2 J4 8 D USB differential Data signal J3 5 J3 shell D USB differential Data signal J4 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 2 4 6 10 12 14 16 20 22 24 26 GND USB Bus GND J3 J5 J4 ...

Page 334: ...Color Inkjet Printer is used to print out reports and exams and is controlled in one of two ways depending on the system configuration Directly via the USB port Via LAN connection The printing operation to the Color Inkjet Printer is performed by pressing the Alt P keys Normally inkjet printers are used to print out Reports but in some cases also ultrasound images 5 10 1 2 DVD CD Reader and CD Wri...

Page 335: ...ess Network Adapter The Vivid i n and Vivid q N supports a unique type of Wireless Network Adapter For details regarding type and installation refer to Wireless USB Network Adaptor on page 3 35 The Wireless Network Adapter provides a mobile network connection to the local area network The Wireless Network Adapter is supported via the USB port ...

Page 336: ...ts has been designed to provide the best thermal conduction allowing the maximum amount of heat to be removed from the system It is crucial that the system ventilation access is not blocked a distance of at least 20 cm from right and left sides of the system should be kept clear In addition operating the system in an unclean environment could introduce foreign objects into the unit and create bloc...

Page 337: ...6 Service Adjustments Section 6 1 Overview 6 1 1 Purpose of Chapter 6 This section provides references to the procedures required for testing and adjusting the Vivid i n Vivid q N portable ultrasound scanner These tests are optional and may be used to check the system for errors NOTE There are no voltage adjustments required ...

Page 338: ...er as stated in the following procedures NOTE Incorrect calibration of the Vivid i n Vivid q N monitor may also result in less than optimal image quality when reviewing images on a DICOM or EchoPAC station Access the Calibration Screen by pressing ALT Q then follow the relevant procedure instructions described below NOTE On systems with older software versions press CTRL ALT P to enter the Calibra...

Page 339: ... left numbers 1 through 6 see Figure 6 1 and the white bars on the right numbers 1 through 12 are all just visible 6 2 3 Testing Your Contrast and Brightness Settings Observe the V watermark on the lower part of the monitor beside the soft keys on the screen The monitor is set to optimal settings when you just start seeing the V If you do not see the V the monitor is set too dark If the V is clear...

Page 340: ...nts and Calibration To adjust the Blue Tint hold down ALT then use the and keys as required Repeatedly pressing will decrease the blue tint repeatedly pressing will increase the blue tint To assist in making blue tint adjustments an adjustment bar is displayed below the screen Figure 6 5 LCD Display Blue Tint Adjustment ...

Page 341: ... installations System Vivid i n Monitor BOE HYDIS 1 NOTE On older Vivid i q systems where the IDTECH 1 monitor is supplied the Blue Tint 2 option is not applicable NOTE The Blue Tint 2 option is incompatible with IDTECH 1 2 monitors therefore on these systems the Blue Tint 2 option must be turned off To identify which monitor is installed on the system 1 Log on as Administrator 2 Press Config Abou...

Page 342: ...ss the Display Calibration Screen 2 To adjust the blue tint hold down ALT then use the and keys as required Repeatedly pressing will decrease the blue tint repeatedly pressing will increase the blue tint Note To assist in making blue tint adjustments an adjustment bar is displayed below the screen see Figure 6 8 3 Press Alt Q to exit to the Display Calibration Screen Figure 6 7 LCD Display Calibra...

Page 343: ...VID Q N SERVICE MANUAL Chapter 6 Service Adjustments 6 7 6 2 5 2 Turning Off the Blue Tint 2 Option 1 Press Config Service The Service Screen opens see Figure 6 9 2 Under LCD Setup de select Blue Tint 2 to turn off this option Figure 6 9 Service Screen LCD Setup ...

Page 344: ...B Scenario B The user prefers to use the External Monitor as the Primary monitor but does not have access to it and therefore cannot make separate adjustments to it using the controls on the External Monitor itself Moreover the calibration preset currently in use on the Vivid i n Vivid q N system is not optimal for the External Monitor In this scenario adjustments Brightness only can be made to th...

Page 345: ...e of the following solutions 1 Purchase a standard VGA isolated splitter Do not use the GE Healthcare VGA isolated splitter P N 5390022 as this has a special pin configuration dedicated for use with CardioLab MacLab Carto 2 Connect the device to an isolation transformer Ensure that the device is not connected to another non medical device which can cause isolation failure through the third device ...

Page 346: ...ht Intensity on page 3 76 6 2 8 Configuring Print Orientation For information on Print Orientation Portrait or Landscape refer to Configuring Printing Orientation Portrait or Landscape and Paper Size on page 3 23 6 2 9 Video System Adjustments 6 2 9 1 Video System Adjustment Procedure For details on Video System adjustments PAL or NTSC setup refer to Figure 3 46 on page 3 49 6 2 10 Beamformer Cali...

Page 347: ...the end of the working day then allow it to continue to run automatically overnight For details see Figure 6 10 below IMPORTANT During the Battery Calibration procedure no scanning can be performed Note For step by step instructions see Performing the Battery Calibration Procedure on page 6 12 Figure 6 10 Recommendations Battery Calibration Procedure May take up to 4 hours May take up to 2 hours B...

Page 348: ...w the procedure to be completed overnight as outlined in Figure 6 10 1 Click on the battery icon at the bottom of the scanning screen OR from the message window shown below click Recalibrate The Battery Calibration Wizard Page 1 opens Figure 6 11 2 Make sure the AC power cable is connected to the Vivid i n Vivid q N then click Next to commence the first stage of the process Full Charge to maximum ...

Page 349: ...ice Adjustments 6 13 3 Wait for the confirmation window to be displayed showing that the battery is fully charged See Battery Calibration Wizard Page 2 in Figure 6 12 4 Click Next The Battery Calibration Wizard Page 3 opens see Figure 6 13 on page 6 14 Figure 6 12 Battery Calibration Wizard Page 2 ...

Page 350: ...isplay Adjustments and Calibration 5 Disconnect the AC power cable from the Vivid i n Vivid q N then click Next The second phase of the process Full Discharge commences and the Battery Calibration Wizard Page 3 is displayed as shown in Figure 6 14 on page 6 15 Figure 6 13 Battery Calibration Wizard Page 3 ...

Page 351: ...scharge process The Vivid i n Vivid q N system automatically shuts down at the end of this process the battery calibration procedure is now complete 7 The following morning re connect the AC power cable and turn ON power to the system Note The system may now be used normally without any dependence on battery capacity status Figure 6 14 Battery Calibration Wizard Page 3 ...

Page 352: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 6 16 Section 6 2 LCD Display Adjustments and Calibration This page was intentionally left blank ...

Page 353: ...te system problems and failures The Vivid i n Vivid q N ultrasound unit includes built in diagnostic tools that can be accessed from the scanner application Table 7 1 Contents in Chapter 7 Section Description Page Number 7 1 Overview 7 1 7 2 Diagnostics 7 2 7 3 Performing Full System Diagnostics 7 19 7 4 Performing Front End FE Diagnostics 7 24 7 5 Monitoring Tests 7 86 7 6 Performing Back End Dia...

Page 354: ... can be achieved When the Back End is operating you can verify proper voltages and temperatures in the Front End and in the Power Supply To check the Front End parts the tests must follow a logical sequence in order to identify the faulty component For example testing communication between the Front End and the Back End begins by testing the RFI board and the PCI bus before testing all the other b...

Page 355: ...ally updated information The Data Flow Map area on the right is designed to provide an instant visual representation of the system components the color illumination of graphic buttons light blue light gray and status indicators green red yellow indicates which component or set of components is currently undergoing diagnostic testing see the examples in Figure 7 2 on page 7 4 In this way it is easy...

Page 356: ...n on page 7 10 Note When performing a specific test for an intermittent problem it is highly recommended to use the Loop option use the combo box or type the required number of loops as appropriate Looping the test will assist in detecting the problem Note Once the test has been completed a log can be viewed as described in the View Log section on page 7 6 Figure 7 2 Data Flow Map Graphic Color Il...

Page 357: ...formed Diagnostic Test Tree Displays a menu of diagnostic test sequences and tests available for selection Execute Displays the name of the currently selected diagnostic test or test sequence Controls and Special Instructions Displays any special instructions for checks required before performing the test For example Make sure system is calibrated Disconnect probe and press Start The area also inc...

Page 358: ...ncluded in the currently selected test sequence Result Lists test results as either PASS or FAIL Loop Lists name of loop and the sub tests within the loop Date Date test was performed Tests Scroll Bar and Arrows Enables scrolling through the list of Test Log Files A test sequence may have up to four reports Since only one specific test may have failed the scroll arrows enable jumping directly from...

Page 359: ... one by one Erase Button Clears all test sequence log files for that specific currently selected test listed in the Tests window passed and failed tests It is not recommended to clear the test sequence log files unless a test failed due to user intervention or as a result of a false alarm It is necessary to erase each log file separately Exit Button Closes the View Log dialog box Returns to the Di...

Page 360: ...al amplitude first harmonic is displayed in the Report window When there is more than one report listed the first report is displayed by default Double clicking on another report for example Test signal amplitude second harmonic will evoke the display of that report in the Report window as shown in the example in Figure 7 4 below Figure 7 3 View Log Dialog Box ...

Page 361: ... Left clicking on a currently selected channel marks a red square on the report graph and also evokes the display of the following information relative to that specific point Average Deviation standard deviation from the Average Max and Min The signal should be within the limits marked in the white area the dotted line represents the Average as shown in Figure 7 4 Figure 7 4 ADC Test Viewing Repor...

Page 362: ... Test requires approximately 10 minutes for completion 7 2 4 1 1 Full System Test FE and BEP Performs a full sequence of complete Front End and Back End tests automatically that do not require user intervention This is particularly recommended for detecting problems that are intermittent in nature since a test may be paused at any stage and any detected failures reviewed For test procedure instruc...

Page 363: ...s a complete test sequence and includes all the RFI board tests described below For test procedure instructions see the Full RFI Test section page 7 31 RFI FE Cache Test Tests the cache memory The cache memory for the PM TR32A TR32B and the RFI boards is located on the RFI board This test sends a specific pattern into the cache memory and the expected result is to see a predefined pattern as the o...

Page 364: ...tructions see the System BF Digital Test section page 7 49 ADC Test The ADC Test is used to check that 1st harmonic 2nd harmonic 3rd harmonic and background noise are within pre defined limits In this test a signal is generated from the RFI into the TR switches For all 64 channels of the system each signal goes through the following pre amplification filtering analog to digital conversion and focu...

Page 365: ...Bend Pass filters A test signal is shifted from the mixer frequency and then the test checks the filter response Attenuation of the filters in different frequencies should be within pre defined limits shown on the Test report For test procedure instructions see the ACW BPF Test section page 7 75 ACW HPF Test The ACW Analog CW HPF Test is used to check the operation of the High Pass Filters 300Hz a...

Page 366: ...supplied by the battery Source of the power being provided to the system i e external power via the AC DC or via the battery indicating the battery is either being charged or discharged For test procedure instructions see the Battery Parameters Test section page 7 102 7 2 4 3 6 Tx Voltages Test The TX Voltages Test is used to test the transmit part of the power supply to make sure that the Tx volt...

Page 367: ...t without Hard Drive This test performs a full system check as well as a Hard Disk scan The test sequence includes all automatic tests including the Hard Disk For test procedure instructions see the Full PC Doctor Test 50 min section page 7 117 CPU Tests The CPU Tests are used to perform a variety of CPU checks as described below For test procedure instructions see the CPU Tests section page 7 119...

Page 368: ...MMX Logical Tests the MXX commands PAND PANDN POR and PXOR If errors are detected the failed commands are listed CPU MMX Shift Tests the MXX commands PSLL and PSRL If errors are detected the failed commands are listed CPU MMX Data Transfer Tests the MXX commands MOVD and MOVQ If errors are detected the failed commands are listed CPU MMX Misc Tests the MXX commands PCMPEQ PCMPGT PACKSS PACKUS and P...

Page 369: ... video buffer with 18 test patterns one pattern at a time These tests fill the entire screen with a single color Video Data Transfer Tests the graphics acceleration part of the video controller This test fills the screen with black and white concentric squares and rectangles of various sizes and colors If errors are detected the locations of the problems are displayed Video Text Output Prints a te...

Page 370: ...he Network test logs as FAILED For test procedure instructions see the Network Test section page 7 130 Interactive Audio Test The Audio Test is performed to check for proper operation of the audio system the following selections are available Left Channel Right Channel Both Channels and MIDI For test procedure instructions see the Audio Test section page 7 132 Monitor Test The Monitor Test is used...

Page 371: ...specially if the Vivid i n Vivid q N system shows signs of an intermittent problem As described in this section the testing sequence can be paused and restarted at any stage providing a quick and easy method of reviewing any detected faults The different types of Full System Diagnostic Tests are described in detail in the Diagnostic Test Window Menu Options section on page 7 10 For instructions on...

Page 372: ...Diagnostic Test window trackball to Full System Test FE and BEP in the list of tests available for selection from the Diagnostic Test Tree see Figure 7 5 below and press Set The Full System Test FE and BEP name is displayed under Execute and all system components are highlighted blue in the Schematic View as shown below Figure 7 5 Full System FE and BEP Test 2 As indicated in the Special Instructi...

Page 373: ...indicators in the Data Flow map corresponding to the various system components will be highlighted in the appropriate color to indicate the current test status as follows Green All recommended tests for these components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these co...

Page 374: ...ON 1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 22 Section 7 3 Performing Full System Diagnostics 8 If necessary use the scroll bars and or arrows to scroll through the list of Test results Figure 7 6 View Test Log Full System Test BEP and FE ...

Page 375: ... then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that board is fa...

Page 376: ...the proper sequence will help to derive the correct conclusions regarding the source of the problem As described in this section the testing sequence can be paused and restarted at any stage providing a quick and easy method of reviewing any detected faults For instructions on performing a specific test refer to the relevant test in the appropriate sub sections Calibration Tests as described in th...

Page 377: ...gure 7 7 below and press Set The DC Offset Calibration name is displayed under Execute the relevant system components to be tested in this case all components are highlighted light blue in the Schematic View as shown below Figure 7 7 DC Offset Calibration Test 2 As indicated in the Special Instructions area disconnect any probes currently connected to the system and make sure that nothing is touch...

Page 378: ...tus as follows Green All recommended tests for these components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components have passed but not all recommended tests have been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume te...

Page 379: ...ts were very close to the upper or lower limits repeat the DC Offset Calibration test again and review the newly acquired results If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either t...

Page 380: ...ct any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may pref...

Page 381: ... Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the DC Offset Verification Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed or failed as shown in the example in Figure 7 10 8 If nec...

Page 382: ...le upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this board is faulty Continue trouble shooting down to the appropriate board If necessary ...

Page 383: ...stic Test Tree see Figure 7 11 below and press Set The Full RFI Test name is displayed under Execute the relevant system component to be tested is highlighted light blue in the Schematic View RFI as shown below 2 As indicated in the Special Instructions area disconnect any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instr...

Page 384: ...ts have been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the Full RFI Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests p...

Page 385: ...and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that board is...

Page 386: ...y probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer te...

Page 387: ...have been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the RFI FE Cache Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests ...

Page 388: ...7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that boa...

Page 389: ...connect any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may...

Page 390: ...se button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the RFI Analog Circuits Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed or failed as shown in the example in Figu...

Page 391: ...and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that board is...

Page 392: ...r the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 5 Trackball to the Start button and press Se...

Page 393: ...n the RFI Demodulator Short Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed or failed as shown in the example in Figure 7 18 8 If necessary use the scroll bars and or arrows to scroll through the list of Test results Note For this test there is one r...

Page 394: ...7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that boa...

Page 395: ... probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer tes...

Page 396: ...rformed When the upper progress bar reaches approximately midway in the Status area a loud high pitched sound will be heard coming from the speakers Note If no sound is heard it is necessary to perform the BEP Audio Test as described in Audio Test on page 7 132 In the event this test also fails the problem is probably in the BEP refer to Performing Back End Diagnostics on the System on page 7 108 ...

Page 397: ...original problem persists contact your local On Line Center for assistance If errors were identified but the results were very close to the upper or lower limits perform calibration as described in DC Offset Calibration on page 7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not mispla...

Page 398: ...ould be touching the probe connectors 1 In the Diagnostic Test window trackball to TR32 BF Digital Test in the list of TR32 BF tests available for selection from the Diagnostic Test Tree see Figure 7 21 below and press Set The TR32 BF Digital Test name is displayed under Execute the relevant system components to be tested are highlighted light blue in the Schematic View TR32A and TR32B as shown be...

Page 399: ... map corresponding to the various system components will be highlighted in the appropriate color to indicate the current test status as follows Green All recommended tests for these components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components have passed but no...

Page 400: ...sists contact your local On Line Center for assistance If errors were identified but the results were very close to the upper or lower limits perform calibration as described in DC Offset Calibration on page 7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning ...

Page 401: ... probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer tes...

Page 402: ...ed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the System BF Digital Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed or fa...

Page 403: ...ion on page 7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation...

Page 404: ...robes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer testi...

Page 405: ...ss Set When the ADC Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed or failed as shown in the example in Figure 7 28 Note The ADC test has four test reports which are listed in Reports on the right By default the First Harmonic report is displayed in...

Page 406: ... list and the corresponding results are shown in the Report window below as shown in Figure 7 29 10 Use the tools provided to scroll through the channels and view the results as previously described 11 When done trackball to the Close button and press Set to close the View Report window 12 Select the third report the report is highlighted in the Reports list and the corresponding results are shown...

Page 407: ...l through the channels and view the results as previously described 14 When done trackball to the Close button and press Set to close the View Report window 15 Select the fourth report the report is highlighted in the Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 31 Figure 7 30 View Test Log Third Harmonic ...

Page 408: ...lower limits perform calibration as described in DC Offset Calibration on page 7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you...

Page 409: ...any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer ...

Page 410: ...assed but not all recommended tests have been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the Rx Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showi...

Page 411: ...orresponding results are shown in the Report window below as shown in Figure 7 29 10 Use the tools provided to scroll through the channels and view the results as previously described 11 When done trackball to the Close button and press Set to close the View Report window 12 After reviewing the test results proceed as follows If no problems were identified it is recommended to re boot the system t...

Page 412: ... faulty or that another board that supports the operation of that board is faulty Continue trouble shooting down to the appropriate board If necessary perform a specific test If the problem still persists contact the GE HEALTHCARE On Line Center for assistance 13 Trackball to the Exit button and press Set to close the View Test Log dialog box 14 Press the Esc button on the keyboard to close the Di...

Page 413: ...ing the probe connectors 1 In the Diagnostic Test window trackball to ACW IQ Symmetry Test in the list of ACW tests available for selection from the Diagnostic Test Tree see Figure 7 32 below and press Set The ACW IQ Symmetry Test name is displayed under Execute the relevant system components to be tested are highlighted light blue in the Schematic View P M as shown below NOTICE IMPORTANT In order...

Page 414: ...side the progress bar in the Status area As the testing sequence progresses the progress bar will advance to reflect the test progress Status indicators in the Data Flow map corresponding to the various system components will be highlighted in the appropriate color to indicate the current test status as follows Green All recommended tests for these components have been completed and no problems we...

Page 415: ...er for assistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that board is faulty...

Page 416: ...that nothing is touching the probe connectors 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The ACW Matrix Test commences While the test proceeds the name of the t...

Page 417: ...ommended tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you paused trackball to the Start button and press Set When the ACW Matrix Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of...

Page 418: ...rea the dotted line represents the Average as shown in Figure 7 38 7 When done trackball to the Close button and press Set to close the View Report window 8 Select the second report Signal RMS the report is highlighted in the Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 39 9 Use the tools provided to scroll through the channels and view the r...

Page 419: ...ance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that board is faulty Continue tro...

Page 420: ...re that nothing is touching the probe connectors 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The ACW Delay Range Test commences While the test proceeds the name ...

Page 421: ... tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you paused trackball to the Start button and press Set When the ACW Delay Range Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of wh...

Page 422: ...a the dotted line represents the Average as shown in Figure 7 42 7 When done trackball to the Close button and press Set to close the View Report window 8 Select the second report Signal RMS the report is highlighted in the Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 43 9 Use the tools provided to scroll through the channels and view the res...

Page 423: ...ce If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that board is faulty Continue troub...

Page 424: ...lue in the Schematic View P M as shown below 2 As indicated in the Special Instructions area disconnect any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer testing to continue o...

Page 425: ...nents have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components have passed but not all recommended tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you paused trackball to the Sta...

Page 426: ...Gain Test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that board is faulty Continu...

Page 427: ...View P M as shown below 2 As indicated in the Special Instructions area disconnect any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer testing to continue on failure and later y...

Page 428: ...e components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components have passed but not all recommended tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you paused trackball to ...

Page 429: ...e Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 49 9 Use the tools provided to scroll through the channels and view the results as previously described 10 When done trackball to the Close button and press Set to close the View Report window 11 Select the third report BPF2 the report is highlighted in the Reports list and the corresponding resu...

Page 430: ...through the channels and view the results as previously described 13 When done trackball to the Close button and press Set to close the View Report window 14 Select the fourth report Signal SNR the report is highlighted in the Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 51 Figure 7 50 View Test Log ACW BPF Test BPF2 ...

Page 431: ...ssistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that board is faulty Continu...

Page 432: ...atic View P M as shown below 2 As indicated in the Special Instructions area disconnect any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer testing to continue on failure and la...

Page 433: ...components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components have passed but not all recommended tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you paused trackball to th...

Page 434: ... 7 25 If the test passed but the original problem persists contact your local On Line Center for assistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you ...

Page 435: ...ny probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 4 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer t...

Page 436: ... been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the Pencil RX Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed ...

Page 437: ...able upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of that board is faulty Continue trouble shooting down to the appropriate board If necessar...

Page 438: ...is displayed under Execute all system components to be tested are highlighted light blue in the Schematic View as shown below 2 As indicated in the Special Instructions area disconnect any probes currently connected to the system and make sure that nothing is touching the probe connectors 3 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from...

Page 439: ... these components during the performance of the test Yellow The tests executed so far on these components have passed but not all recommended tests have been performed 6 At any stage trackball to the Pause button and press Set if required 7 To resume testing from the point where you paused trackball to the Start button and press Set When the Full System Monitoring Test sequence is complete the Dia...

Page 440: ...assed and that result values are within the pre defined Max and Min limits as shown in Figure 7 59 10 Select the second report Temperatures the report is highlighted in the Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 60 Figure 7 59 View Test Log Full System Monitoring Test Fixed Voltages ...

Page 441: ...b test passed and that result values are within the pre defined Max and Min limits as shown in Figure 7 60 12 Select the third report Fans Speed the report is highlighted in the Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 61 Figure 7 60 View Test Log Full System Monitoring Test Temperatures ...

Page 442: ...assed and that result values are within the pre defined Max and Min limits as shown in Figure 7 59 14 Select the fourth report Battery Parameters the report is highlighted in the Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 62 Figure 7 61 View Test Log Full System Monitoring Test Fans Speed ...

Page 443: ... passed and that result values are within the pre defined Upper and Lower limits as shown in Figure 7 62 16 Select the fifth report Tx Voltages the report is highlighted in the Reports list and the corresponding results are shown in the Report window below as shown in Figure 7 63 Figure 7 62 View Test Log Full System Monitoring Test Battery Parameters ...

Page 444: ...erform calibration as described in DC Offset Calibration on page 7 25 and then repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is fau...

Page 445: ...Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Fixed Voltages Test commences While the test proceeds the name of the test currently in progress Fixed Voltages is displayed in the Execute field and the current Loop and specific test are shown below The messages Init Done and Please Wait are displayed ...

Page 446: ... 7 65 Note For this test there is one report Voltages which is listed in the Reports window on the right By default this report is displayed in the Reports window below 7 Check that the test passed and that result values are within the pre defined Max and Min limits as shown in Figure 7 65 8 When done trackball to the Exit button and press Set to close the View Report window 9 After reviewing the ...

Page 447: ... and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this board is faulty Continue trouble shooting down to the appropriate board If necessary perform a specific test If the problem still persists contact the GE HEALTHCARE On Line Center for assistance 10 Trackball to the Exit button and...

Page 448: ...ely you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Temperatures Test commences While the test proceeds the name of the test currently in progress Temperatures is displayed in the Execute field and the current Loop and specific test are shown below The messages Init Done and Please Wait are displayed beside the pro...

Page 449: ...complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed or failed as shown in the example in Figure 7 67 Note For this test there is one report Temperatures which is listed in the Reports window on the right By default this report is displayed in the Reports window below 7 C...

Page 450: ...hen repeat the test If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this board is faul...

Page 451: ...natively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Fans Speed Test commences While the test proceeds the name of the test currently in progress Fans Speed is displayed in the Execute field and the current Loop and specific test are shown below The messages Init Done and Please Wait are displayed beside the pr...

Page 452: ... is one report Fans Status which is listed in the Reports window on the right By default this report is displayed in the Reports window below 7 Check that the test passed and that result values for each fan are within the pre defined Max and Min limits as shown in Figure 7 69 8 When done trackball to the Exit button and press Set to close the View Report window 9 After reviewing the test results p...

Page 453: ...s and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this board is faulty Continue trouble shooting down to the appropriate board If necessary perform a specific test If the problem still persists contact the GE HEALTHCARE On Line Center for assistance 10 Trackball to the Exit button an...

Page 454: ...ate the Loops checkbox and select the required number of Loops from the combo box 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Battery Parameters Test commenc...

Page 455: ...To resume testing from the point where you paused trackball to the Start button and press Set When the Battery Parameters Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed or failed as shown in the example in Figure 7 71 Note For this test there is one...

Page 456: ... that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the test fails again repeat the Fan Speed Test If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this board is faulty Continue trouble shooting down to the appropriate ...

Page 457: ...vate the Loops checkbox and select the required number of Loops from the combo box 3 Adjacent to On Fail select either Stop or Continue as required Stopping on failure enables you to review a failed test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Tx Voltages Test commences Whi...

Page 458: ... 6 To resume testing from the point where you paused trackball to the Start button and press Set When the Tx Voltages Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests passed or failed as shown in the example in Figure 7 73 Note For this test there is one rep...

Page 459: ... upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this board is faulty Continue trouble shooting down to the appropriate board If necessary pe...

Page 460: ... on page 7 25 7 6 1 1 Extended Keyboard Test NOTE For a description of this test refer to the information in the Extended Keyboard Test on page 7 14 1 In the Diagnostic Test window trackball to Extended Keyboard Test in the list of BEP tests available for selection from the Diagnostic Test Tree see Figure 7 74 below and press Set The Extended Keyboard Test name is displayed under Execute the relev...

Page 461: ...nfluence on any of the other ATGC sliders above or below i e all other sliders remain stationary 4 Repeat step 3 to test each of the other ATGC sliders 5 Check all keyboard keys by pressing each in turn as follows a Press one key make sure the depressed key is shown as depressed in the Test Screen b Release the key make sure the released key is no longer showing as depressed in the Test Screen 6 R...

Page 462: ...are activated illuminated green Patient CW Application etc alternatively check each one one by one individually b trackball to the Backlights button under Lights Control and press Set Make sure that all the extended keyboard backlighting is correctly illuminated 13 Trackball to the Battery Light button top right of Test Screen and press Set Make sure that all the corresponding LEDs is activated il...

Page 463: ...yed as shown in Figure 7 76 below 17 Trackball to the Exit button and press Set to close the View Test Log dialog box 18 Proceed as follows If the test passed but the original problem persists contact your local On Line Center for assistance 19 Press the Esc button on the keyboard to close the Diagnostic Test Window Figure 7 76 View Test Log Extended Keyboard Test ...

Page 464: ... see Figure 7 77 below and press Set The ECG Test name is displayed under Execute the relevant system components to be tested are highlighted light blue in the Schematic View BEP Peripherals Keyboard as shown below 4 Under the Special Instructions area activate the Loops checkbox and select the required number of Loops from the combo box 5 Adjacent to On Fail select either Stop or Continue as requ...

Page 465: ...ostics Troubleshooting 7 113 7 Make sure that the resolution scale is set to 1 Sec as shown in Figure 7 78 above 8 Trackball to the Open Device button and press Set The ECG signal is displayed in the Data area as shown in the example below Figure 7 78 ECG Device Test Screen Figure 7 79 Test ECG Data ...

Page 466: ...gnal is displayed on the screen as shown in Figure 7 79 10 When done trackball to the Exit button and press Set 11 When prompted by the following message select Yes or No as applicable 12 Proceed as follows If the test passed but the original problem persists contact your local On Line Center for assistance 13 Press the Esc button on the keyboard to close the Diagnostic Test Window ...

Page 467: ...cription of this test refer to Full Back End Test without Hard Disk on page 7 15 1 In the Diagnostic Test window trackball to Full Back End Test w o Hard Disk Scan in the list of PC Doctor tests available for selection from the Diagnostic Test Tree see Figure 7 80 below and press Set The Test name is displayed under Execute the relevant system components to be tested are highlighted light blue in ...

Page 468: ...far on these components have passed but not all recommended tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you paused trackball to the Start button and press Set When the Full Back End Test w o Hard Disk Scan sequence is complete the Diagnostic Test window displays the Finished message the View Test Log wind...

Page 469: ...at supports the operation of this board is faulty Continue trouble shooting down to the appropriate board If necessary perform a specific test If the problem still persists contact the GE HEALTHCARE On Line Center for assistance 10 Trackball to the Exit button and press Set to close the View Test Log dialog box 11 Press the Esc button on the keyboard to close the Diagnostic Test Window 7 6 1 3 2 F...

Page 470: ...mended tests for these components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components have passed but not all recommended tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To resume testing from the point where you...

Page 471: ... or that another board that supports the operation of this board is faulty Continue trouble shooting down to the appropriate board If necessary perform a specific test If the problem still persists contact the GE HEALTHCARE On Line Center for assistance 10 Trackball to the Exit button and press Set to close the View Test Log dialog box 11 Press the Esc button on the keyboard to close the Diagnosti...

Page 472: ...t test status as follows Green All recommended tests for these components have been completed and no problems were identified Red Problems were identified in these components during the performance of the test Yellow The tests executed so far on these components have passed but not all recommended tests have been performed 5 At any stage trackball to the Pause button and press Set if required 6 To...

Page 473: ...n Line Center for assistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this boar...

Page 474: ...moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Hard Drive Full Test w o Scan commences While the test proceeds the name of the test currently in progress is displayed in the Execute field and the current Loop and specific test are shown below The messages Init Done and Please Wait are...

Page 475: ...quired Linear Seek Random Seek Funnel Scan 8 When done trackball to the Exit button and press Set to close the View Report window 9 After reviewing the test results proceed as follows If no problems were identified it is recommended to re boot the system this is obligatory when prompted If the test passed but the original problem persists contact your local On Line Center for assistance If errors ...

Page 476: ...ed test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Full Memory Test commences While the test proceeds the name of the test currently in progress is displayed in the Execute field and the current Loop and specific test are shown below The messages Init Done and Please Wait are ...

Page 477: ...ed Memory Pattern Test Memory Parity Test 8 When done trackball to the Exit button and press Set to close the View Report window 9 After reviewing the test results proceed as follows If no problems were identified it is recommended to re boot the system this is obligatory when prompted If the test passed but the original problem persists contact your local On Line Center for assistance If errors w...

Page 478: ...d test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Full Video Test commences While the test proceeds the name of the test currently in progress is displayed in the Execute field and the current Loop and specific test are shown below The messages Init Done and Please Wait are di...

Page 479: ...emory Video Data Transfer Video Text Output 8 When done trackball to the Exit button and press Set to close the View Report window 9 After reviewing the test results proceed as follows If no problems were identified it is recommended to re boot the system this is obligatory when prompted If the test passed but the original problem persists contact your local On Line Center for assistance If errors...

Page 480: ...iled test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The PCI Test commences While the test proceeds the name of the test currently in progress is displayed in the Execute field and the current Loop and specific test are shown below The messages Init Done and Please Wait are displa...

Page 481: ...esults proceed as follows If no problems were identified it is recommended to re boot the system this is obligatory when prompted If the test passed but the original problem persists contact your local On Line Center for assistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning proper...

Page 482: ...failed test the moment it fails Alternatively you may prefer testing to continue on failure and later you may review the report 4 Trackball to the Start button and press Set The Network Test commences While the test proceeds the name of the test currently in progress is displayed in the Execute field and the current Loop and specific test are shown below The messages Init Done and Please Wait are ...

Page 483: ... results proceed as follows If no problems were identified it is recommended to re boot the system this is obligatory when prompted If the test passed but the original problem persists contact your local On Line Center for assistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning prop...

Page 484: ...see Figure 7 96 below and press Set The Test name is displayed under Execute the relevant system components to be tested are highlighted light blue in the Schematic View BEP Peripherals Keyboard as shown below 2 Trackball to the Start button and press Set The interactive PC Dr for Windows Audio Test screen is displayed as shown in Figure 7 97 3 Trackball to the Left Channel Right Channel Both Chan...

Page 485: ...ter reviewing the test results proceed as follows If no problems were identified it is recommended to re boot the system this is obligatory when prompted If the test passed but the original problem persists contact your local On Line Center for assistance If errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplace...

Page 486: ... in the list of Interactive PC Doctor tests available for selection from the Diagnostic Test Tree see Figure 7 99 below and press Set The Test name is displayed under Execute the relevant system components to be tested are highlighted light blue in the Schematic View BEP Peripherals Keyboard as shown below 2 Trackball to the Start button and press Set The interactive PC Dr for Windows Monitor Test...

Page 487: ...st 5 When in the displayed screen press any key to return to PC Dr for Windows Monitor Test screen 6 When done trackball to the Close button and press Set to exit the test When the Monitor Test sequence is complete the Diagnostic Test window displays the Finished message the View Test Log window opens automatically listing all tests performed and showing details of which tests and sub tests passed...

Page 488: ... errors were identified results exceeding acceptable upper or lower limits verify that all system cables and boards are connected not misplaced and functioning properly as described in Components and Function Theory If the problem persists this indicates that either the specific board you tested is faulty or that another board that supports the operation of this board is faulty Continue trouble sh...

Page 489: ...Site ExC is only available in English at this time there is no multilingual platform available 7 7 1 1 Service Log In 1 From the Scanning Screen click the Global Service icon in the status bar at the bottom of the screen The Service Login dialog box is displayed as shown below Figure 7 102 Service Login Dialog Box 2 From the Select User Level drop down list select GE Service Then in the Enter Pass...

Page 490: ... I N AND VIVID Q N SERVICE MANUAL 7 138 Section 7 7 InSite ExC The lower section next page of the GEMS Service Home Page is illustrated in Figure 7 104 Figure 7 103 GEMS Service Home Page upper section Figure 7 104 GEMS Service Home Page lower section ...

Page 491: ...curity The service interface has various different security and user levels Users are only granted access to the tools they are authorized to use as described in the following table 7 7 2 Error Logs Page The Error Logs page enables the user to view the various log categories for problem analysis When the Error Logs page is selected the Log Viewer selector is displayed in the left pane of the scree...

Page 492: ...e 7 107 Click a category to navigate to the desired log category Figure 7 106 Log Viewer 7 7 2 1 Examples and Brief Description of Selected Categories When selecting System in the Logs drop down menu the following tabs are displayed as shown in Figure 7 108 The following menu categories are explained in the sections below System Logs on page 7 141 Utilities on page 7 142 Search on page 7 143 Exit ...

Page 493: ...7 7 2 1 1 System Logs Displays all the system logs including errors and additional details see Figure 7 109 Figure 7 109 System Logs Note When a log category has been selected it is possible to select additional log categories from the Log menu that appears at the top of the current screen see the red rectangle in Figure 7 109 ...

Page 494: ... form For example select Power or Temperature from the Log option on the Logs Viewer menu and then select Plot Log to display the results as shown in Figure 7 110 Figure 7 110 Plot Log Window Plot Page Enables you to view the results of the System log in graphic form showing the distribution of the log information according to packages For example select System from the Log option on the Logs View...

Page 495: ...gs Viewer menu enables you to enter case sensitive text that you want to find from the log page that you are currently viewing as shown below Figure 7 112 Search for Text in Log or Page 7 7 2 1 4 Exit The Exit option on the Log Viewer menu enables you to exit the Log Viewer Select the Exit Log Viewer option to return to the Common Service Desktop Figure 7 113 Exit Option ...

Page 496: ...ed without the user s intervention 7 7 3 1 Diagnostics Page Overview The left pane of the Diagnostics page is shown in Figure 7 114 below There are five main categories Calibration Full System Test Front End Monitoring BEP You can expand each category to drill down to the various diagnostic tests by clicking on the plus sign adjacent each category For an example see Figure 7 115 When a diagnostic ...

Page 497: ...ially grey Upon completion of a diagnostic the background color changes to indicate the updated status either Red Fail Green Pass or Grey Abort Loop Count Field Enables you to enter the number of times that the diagnostic test should be performed Any numeric value between one and four digits can be entered This field is queried after you press Execute to start a diagnostic test Execute Button Enab...

Page 498: ...ion on page 7 25 DC Offset Verification on page 7 28 7 7 3 2 2 Front End The Front End directory tree showing all the Front End diagnostic tests is shown in Figure 7 116 The Front End diagnostic tests are described in the following sections RFI RFI FE Cache Test on page 7 34 RFI Analog Circuits Test on page 7 37 RFI Demodulator Short Test on page 7 40 RFI Audio Test on page 7 43 TR32 TR32 tests wi...

Page 499: ... Test on page 7 12 ACW Matrix Test on page 7 13 ACW Delay Range Test on page 7 13 ACW Pre Mixer Gain Test on page 7 13 ACW BPF Test on page 7 13 ACW HPF Test on page 7 13 Pencil Pencil Rx Test on page 7 13 7 7 3 2 3 Monitoring The diagnostic tests accessed under Monitoring are described in the following sections Full System Monitoring Test on page 7 86 Fixed Voltages Test on page 7 93 Temperatures...

Page 500: ...ests is shown in Figure 7 117 The BEP diagnostic tests are described in the following sections Extended Keyboard Test on page 7 14 ECG Test on page 7 14 PC Doctor Automatic Full Back End Test w o Hard Disk Scan on page 7 115 Full PC Doctor Test 50 min on page 7 117 CPU Tests on page 7 119 Registers Level 2 Cache Math Register MMX SSE SSE2 Figure 7 117 BEP Diagnostics Directory ...

Page 501: ...e detected the address of the block in which the error was detected is shown During testing these diagnostics will also look for parity errors and other exceptions If encountered these are added to the test log as errors see Memory Test section page 7 124 Full Memory Test Pattern Test Uses several test patterns to test as much memory as possible Advanced Pattern Test Video Video Test All Test the ...

Page 502: ...le or DSL modems The Network Communications Test checks if your PC s network card works and if your PC properly communicates over a network It makes a list of all the remote PCs on the network that the NIC uses and sends information in the form of several data packets from the tested PC to the host PCs These data packets are then returned by the host PCs to the tested PC where the contents of the ...

Page 503: ... Quality Page 7 7 5 Calibration Page Figure 7 119 Calibration Page Note This page is not populated in this version When the Calibration page is selected as shown in Figure 7 119 although not populated you can perform different calibration procedures as described in Chapter 6 For additional calibration procedures refer to the following sections DC Offset Calibration see DC Offset Calibration on pag...

Page 504: ...hen the Configuration page is selected you can access the Software Options Interface and install software options for the specific Vivid i n Vivid q N system No changes should be made in this page without the help of GE Medical Systems Service Personnel as any modifications made by untrained personnel can cause damage to the ultrasound unit Figure 7 120 Configuration Page ...

Page 505: ...uptive Mode when the OLC or field engineer requests the user to invoke the Disruptive mode the user will be asked to confirm Disruptive Mode by clicking the Yes button as shown in Figure 7 122 Disruptive mode enables the OLC or field engineer to take control of the system for problem solving purposes NOTE If you need to enter Local System Diagnostics while the system is in Disruptive mode you must...

Page 506: ...RE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 154 Section 7 7 InSite ExC Disk Usage displays available disks and their utilization in the system see Figure 7 123 Figure 7 123 Disk Usage ...

Page 507: ...TION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 155 IP Configuration displays IP information required by the FE or OLC see Figure 7 124 Figure 7 124 IP Configuration ...

Page 508: ... Section 7 7 InSite ExC Network Status displays information concerning the system s communication with the network see Figure 7 125 Windows Services information concerning Windows applications invoked by the system see Figure 7 126 Figure 7 125 Network Status Figure 7 126 Windows Services ...

Page 509: ...ter 7 Diagnostics Troubleshooting 7 157 User Accounts states those users registered on the system see Figure 7 127 Shared Resources indicates those system resources accessible from the network see Shared Resources section page 7 157 Figure 7 127 User Accounts Figure 7 128 Shared Resources ...

Page 510: ... 1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 158 Section 7 7 InSite ExC System Shutdown see Figure 7 129 Disk Defragmenter to invoke Windows disk defragmenter utility see Figure 7 130 Figure 7 129 System Shutdown Figure 7 130 Disk Defragmenter ...

Page 511: ...shooting 7 159 Gather Logs Utility accumulates the logs available in the system in D Export from where the OLC can download them see Figure 7 131 Image Viewer Utility enables the user to save images for further viewing by the OLC see Figure 7 131 Figure 7 131 Gather Logs Utility Figure 7 132 Image Viewer Utility ...

Page 512: ...e OLC to compress images for downloading and deleting redundant files from D Export see Figure 7 133 Scanner Documentation Interface enables the system to read a DOC CD if present in a CD DVD device Figure 7 134 Distinct Network Monitor This feature is not supported Figure 7 133 Image Compress and Delete Utility Figure 7 134 Scanner Documentation Interface ...

Page 513: ...VID Q N SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 161 Virtual Console Observation this is started once the Disruptive mode has been invoked See Figure 7 135 Telnet Server Control this feature is not supported Figure 7 135 Virtual Console Observation ...

Page 514: ... pane of the Invoke Software screen Figure 7 136 above 3 From your PC open Grub Challenger MFC see Figure 7 137 4 Type the Challenge string in the Challenge field A Response code is displayed below the Challenge string in the Grub Challenger screen see Figure 7 137 5 Make a note of the displayed Response code Figure 7 136 Invoke Software Reload Figure 7 137 Grub Challenger MFC Challenge String Res...

Page 515: ...eld in the pane on the right refer to Figure 7 136 The following screen Figure 7 138 displays a message stating that the Reload Setup was successful 7 Reboot the system as explained in the User Manual Vivid i n Vivid q Nsystem see Switching On Off Vivid i n Vivid q N User Manual R2424431 100 Rev 1 page 79 The software reload process begins automatically Figure 7 138 Invoke SW Reload Confirmation M...

Page 516: ...ON 1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 164 Section 7 7 InSite ExC 7 7 7 2 Scanner Utilities Resource File Utility see Figure 7 139 7 7 7 3 DICOM Verify Refer to Figure 7 140 Figure 7 139 Resource File Utility Figure 7 140 DICOM Verify ...

Page 517: ...t replacement details and ordering information is displayed as described in Chapter 9 Renewal Parts Figure 7 141 Replacement Page Note This page is not populated in this version 7 7 9 PM Page When the PM page is selected as shown below information about planned proactive and preventive maintenance is displayed as described in Chapter 10 Figure 7 142 PM Page Note This page is not populated in this ...

Page 518: ... prompt line as shown in the example in Figure 7 143 The bookmark will allow users to place a time stamp where a particular problem has occurred while continuing with their current activity with minimum interruption Note The log file bookmark will later make it easy for service engineers to locate the location of the failure logging more precisely 7 8 2 Extracting Logs Whenever log extraction is n...

Page 519: ... drop down menu select the required export destination DVD CD or Memory Stick then click Save and Export A Zip file named logfile_ date _ time zip is copied onto the selected removable media Note Previously users were required to save log files before exporting them With version BT06 these two operations are performed as one step 5 When done click the Exit button The System Problem Reporting dialo...

Page 520: ...External position as shown in Figure 7 145 below 2 Insert the blank DVD CD media into the DVD CD RW drive tray 3 On the Vivid i n Vivid q N system press the Config key 4 Select the Connectivity Tab and activate the Tools Tab 5 Under Removable Media select the media type CD DVD Writeable from the Media drop down menu as shown in Figure 7 146 6 Click the Format button Figure 7 145 Power Switch in Ex...

Page 521: ...vides recommendations for activating a specific Extensive Log category Table 7 4 Extensive Log Categories and Failure Scenarios Extensive Log Category Type of Failure Archive Loss of patient information or images System lock up due to database corruption Display of message warning that the archive is corrupted Any suspicion that there may be a problem with the archive Connectivity Network is not s...

Page 522: ... the appropriate checkboxes to activate the categories for which detailed analysis is required refer to the recommendations in Table 7 4 Note The impact of activating a checkbox would be to generate larger detailed log files After usage make sure to return to the default settings unflagged to conserve the size of the log files After 7 days the system will automatically return to the default settin...

Page 523: ...tting the Log File Ranges Note By default the log files may be very large as they are not limited by time date The user is able to limit the size timeframe of the log files by choosing the required settings as described below 1 Activate the Export Logs Using checkbox as shown in Figure 7 148 above 2 Proceed to set the required Log Range as follows Either Activate the Time Range radio button then u...

Page 524: ... press Ctrl E to eject the media Table 7 5 Attachment Categories and Recommendations for Use Category Type of Information and Recommended Use Diagnostic Reports Should be used whenever a diagnostic test has failed and it is necessary to forward the information to GE Medical Systems for analysis Crash and Memory Dumps Useful for any type of memory crash or system lock up problem Keyboard Strokes Lo...

Page 525: ...e GE ftp site NOTE Successful upload of logs to the GE ftp site can only be achieved when the designated PC is connected to the Internet and when all security measures such as firewalls taken at the local network do not prevent the utility from performing the upload To Perform the Upload Procedure 1 Place the media in any Windows based PC and run the UploadLog exe The Upload Log dialog is displaye...

Page 526: ... 6 Inform the local GE personnel that you have uploaded files to the GE ftp site Note In the event that the upload process fails check your local network security settings to make sure that they did not prevent the upload Otherwise contact GE personnel 7 8 2 4 DB Scan To Perform the Database Scan Procedure 1 Press ALT D The System Problem Reporting dialog opens Figure 7 151 Upload Complete Figure ...

Page 527: ...t During this procedure a log file which contains information about the integrity of the database will automatically be created and stored in the Logs directory in the folder name Log If required this log file may be sent as an attachment when exporting log files For more details see Setting the Log File Attachments on page 7 172 7 8 2 5 Other Logs Other available logs can be viewed via the Common...

Page 528: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 7 176 Section 7 8 Automatic Error Log This page was intentionally left blank ...

Page 529: ...8 3 Battery Replacement Procedure 8 5 Hard Disk Replacement Procedure 8 7 Control Panel and Keyboard Replacement Procedure 8 9 Rear Cover Latch Assembly Replacement Procedure 8 19 LCD Display Cover Hinges Replacement Procedure 8 21 Bottom Assembly Replacement Procedure 8 25 LCD Display Frame Replacement Procedure 8 27 8 3 Internal Component Replacement Procedures 8 29 P M Probe and MUX Board Repla...

Page 530: ...ERVICE MANUAL 8 2 LCD Flex Cable Replacement Procedure 8 58 8 4 SafeLock Cart Components Replacement 8 61 8 5 Software Loading 8 97 8 6 Performing a Complete System Back up 8 141 8 7 Peripherals 8 148 Table 8 1 Contents in Chapter 8 cont d Section Description Page Number ...

Page 531: ... handle 8 2 1 1 Tools Use the appropriate flat and Phillips type screw drivers as indicated in the Bearing Handle replacement procedure 8 2 1 2 Time Required 10 minutes 8 2 1 3 Preparations None 8 2 1 4 Bearing Handle Removal Procedure 1 Close the Vivid i n Vivid q N LCD cover making sure both front latches are securely fastened Figure 8 1 2 Release the captive screw on either side of the Bearing ...

Page 532: ...ing Handle in position using both hands to pull the Bearing Handle slightly apart in order to slot the ends into the handle holes on the Vivid i n casing 2 Carefully fasten each captive screw on either side of the Bearing Handle to secure the handle in position by turning 1 4 turn in a clockwise direction Note When tightening the captive screws do not use excessive force do not apply more than 1 4...

Page 533: ...ultrasound scanner 2 Close the Vivid i n Vivid q N LCD cover making sure both front latches are securely fastened 3 Carefully turn the Vivid i n Vivid q N over placing it face down 4 While holding the Vivid i n Vivid q N steady pull the locking latch away from the battery and slide the battery sideways out of the battery compartment to remove it completely from the system NOTE WHEN REMOVING A DEFE...

Page 534: ...vid i n Vivid q N lying face down refer to Figure 8 5 on page 8 5 hold the unit steady with one hand while sliding the battery sideways into the battery compartment 2 Make sure the battery is properly inserted all the way before pushing the locking latch to the closed position 3 Turn the Vivid i n Vivid q N over release the front latches and open the LCD cover 4 Reconnect the AC adapter and all ca...

Page 535: ...8 2 3 3 Preparation Shut down the Vivid i n Vivid q N ultrasound unit as described in System Setup 8 2 3 4 Hard Disk Removal Procedure 1 Make sure the Vivid i n Vivid q N LCD Display cover is securely closed and that the scanner is standing firmly on a flat surface 2 Working from the rear of the scanner rear panel use a small Phillips screwdriver to loosen and remove the two screws that secure the...

Page 536: ...hard disk module holder see Figure 8 7 on page 8 7 2 Return and fasten the two screws previously removed to secure the Hard Disk in position refer to Figure 8 6 on page 8 7 Note When a new hard disk arrives from the manufacturer usually it is partitioned and ready for software installation Refer to Software Loading on page 8 97 for software installation instructions Note IMPORTANT if both the Hard...

Page 537: ...ard Removal Procedure 1 Remove the bearing handle as described in the Bearing Handle Removal Procedure on page 8 3 2 Place the Vivid i n Vivid q N portable ultrasound scanner face down and remove the battery as described in the Battery Removal Procedure on page 8 5 3 Loosen and remove each of the 9 screws that fasten the Control Panel and Keyboard onto the Bottom Assembly as indicated in Figure 8 ...

Page 538: ...14 NOTE When removing the Control Panel and Keyboard during other replacement procedures Step 6 will only be required if indicated 6 Lift the Control Panel and Keyboard up to gain access to the cable connector underneath and disconnect the control panel keyboard cable from the system as shown in Figure 8 15 Figure 8 14 Removing the Control Panel and Keyboard Figure 8 15 Disconnecting the Control P...

Page 539: ... in the Bottom Assembly 3 Close the Vivid i n Vivid q N LCD Display cover making sure both front latches are securely fastened 4 Carefully turn the Vivid i n Vivid q Nover placing it face down 5 Return and fasten each of the 9 screws that secure the Control Panel and Keyboard onto the Bottom Assembly refer to Figure 8 13 above 6 Re install the battery as described in the Battery Installation Proce...

Page 540: ...river between the keycap you wish to remove and its neighboring keycap 2 Use the screw driver to gently lever off the defective keycap until it is completely loosened from its base see Figure 8 16 below 3 Lift off and remove the defective keycap 8 2 4 5 2 Keycap Installation Procedure 1 Carefully place the appropriate new keycap in position on the keyboard taking care to place the plastic alignmen...

Page 541: ... of the bracket that fasten the trackball assembly to the base of the Control Panel as shown in Figure 8 17 below 3 Disconnect the three trackball cable connectors from the trackball assembly carefully pull them out and remove the trackball assembly from the Control Panel 8 2 4 5 4 Trackball Installation Procedure 1 Return a new trackball assembly to the correct location on the Control Panel 2 Car...

Page 542: ...and remove the defective keycap see Figure 8 18 below 8 2 4 5 6 ATGC Slider Keycap Installation Procedure 1 Carefully place a new slider keycap in position on the slider base taking care to place the plastic alignment pin in the correct location 2 Gently push the slider keycap down until it snaps into position 8 2 4 5 7 ATGC Rotary Knob Removal Procedure NOTE This section describes the replacement...

Page 543: ...e the defective knob see Figure 8 19 below 8 2 4 5 8 ATGC Rotary Knob Installation Procedure 1 Carefully place a new ATGC Rotary Knob in position on the keyboard taking care to place the plastic alignment pin in the correct location 2 Gently push the ATGC Rotary Knob down until it snaps into position Figure 8 19 ATGC Rotary Knob Replacement ...

Page 544: ...s res as shown below NOTE In the event that the traces res file is not deleted prior to installing the new ECG Respiratory board the following error message will later be displayed 2 Shut down the Vivid i n Vivid q N ultrasound unit as described in System Setup Removal Procedure 1 Lift the Control Panel and Keyboard up to gain access to the cable connector underneath and disconnect the keyboard co...

Page 545: ...t location on the base of the Control Panel 3 Re connect the ECG cable see Figure 8 21 above 4 Re connect the Flex Cable previously removed from the ECG Board to the new ECG Respiratory Board 5 Return and fasten the four screws two long and two short on either side of the bracket to secure the ECG Respiratory Module assembly to the base of the Control Panel refer to Figure 8 20 on page 8 16 When p...

Page 546: ...as shown in Figure 8 22 11 Proceed to perform the following functionality tests Isolated Patient Lead Source Leakage Lead to Ground on page 10 27 Isolated Patient Lead Source Leakage Lead to Lead on page 10 28 Isolated Patient Lead Sink Leakage Isolation Test on page 10 28 SafeLock Cart Grounding Continuity on page 10 23 only if the Vivid i n Vivid q N system is currently mounted on the SafeLock c...

Page 547: ...and Keyboard Removal Procedure on page 8 9 4 Remove the LCD Display Frame as described in the LCD Display Frame Removal Procedure on page 8 27 5 Remove the LCD Display see the LCD Display Removal Procedure on page 8 53 6 Loosen and remove the two screws on each of the left and right hinges attached to the Rear Cover and Latch Assembly as shown in Figure 8 23 below 7 Lift the Rear Cover and Latch A...

Page 548: ...6 Return the Bearing Handle as described in the Bearing Handle Installation Procedure on page 8 4 7 Reconnect the AC adapter and all cables to the Vivid i n Vivid q N portable ultrasound scanner 8 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 only if the Vivid i n Vivid q N system is currently mounted on the SafeLock cart SafeLock Cart Chassi...

Page 549: ...inutes 8 2 6 3 Preparations Shut down the Vivid i n Vivid q N ultrasound unit as described in System Setup 8 2 6 4 LCD Display Cover Hinges Removal Procedure 1 Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 2 Disconnect the LCD Flex ribbon cable from the BEP as shown in Figure 8 24 Use a flat screwdriver to carefully release the c...

Page 550: ...en and remove the 2 screws that secure the left hinge to the LCD Display cover Figure 8 26 then carefully remove the hinge 6 Loosen and remove the 2 screws that secure the right hinge to the LCD Display cover Figure 8 26 then carefully remove the hinge Figure 8 25 LCD Display Hinges Secured to Base Assembly Location of Screws Figure 8 26 Hinges Secured to LCD Display Cover Location of Screws Figur...

Page 551: ...cover by raising and lowering it while exerting pressure to slide the LCD Display Cover until the hinge holes are aligned with the Bottom Assembly screw holes see right hinge in Figure 8 29 on page 8 24 NOTE The axle hinges should be adjusted so that on each hinge the axle should protrude equally as shown in Figure 8 28 below 4 On the Left hinge return and fasten the 2 screws previously removed to...

Page 552: ...ge 8 22 Do not over tighten the screws 9 Reconnect the LCD flex cable to the BEP refer to Figure 8 24 on page 8 21 10 Refit the Control Panel and Keyboard as described in the Control Panel and Keyboard Installation Procedure on page 8 11 11 Close and open the lid a few times to ensure proper closure of the LCD Display cover 12 Reconnect the AC adapter and all cables to the Vivid i n Vivid q N port...

Page 553: ...the HVPS as described in the HVPS Removal Procedure on page 8 46 9 Remove the fans as described in the Fan Removal Procedure on page 8 49 10 Remove the LCD Display Frame as described in the LCD Display Frame Removal Procedure on page 8 27 11 Separate the Bottom Assembly from the LCD Display rear cover by loosening and removing the two screws on each of the left and right hinges attached to the Rea...

Page 554: ... 31 10 Return the Control Panel and Keyboard as described in the Control Panel and Keyboard Installation Procedure on page 8 11 11 Return the hard disk as described in the Hard Disk Installation Procedure on page 8 8 12 Return the bearing handle as described in the Bearing Handle Installation Procedure on page 8 4 13 Re install the battery as described in the Battery Installation Procedure on page...

Page 555: ...ivid i n Vivid q N ultrasound unit as described in System Setup 8 2 8 4 LCD Display Frame Removal Procedure 1 With the LCD Display in the open upright position use a flat screwdriver to remove the screw cap from each of the four screws that fasten the LCD display frame in position as shown in Figure 8 30 below Figure 8 30 Removing the Screw Caps from the LCD Display Frame 2 Loosen and remove the f...

Page 556: ...ion return the LCD display frame to the correct position and using both hands press the frame under the securing clips on all four sides to fasten it against the LCD display Refer to Figure 8 32 on page 8 28 2 Return and tighten the four screws previously removed 3 With the LCD Display in the open upright position return the screw cap to each of the four screws that fasten the LCD frame in positio...

Page 557: ...en tightening screws do not overtighten or use excessive force CAUTION DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS 1 ALWAYS CONNECT YOURSELF VIA AN ARM WRIST STRAP TO THE ADVISED ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER TO THE RIGHT OF THE POWER CONNECTOR 2 FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT ...

Page 558: ...cedure on page 8 3 2 Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Loosen and remove the four screws that secure the P M Board to the RFI Board as shown in Figure 8 33 4 Loosen the two screws that fasten the P M Board to the Bottom Assembly as shown in Figure 8 34 below Figure 8 33 Removing the Screws that Secure the P M Board ...

Page 559: ... into position on the Bottom Assembly Make sure it is properly seated before returning and fastening the two screws previously removed Refer to Figure 8 35 and Figure 8 34 above 2 Return and fasten the four screws to secure the P M Board to the RFI Board as shown in Figure 8 33 on page 8 30 3 Return the Control Panel and Keyboard as described in the Control Panel and Keyboard Installation Procedur...

Page 560: ... the board as shown in Figure 8 36 below CAUTION When removing the boards ESD may cause damage to a board Always have the ESD hand strip connected to the machine chassis and to your hand 8 3 2 1 Tools Use the appropriate Phillips and flat screw drivers as indicated in the bottom assembly replacement procedures 8 3 2 2 Time Required 15 minutes 8 3 2 3 Preparations Shut down the Vivid i n Vivid q N ...

Page 561: ...e the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Loosen the eight screws that secure the TR32 and RFI Boards to the Bottom Assembly as shown in Figure 8 37 below 4 Remove the eight screws see Figure 8 38 5 Disconnect the PCI cable flex BEP to RFI from the cable connector on RFI Board as shown in Figure 8 37 Loosen Screws that Secure TR...

Page 562: ...n DC Power Supply as shown in Figure 8 40 below 7 Gently lift one side of the P M Board upwards and hinge it open to gain access to the RFI and TR32 Board assembly beneath as shown in Figure 8 42 below The RFI and TR32 Boards are seated together in the system as one unit Figure 8 39 Disconnecting the PCI Cable from the RFI Board Figure 8 40 Disconnecting the DC Power Supply Figure 8 41 Accessing t...

Page 563: ...ard assembly upwards and remove it from the Bottom Assembly as shown in Figure 8 42 below 9 Remove the RFI Board from the TR32 Boards separate by gently pulling them apart as shown in Figure 8 43 below Figure 8 42 Removing the RFI and TR32 Boards from the Bottom Assembly Figure 8 43 Removing the RFI Board from the TR32 Boards TR32 Boards RFI Board ...

Page 564: ...ernal Component Replacement Procedures 10 Separate the two TR32 Boards from each other as shown in Figure 8 44 below Note Replace the defective board as described in the appropriate steps of the following procedure Figure 8 44 Separating the Two TR32 Boards from Each Other RFI Board TR32 Board TR32 Board ...

Page 565: ...sembly While performing this step gently lift one side of the P M Board upwards and hinge it open in order to insert the TR32 and RFI Board assembly underneath Refer to Figure 8 41 and Figure 8 42 on page 8 35 4 Reconnect each of the two HVPS cables to the DC Power Supply see Figure 8 40 on page 8 34 5 Reconnect the PCI cable flex BEP to RFI to the cable connector on RFI Board see Figure 8 39 on p...

Page 566: ...edure on page 8 3 3 Make sure the Vivid i n Vivid q N LCD Display cover is securely closed and place the scanner face down Remove the battery as described in the Battery Removal Procedure on page 8 5 4 Remove the Hard Disk as described in the Hard Disk Removal Procedure on page 8 7 5 Loosen and remove the two securing nuts one on each side of the VGA Connector located on the Vivid i n Vivid q N re...

Page 567: ...he BEP to RFI Flex cable c Disconnect the HVPS to BEP power cable d Disconnect the Fan Speaker to BEP power cable 8 Unscrew and remove the four screws that secure the BEP to the Bottom Assembly as shown in Figure 8 46 Disconnecting Cables from the BEP LCD Flex cable BEP to RFI Flex cable HVPS to BEP Power cable Fan Speaker to BEP cable ...

Page 568: ... tab illustrated in Figure 8 48 that serves to seat the BEP firmly in position It is necessary to loosen and release this tab lift it up using a screw driver if necessary in order to facilitate complete removal of the BEP 9 Using both hands lift the BEP out of the Bottom Assembly and remove it from the system as shown Figure 8 47 Removing the Screws from the BEP Figure 8 48 Releasing the Plastic S...

Page 569: ...low NOTE When removing the BEP for replacement with the SATA BEP disconnect the control panel keyboard cable Figure 8 50 from the parallel BEP you have just removed and keep this cable for installing on the SATA BEP Figure 8 49 Removing the BEP from the Bottom Assembly Figure 8 50 Control Panel Keyboard Cable Control Panel Keyboard Cable ...

Page 570: ...ation The bracket is shown on the left in Figure 8 51 in the picture on the right the bracket is shown in position in the Vivid i n Vivid q N scanner Notice that the bracket sits on the PCMCIA slot flap which is pressed open while the adaptor bracket is positioned on the flap itself To install the adaptor bracket the flap is pressed open Figure 8 52 and the adaptor bracket is placed in position Th...

Page 571: ... to fit the access codes already registered in the Hard Disk Only qualified personnel should perform these tasks 2 Check that the BEP is seated correctly then return and fasten the four screws previously removed to secure the BEP to the Bottom Assembly refer to Figure 8 47 on page 8 40 3 Re connect all the cables to the BEP Proceed as follows refer to Figure 8 46 on page 8 39 a Re connect the Spea...

Page 572: ...on page 8 118 Note If both the Back End and the Hard Disk are being replaced simultaneously continue to the next step If only the Back End is being replaced skip step 12 and proceed to step 13 below 12 Perform the steps required to burn the LCD type onto the newly installed Back End as described in step 9 on page 8 56 through step 12 on page 8 57 13 Perform a complete functional check of the syste...

Page 573: ...tup 8 3 4 4 BIOS Battery Removal Procedure 1 Remove the BEP as described in the BEP Removal Procedure on page 8 38 2 Take out the BIOS battery located on the bottom of the BEP as shown in Figure 8 54 8 3 4 5 BIOS Battery Installation Procedure NOTE Before proceeding refer to the Preparation Procedure on page 8 43 1 Fit a new BIOS battery in position on the BEP refer to Figure 8 54 2 Re install the...

Page 574: ... HVPS Removal Procedure 1 Make sure the Vivid i n Vivid q N LCD Display cover is securely closed and place the scanner face down Remove the battery as described in the Battery Removal Procedure on page 8 5 2 Remove the bearing handle as described in the Bearing Handle Removal Procedure on page 8 3 3 Disconnect each of the two Power Supply cables from the DC Power Supply HVPS to BEP and HVPS to RFI...

Page 575: ... Bottom Assembly as shown in Figure 8 57 below 6 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 only if the Vivid i n Vivid q N system is currently mounted on the SafeLock cart SafeLock Cart Chassis Current Leakage Test on page 10 25 Calibration Tests on page 7 25 Full System Test on page 7 10 Figure 8 57 Removing the DC Power Supply ...

Page 576: ... 8 46 3 Reconnect each of the two Power Supply cables to the DC Power Supply HVPS to BEP and HVPS to RFI as shown in Figure 8 55 on page 8 46 4 Return the bearing handle as described in the Bearing Handle Installation Procedure on page 8 4 5 Make sure the Vivid i n Vivid q N LCD Display cover is securely closed and place the scanner face down Re install the battery as described in the Battery Inst...

Page 577: ...nutes 8 3 6 3 Preparation Shut down the Vivid i n Vivid q N ultrasound unit as described in System Setup 8 3 6 4 Fan Removal Procedure 1 Remove the bearing handle as described in the Bearing Handle Removal Procedure on page 8 3 2 Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Unscrew and remove the two screws that secure the Fan...

Page 578: ... upwards from the front and then the rear to remove it from the Bottom Assembly lay it down adjacent to the system ready for the next step as shown in Figure 8 60 below Figure 8 60 Removing the Fan Assembly 6 Disconnect the Fan cable from the BEP and completely remove the Fan Assembly from the system as shown in Figure 8 61 below Figure 8 61 Disconnecting the Fan Cable from the BEP ...

Page 579: ...Fan Assembly is seated in the correct position Return and fasten the two screws previously removed to secure the Fan Assembly to the Bottom Assembly as shown in Figure 8 58 on page 8 49 4 Reconnect the speaker cable to the fan assembly as shown in Figure 8 59 5 Return the Control Panel and Keyboard as described in the Control Panel and Keyboard Installation Procedure on page 8 11 6 Return the bear...

Page 580: ...ibed in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Loosen and remove the two screws on each of the speaker housings four in total 4 Disconnect the speaker cables from the fan assembly 5 Gently lift the speaker assembly towards you and remove from the system 8 3 7 5 Speaker Assembly Installation Procedure 1 Return a replacement speaker assembly to the original position ensuring ...

Page 581: ...e as described in the Bearing Handle Removal Procedure on page 8 3 2 Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Remove the LCD Display Frame as described in the LCD Display Frame Removal Procedure on page 8 27 4 Loosen and remove the four screws that fasten the LCD Display to the Rear Cover Latch Assembly as shown in shown i...

Page 582: ... 8 3 Internal Component Replacement Procedures 6 Using both hands carefully lift the LCD Display upwards and remove it from the Rear Cover Latch Assembly as shown in Figure 8 64 below Take care not to damage the LCD Flex ribbon cable when performing this step Figure 8 64 Removing the LCD Display ...

Page 583: ...when performing this step 2 Use both hands to re connect the LCD Flex ribbon cable to the BEP refer to Figure 8 63 on page 8 53 3 Return and fasten the four screws to secure the LCD Display to the Rear Cover Latch Assembly refer to Figure 8 62 on page 8 53 4 Return the LCD Display Frame as described in the LCD Display Frame Installation Procedure on page 8 28 5 Check that the LCD Display can be ti...

Page 584: ... 3 Internal Component Replacement Procedures 9 From Windows Explorer go to the directory C Marlin Target bin see Figure 8 65 below 10 Double click on SetLCDType exe to run this utility The Set LCD Type dialog opens Figure 8 66 Figure 8 65 Selecting SetLCDType Utility Figure 8 66 Set LCD Type Dialog ...

Page 585: ...tion procedure proceed to calibrate the newly installed LCD Display by making any adjustments required to the Brightness Contrast and or Blue Tint as described in the LCD Display Adjustments and Calibration on page 6 2 14 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 only if the Vivid i n Vivid q N system is currently mounted on the SafeLock ...

Page 586: ...Flex Cable Removal Procedure 1 Disconnect the AC cable and battery 2 Remove the Control Panel and Keyboard as described in the Control Panel and Keyboard Removal Procedure on page 8 9 3 Remove the LCD Display Frame as described in the LCD Display Frame Removal Procedure on page 8 27 4 Remove the LCD Display as described in the LCD Display Removal Procedure on page 8 53 5 Using both hands carefully...

Page 587: ...anel and disconnect the flex cable see Figure 8 69 Figure 8 69 Removing Tape from LCD Rear Panel 2 Connect the new flex cable to LCD rear panel connector as shown in Figure 8 70 3 Connect the flex cable to the LCD rear panel connector 4 Adhere the kapton tape on the LCD rear panel connector see Figure 8 71 Figure 8 70 Connecting the Flex Cable to the LCD Rear Panel ...

Page 588: ...ame as described in the LCD Display Frame Installation Procedure on page 8 28 9 Check that the LCD Display can be tilted upwards and downwards and that it holds its set position 10 Return the Control Panel and Keyboard as described in the Control Panel and Keyboard Installation Procedure on page 8 11 11 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page...

Page 589: ...pment must not be operated within 15 feet of any magnetic equipment DANGER Electrical hazards exist at several points in the system Familiarize yourself with all hazardous voltages and high current levels before removing any of the covers DANGER DO NOT wear the ESD wrist band strap when removing parts from the power supply unit Before removing any part of the power unit turn the power OFF and disc...

Page 590: ...Lock Cart Components Label Item Label Item 1 Top Shelf Assembly with handle 8 Rear Lower Cover 2 ECG Cable Protector 9 Bottom Cover Front 3 Peripheral Power Outlet 10 Bottom Cover Rear AC Assembly Cover 4 LAN Isolation Box 11 AC Power Cable Bumper Protector 5 AC Cable Hook 12 AC Distribution Assembly behind cover 6 Probe Shelf with probe holders 13 SafeLock Cart Bottom Assembly 7 Rear Upper Cover ...

Page 591: ... Cable Hook see page 8 65 AC Power Cable Bumper Protector see page 8 66 ECG Cable Protector see page 8 68 Lower Cover Rear see page 8 71 AC Distribution Assembly see page 8 72 Peripheral Power Cable see page 8 74 Main Cable Harness see page 8 79 USB Board see page 8 80 Gas Spring see page 8 82 DVD Holder see page 8 85 LAN Isolation Box Holder see page 8 86 Top Shelf Assembly see page 8 87 Handrest...

Page 592: ...ver and while supporting the cover with one hand carefully remove the cover see Figure 8 73 8 4 3 4 Upper Cover Rear Installation Procedure 1 Return the cover to the correct position on the rear of the SafeLock Cart see Figure 8 73 2 While holding the cover to prevent it from falling return and fasten the two upper screws see Figure 8 73 3 Fasten the four remaining screws that secure the upper cov...

Page 593: ...om the Upper Cover rear unscrew the four AC Cable Hook retaining screws see Figure 8 74 and remove the cable hook holder 8 4 4 3 AC Cable Hook Installation Procedure 1 Position the AC Cable Hook on the Upper Cover rear and return and fasten the four retaining screws 2 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 SafeLock Cart Chassis Current...

Page 594: ...ector upwards to release it from the shoulder screws refer to Figure 8 77 and Figure 8 75 2 Remove the Bumper Protector from the Cart 8 4 5 3 AC Power Cable Bumper Protector Installation Procedure NOTE If this is a first time installation of the bumper protector it is necessary to remove the two existing Phillips screws located on the rear of the SafeLock Cart AC Cable connection area as shown in ...

Page 595: ...Protector onto the shoulder screws by aligning the mounting tabs ears with the screws then pushing the bumper downwards until it is seated securely See Figure 8 77 4 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 SafeLock Cart Chassis Current Leakage Test on page 10 25 Figure 8 77 AC Power Cable Bumper Protector Mounted on SafeLock Cart AC Pow...

Page 596: ...otector Installation Procedure If this is a first time installation of the ECG Cable Protector it is necessary to remove the two plastic guide holders from the Top Shelf Assembly as shown in Figure 8 78 While facing the SafeLock Cart i e working from the front of the Cart these are located on the right side of the SafeLock Cart 1 Thoroughly clean the surface of the Top Shelf Assembly in particular...

Page 597: ...n the rear securing screw then the front securing screw 3 Return and fasten the rear securing screw Figure 8 80 then the front securing screw Figure 8 81 illustrates the ECG Cable Protector secured in position on the SafeLock Cart Figure 8 79 Placing the ECG Cable Protector in Position CAUTION When placing the ECG Cable Protector in position take care not to damage the plastic Figure 8 80 Fastenin...

Page 598: ...nt 4 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 SafeLock Cart Chassis Current Leakage Test on page 10 25 Figure 8 81 ECG Cable Protector Secured on SafeLock Cart ECG Cable Protector Mounted on SafeLock Cart ECG Cable Protector as seen when in Use ECG Cable Connected to Scanner ...

Page 599: ...g screws of the lower cover rear and while supporting the cover with one hand gently remove the cover see Figure 8 82 8 4 7 4 Lower Cover Rear Installation Procedure 1 Return the cover to the correct position on the rear of the SafeLock Cart see Figure 8 82 2 While holding the cover to prevent it from falling fasten the retaining screws of the lower cover rear 3 Proceed to perform the following fu...

Page 600: ... of the cart rear 2 Remove the Upper Cover Rear as described in the Upper Cover Rear Removal Procedure on page 8 64 3 Remove the Lower Cover Rear as described in the Lower Cover Rear Removal Procedure on page 8 71 4 Remove the three retaining screws of the AC Assembly Cover and remove the cover Figure 8 83 below The AC Distribution Assembly is now exposed as shown in Figure 8 84 5 Cut all cable ti...

Page 601: ...Fasten the Ground screw to the AC Distribution Assembly Figure 8 84 above 3 Reconnect the peripheral power cable to the AC Distribution Assembly Figure 8 84 above 4 Secure all cables in position with new cable ties 5 Return the AC Assembly cover and fasten the three screws previously removed 6 Return the Lower Cover Rear as described in the Lower Cover Rear Installation Procedure on page 8 71 7 Re...

Page 602: ...ar of the SafeLock Cart as described in Upper Cover Rear Removal Procedure on page 8 64 and Lower Cover Rear Removal Procedure on page 8 71 8 4 9 3 Peripheral Power Cable Removal Procedure 1 Disconnect the Peripheral Power cable from the AC Distribution Assembly 2 Cut any cable ties that secure the cable 3 Disconnect all grounding connections 4 Remove the Peripheral Power socket by squeezing the p...

Page 603: ...cable and power socket in position in the SafeLock s Upper Cover front see item 4 in Figure 8 72 page 8 62 2 With the AC power socket plugged into the cart Figure 8 86 left picture connect the ground cable to the body of the cart 3 Route the Peripheral AC Outlet cable as shown in the middle picture of Figure 8 86 4 Secure the cable with the cable clamp in the direction shown in Figure 8 87 5 Route...

Page 604: ...d downwards Figure 8 88 Figure 8 88 Routing of Cable Downwards inside Rear Upper Cover 7 Fasten the Peripheral AC Outlet cable to the Main Harness cable as shown in Figure 8 89 Figure 8 89 Peripheral AC Outlet Cable Fastened to Main Harness Cable 8 Plug the Peripheral AC Outlet cable into the power supply and connect the Ground lead to the body of the cart Figure 8 90 Note use of cable fasteners a...

Page 605: ...ng it into a u shape and fasten together with tie wraps as shown in Figure 8 91 11 Check that the routing procedure has been implemented correctly lower the cart s top shelf and make sure that the cables do not interfere with the movement of the cart s top shelf Figure 8 90 Peripheral AC Outlet and Ground Cable Connected Figure 8 91 USB Power Cable Folded and Fastened with Tie wraps ...

Page 606: ...ision 3 Chapter 8 section 8 4 3 4 13 Return the Upper Cover rear and Lower Cover rear of the SafeLock Cart as described in Upper Cover Rear Installation Procedure on page 8 64 and Lower Cover Rear Installation Procedure on page 8 71 14 Once more slide the top shelf up and down a number of times and check that the movement is smooth and that there is no interference of the cables with the up down m...

Page 607: ...ard as shown in Figure 8 93 4 From above the Top Shelf carefully push the Main Cable Harness Vivid i n Vivid q N power supply cable through the top shelf aperture and gently pull it through from below 5 From below the Top Shelf carefully push the DVD connectors through the front upper cover 6 Remove the Main Cable Harness from the cart 8 4 10 4 Main Cable Harness Installation Procedure 1 From the ...

Page 608: ...Upper Cover Rear Removal Procedure on page 8 64 3 Disconnect the P6 cable and USB connector from the USB board as shown in Figure 8 93 4 Remove the upper and lower bolts that attach the USB board to the upper frame as shown in Figure 8 93 5 Remove the USB board from the SafeLock Cart 8 4 11 4 USB Board Installation Procedure 1 Return the replacement USB board to its position on the SafeLock Cart c...

Page 609: ...nt Procedures 8 81 5 Reconnect the AC power cable to the AC input assembly at the base of the cart rear and turn ON power to the SafeLock Cart 6 Proceed to perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 SafeLock Cart Chassis Current Leakage Test on page 10 25 ...

Page 610: ...ure on page 8 64 3 Remove the Lower Cover Rear as described in the Lower Cover Rear Removal Procedure on page 8 71 4 Disconnect the Main Harness cable from the AC unit cut both tie clips and free the Main Cable Harness so that it does not interfere with the Gas Spring 5 Gradually raise the Upper Assembly until the lower Gas Spring securing screw is accessible 6 Lie the SafeLock cart on its left si...

Page 611: ...e Gas Spring locking nut counter clockwise until it rises about a quarter way upwards on the thread 3 Insert the Gas Spring handle into the Gas Spring Housing Assembly until it is flush with the outer cylindrical surface of the housing assembly 4 Attach the Gas Spring to the housing assembly by turning it clockwise until the locking nut makes contact with the housing assembly see Figure 8 96 Figur...

Page 612: ...around the Gas Spring and then reconnect it to the AC unit 8 Secure all cables in position with new cable ties 9 Return the Lower Cover Rear as described in the Lower Cover Rear Installation Procedure on page 8 71 10 Return the Upper Cover Rear as described in the Upper Cover Rear Installation Procedure on page 8 64 11 Proceed to perform the following functionality tests SafeLock Cart Grounding Co...

Page 613: ...g the DVD Holder with one hand remove the four retaining screws as shown in Figure 8 98 2 Remove the DVD Holder from the base of the Top Shelf 8 4 13 3 DVD Holder Installation Procedure 1 Working from beneath the Top Shelf align the replacement DVD Holder in the correct position 2 Return and fasten the four retaining screws as shown in Figure 8 98 to secure the DVD Holder firmly in position Figure...

Page 614: ...ove the two LAN Holder retaining screws see Figure 8 99 and remove the LAN Holder 3 Remove the LAN Isolation Box 8 4 14 3 LAN Isolation Box and Holder Installation Procedure 1 Position the LAN Isolation Box in the LAN Holder and replace the LAN Holder with the Isolation Box in their appropriate location beneath the Top Shelf 2 Return and tighten the two retaining screws to secure the LAN Holder 3 ...

Page 615: ...e 8 64 4 Remove the Lower Cover Rear as described in the Upper Cover Rear Removal Procedure on page 8 64 5 Remove the USB board as described in the USB Board Removal Procedure on page 8 80 6 Remove the gas spring as described in Gas Spring Replacement Procedure on page 8 82 7 Disconnect the two cables P1 cable and peripheral power cable from the top of the AC Input Assembly as described in AC Dist...

Page 616: ... 8 82 4 Return the USB Box as described in the USB Board Installation Procedure on page 8 80 5 Fasten all cables in position using new cable ties 6 Return the Lower Cover Rear as described in the Lower Cover Rear Installation Procedure on page 8 71 7 Return the Upper Cover Rear as described in the Upper Cover Rear Installation Procedure on page 8 64 8 Reconnect all required peripherals to the Safe...

Page 617: ...ove all peripherals from the Vivid i n Vivid q N portable ultrasound scanner 2 Working from the front of the SafeLock Cart and below the main shelf loosen and remove the three screws that secure the handrest to the main shelf see Figure 8 102 below 3 Remove the handrest 8 4 16 4 Handrest Installation Procedure 1 Install a replacement handrest on the SafeLock Cart s main shelf making sure the three...

Page 618: ...vid q N ultrasound unit System Setup 8 4 17 3 Probe Shelf Assembly Removal Procedure 1 Remove all peripherals from the Vivid i n Vivid q N portable ultrasound scanner 2 Working from the front of the SafeLock Cart and underneath the probe shelf assembly loosen and remove the two lower screws highlighted that secure the shelf in position Figure 8 103 below 3 While supporting the shelf with one hand ...

Page 619: ...e hand return and fasten the two upper screws previously removed 3 Return and fasten the two lower screws one on either side making sure that the shelf is firmly secured in position 4 Place the required peripheral on the probe shelf assembly 5 Re connect all peripherals to the Vivid i n Vivid q N portable ultrasound scanner 6 Proceed to perform the following functionality tests SafeLock Cart Groun...

Page 620: ...8 2 Probe Cable Hook Removal Procedure 1 From below the Probe Shelf remove the Probe Cable Hook retaining screw see Figure 8 105 on page 8 92 2 Gently pull the Probe Cable Hook upwards to remove it 8 4 18 3 Probe Cable Hook Installation Procedure 1 Insert the Probe Cable Hook in its position on the Probe Shelf 2 From below return and replace the Probe Cable Hook retaining screw Figure 8 105 Probe ...

Page 621: ... i n Vivid q N ultrasound unit as described in System Setup 8 4 19 3 Front Wheel Removal Procedure 1 Disconnect mains power from the AC adapter 2 Disconnect the AC adapter and all cables and peripherals from the Vivid i n Vivid q N portable ultrasound scanner 3 Remove the Vivid i n Vivid q N and all peripherals from the SafeLock Cart 4 Carefully lay the SafeLock Cart on its back 5 Loosen and remov...

Page 622: ... previously removed and make sure that wheel is secured firmly to the chassis 3 Stand the SafeLock Cart upright 4 Return the Vivid i n Vivid q N and all peripherals to their appropriate positions on the SafeLock Cart 5 Re connect the AC adapter and all cables and peripherals to the Vivid i n Vivid q N portable ultrasound scanner 6 Re connect mains power to the AC adapter 7 Proceed to perform the f...

Page 623: ... Removal Procedure 1 Disconnect mains power from the AC adapter 2 Disconnect the AC adapter and all cables and peripherals from the Vivid i n Vivid q N portable ultrasound scanner 3 Remove the Vivid i n Vivid q N and all peripherals from the SafeLock Cart 4 Carefully lay the SafeLock Cart on its side 5 Loosen and remove the four bolts that secure the rear wheel to the SafeLock Cart chassis see Fig...

Page 624: ...ly removed see Figure 8 107 and make sure that wheel is secured firmly to the chassis 3 Stand the SafeLock Cart upright 4 Return the Vivid i n Vivid q N and all peripherals to their appropriate positions on the SafeLock Cart 5 Re connect the AC adapter and all cables and peripherals to the Vivid i n Vivid q N portable ultrasound scanner 6 Re connect mains power to the AC adapter 7 Proceed to perfo...

Page 625: ...ready installed In some cases it is necessary to re install the software for example after replacing the Back end Processor by following the software re installation procedure shown in Table 8 6 on page 8 104 Refer to the following sub sections as appropriate Vivid i n Software Upgrade from Version BT10 to BT11 on page 8 98 Vivid i n Vivid q N Software Re Installation Procedure on page 8 104 Setti...

Page 626: ...riate ordering number HCat as listed in Table 8 3 Table 8 3 Software BT10 to BT11 Upgrade Ordering Number HCat Description H45032BF Vi q German Upgr BT11 H45032BG Vi q Italian Upgr BT11 H45032BH Vi q Spanish Upgr BT11 H45032BJ Vi q Spanish LA Upgr BT11 H45032BK Vi q Portuguese Upgr BT11 H45032BL Vi q Brazilian Upgr BT11 H45032BM Vi q Japanese Upgr BT11 H45032BN Vi q Chinese Upgr BT11 H45032BP Vi q...

Page 627: ...T11 H45032CH Vi q Serbian Upgr BT11 H45032CJ Vi q Korean Upgr BT11 H45032CK Vi q Estonian Upgr BT11 H45032CL Vi q Hebrew Upgr BT11 H45032CM Vi q Croatian Upgr BT11 H45032CN Vi q Slovenian Upgr BT11 H45032CP Vi q English Upgr BT11 H45032CR Vi q French Upgr BT11 Table 8 4 Upgrade Kit P N H45031XX Contentsa a The Vivid i n Vivid q N Release Note is also provided with the Upgrade Kit and in the DOC CD...

Page 628: ...Wireless network settings if present Trackball speed InSite ExC settings NOTE The following upgrade procedure is based on the Vivid q N system The procedure is the same as for the Vivid i n system 8 5 1 1 2 Vivid i n Vivid q N Software Upgrade Procedure BT10 to BT11 Table 8 5 Step by Step Vivid i n Vivid q N Software Upgrade Procedure Step Instruction Expected Result 1 Perform a complete shutdown ...

Page 629: ...password use lower case letters A message is displayed The DVD then begins the software installation process and a progress indicator is displayed as shown below This stage of the process takes approximately 5 mins Note Do not stop the software installation operation during this process 10 Press CTRL ALT DELETE The Ghosting process continues for approximately 10 minutes The system automatically lo...

Page 630: ...e version corresponds with the System SW version you have installed check against the CD label on the installation disc and that it was correctly installed 14 Restore all Presets as per user manual Note When performing Restore click Admin at the bottom of the screen then click the Restore tab Select System Configuration and then de select GSP Resources 15 Install security patch as required 16 Inst...

Page 631: ...tion 8 5 1 1 1 on page 8 100 Verify proper Connectivity functionality refer to Connectivity Setup on page 3 85 18 Proceed to perform the following functionality tests Software Configuration Checks on page 4 37 Calibration Tests on page 7 25 Full System Test on page 7 10 Table 8 5 Step by Step Vivid i n Vivid q N Software Upgrade Procedure cont d Step Instruction Expected Result ...

Page 632: ...the Eject button of the DVD CD RW drive and remove any media if present 3 Insert the CD provided with the software kit into the DVD CD RW The DVD CD RW drive loads the software CD 4 Hold down the ON OFF button for 5 seconds to shutdown the system The shutdown options are displayed on screen 5 Select Full Shutdown The system powers down 6 Hold down the ON OFF button for three seconds The system pow...

Page 633: ...L ALT DELETE The system re boots 11 Follow the instructions on the screen clicking OK or Cancel when prompted The machine performs a number of reboots before completing the installation As the installation proceeds the system displays various installation screens The installation process will take approximately 15 minutes DO NOT ATTEMPT TO STOP THE INSTALLATION BEFORE COMPLETION When done the foll...

Page 634: ... 10 seconds then re connect the battery and the AC adapter Boot up the system by pressing the On Off button If this is a new installation the system will prompt for the software licence Password as shown below Proceed to step 15 If not a new installation there is no prompt for the software licence Password and the boot up screen is displayed see below Proceed to step 16 15 Type the password provid...

Page 635: ...trol panel A message is displayed prompting for system re boot 19 Click OK and re boot the system 20 If this is a new software version an adhesive Rating Plate is supplied with the software CD Adhere the plate in the appropriate position under the Bottom Assembly 21 Proceed to perform the following functionality tests Software Configuration Checks on page 4 37 Calibration Tests on page 7 25 Full S...

Page 636: ...versions 1 While booting press F2 Config Diag on alphanumeric keyboard 2 Type the password marlin04 3 Exit and save presets 4 Select Exit section and load Setup Defaults 5 Select Main section 6 Set correct System Time and System Date Set Date using next format MM DD YYYY 7 Check that IDE Primary Master is detected 8 Proceed to Setting the BIOS Advanced Parameters on page 8 109 Figure 8 108 BIOS Ma...

Page 637: ...he Advanced menu and press Enter Advanced Chipset Control is highlighted BIOS Parameter Corresponding Figure Chipset Control Figure 8 110 on page 8 110 PCI PNP Configuration Figure 8 111 on page 8 110 Memory Cache Figure 8 113 on page 8 111 I O Device Configuration Figure 8 113 on page 8 111 Keyboard Features Figure 8 114 on page 8 112 Hardware Monitor Figure 8 115 on page 8 112 Watchdog Settings ...

Page 638: ...bove in Figure 8 110 4 Press Esc to return to the Advanced menu home screen Figure 8 109 5 Scroll to PCI PNP Configuration and press Enter to open the PCI PNP Configuration screen 6 Set the parameters in the PCI PNP Configuration screen as shown above in Figure 8 111 7 Press Esc to return to the Advanced menu home screen Figure 8 109 Figure 8 110 BIOS Advanced Chipset Control Parameters Figure 8 1...

Page 639: ... in Figure 8 112 10 Press Esc to return to the Advanced menu home screen Figure 8 109 11 Scroll to I O Device Configuration and press Enter to open the I O Device Configuration screen 12 Set the parameters in the I O Device Configuration screen as shown above in Figure 8 113 13 Press Esc to return to the Advanced menu home screen Figure 8 109 Figure 8 112 BIOS Advanced Memory Cache Parameters Figu...

Page 640: ...creen as shown above in Figure 8 114 16 Press Esc to return to the Advanced menu home screen Figure 8 109 17 Scroll to Hardware Monitor and press Enter to open the Hardware Monitor screen 18 Set the parameters in the Hardware Monitor screen as shown above in Figure 8 115 19 Press Esc to return to the Advanced menu home screen Figure 8 109 Figure 8 114 BIOS Advanced Keyboard Features Parameters Fig...

Page 641: ...og Settings as shown above in Figure 8 116 22 Press Esc to return to the Advanced menu home screen Figure 8 109 23 Scroll to Display Control and press Enter to open the Display Control screen 24 Set the parameters in the Display Control screen as shown above in Figure 8 117 25 Press Esc to return to the Advanced menu home screen Figure 8 109 Figure 8 116 BIOS Advanced Watchdog Settings Figure 8 11...

Page 642: ...Section 8 5 Software Loading 26 Scroll to Miscellaneous and press Enter to open the Miscellaneous screen 27 Set the parameters in the Miscellaneous screen as shown above in Figure 8 118 28 Proceed to Additional BIOS Settings on page 8 115 Figure 8 118 BIOS Advanced Miscellaneous Parameters ...

Page 643: ...e shown in this section 1 From the BIOS main menu see Figure 8 108 select the Advanced menu and press Enter 2 Select the Security menu and press Enter 3 Set the parameters in the Security screen as shown above in Figure 8 119 4 Press Esc to return to the main BIOS screen Figure 8 108 on page 8 108 BIOS Parameter Corresponding Figure Security Figure 8 119 on page 8 115 Power Figure 8 120 on page 8 ...

Page 644: ...the Power screen as shown above in Figure 8 120 7 Press Esc to return to the main BIOS screen Figure 8 108 on page 8 108 8 Select the Boot menu and press Enter 9 Set the parameters in the Boot parameters screen as shown above in Figure 8 121 10 Press Esc to return to the main BIOS screen Figure 8 108 on page 8 108 Figure 8 120 BIOS Power Parameters Figure 8 121 BIOS Boot Parameters ...

Page 645: ...ess F10 to Exit Saving Changes as shown in Figure 8 122 12 Reboot the system and check that the Supervisor PASSWORD was installed properly 13 Proceed to perform the following functionality tests Software Configuration Checks on page 4 37 Calibration Tests on page 7 25 Full System Test on page 7 10 Figure 8 122 BIOS Exit Screen ...

Page 646: ...able damage to the system 1 Make sure the Vivid i n Vivid q N scanner is connected to both battery and AC power 2 Insert the Vivid i n Vivid q N software installation CD in the DVD CD R and close the tray 3 Hold down the ON OFF button on the control console for 3 seconds The system powers ON and when system boot up from the CD is complete the following options are displayed in the DOS screen promp...

Page 647: ...the BIOS update process When the BIOS Update process is complete the following message is displayed 7 Press any key to restart the system The boot up screen is displayed as shown in Figure 8 124 on page 8 120 Note If the system does not restart turn the power OFF then ON In the event that the system boot up screen is not displayed repeat the procedure from step 1 Figure 8 123 BIOS Update in Progre...

Page 648: ... SERVICE MANUAL 8 120 Section 8 5 Software Loading 8 Proceed to perform the following functionality tests Software Configuration Checks on page 4 37 Calibration Tests on page 7 25 Full System Test on page 7 10 Figure 8 124 Boot up Screen after Completion of BIOS Update ...

Page 649: ...ct the DVD CD RW drive to the system s USB and connect the power supply of the drive to power outlet 2 Place the GHOST CD inside the DVD CD RW Tray and close tray 3 Turn ON power to the system and allow it to boot up from the CD A menu will appear prompting for selection of one of four options 4 Select option 1 Vivid_i Full Clean Install System Application Software A menu will appear prompting for...

Page 650: ...pace i e complete hard disk size minus 11GB for partitions C D and F IMPORTANT Do not stop the operation during this process 8 When complete continue with the software installation procedure Refer to the relevant section Note In the event that the above process fails one can perform the hard disk partitioning procedure manually as described below 9 Proceed to perform the following functionality te...

Page 651: ...see page 8 139 it will be DELETED after performing the following procedure 1 Insert the Vivid i n Vivid q N software installation CD in the DVD CD R and close the tray 2 Hold down the ON OFF button on the control console for 3 seconds The system powers ON and when system boot up from the CD is complete the following options are displayed in the DOS screen prompting you to select the required optio...

Page 652: ...the Service Tools Options menu opens 6 Do not select an option but press Ctrl C 7 When the message EMM386 is running Available memory is limited to 32MB is displayed click OK 8 When the message Terminate batch job Y N is displayed type Y 9 At the prompt A type pqmagic and press Enter The Partition Magic Screen is displayed as shown below Figure 8 125 Partition Magic Screen ...

Page 653: ...on click OK 12 Repeat step 10 and step 11 to select and delete each of the remaining partitions in turn including the partition type Extended The non partitioned hard disk will now be shown on the Partition Magic Screen with all disk space listed as Unallocated for example 19 594 9 MB 13 With Unallocated selected highlighted create the first partition by selecting the Create Tool under the Operati...

Page 654: ...ed checkbox 26 When done click OK The User partition and allocated space is now listed in the Partition Magic Screen 27 With Unallocated selected in the Partition Magic Screen repeat step 15 and step 16 to create the next partition using the following Partition Logical Partition Partition Type NTFS 28 Type the Label as SWAP 29 Type the size required This should be 1 2 to 1GB for example 1022 6 30 ...

Page 655: ...N 1 VIVID I N AND VIVID Q N SERVICE MANUAL Chapter 8 Replacement Procedures 8 127 36 Proceed to perform the following functionality tests Software Configuration Checks on page 4 37 Calibration Tests on page 7 25 Full System Test on page 7 10 ...

Page 656: ...If this is first installation format the hard disk as described in Formatting and Partitioning the Hard Disk Automatic Procedure on page 8 121 When done proceed to the next step below 6 Select option 1 Vivid_i Full Clean Install System Application Software A menu will appear prompting for selection of the required option 7 Select option A Install Software The system will prompt you to type in the ...

Page 657: ...ype in the password provided for the specific software and approve by clicking OK 10 Restart the system and commence scanning 11 Proceed to perform the following functionality tests Software Configuration Checks on page 4 37 Calibration Tests on page 7 25 Full System Test on page 7 10 Install Software Service Tools Exit A B C ...

Page 658: ...n into the DVD CD RW drive and close the tray 4 Turn ON the Vivid i n Vivid q N system and allow it to boot up from the DVD The system powers ON and when system boot up from the DVD is complete the following options are displayed in the DOS screen prompting you to select the required option 5 Press 2 to perform installation of Vivid i n Vivid q N Application Software only from DVD The system will ...

Page 659: ...t Procedures 8 131 When done the Start Application screen is displayed as shown in Figure 8 127 6 Click on Install Software and follow the on screen instructions A warning message is displayed as shown in Figure 8 128 Figure 8 127 Start Application Screen Figure 8 128 Start Software Installation Message ...

Page 660: ...Loading 7 Click OK The following message is displayed 8 Click OK to continue The following warning message is displayed WARNING DO NOT INTERRUPT THE INSTALLATION PROCESS Interruption will result in a non functional system 9 Press any key to continue Installation commences and the following message is displayed ...

Page 661: ...ET button 11 Press any key to re boot the system The system re boots into Normal Scanning mode 12 Press Config and then About The newly installed software version is now listed under the System Version tab 13 Remove the DVD from the DVD CD RW drive and close the tray 14 Proceed to perform the following functionality tests Software Configuration Checks on page 4 37 Figure 8 129 Software Installatio...

Page 662: ...nstallation Complete message is displayed refer to Figure 8 129 on page 8 133 7 Move the cursor over the displayed message window and press the SET button 8 Press any key to re boot the system The system re boots into Normal Scanning mode 9 Press Config and then About The newly installed software version is now listed under the System Version tab refer to Figure 8 130 on page 8 133 10 Remove the C...

Page 663: ...n as ADM The default password is ulsadm 4 Press the About button and activate the FwVersion tab verify that PMC version is 19 or above as shown in the example in Figure 8 131 below 5 Turn the system OFF and remove the DVD USB cable and both battery and AC adapters 6 Connect the AC adapter only NO BATTERY 7 Boot up the system in Maintenance mode using the Service Dongle 8 Go to the following direct...

Page 664: ...elow Figure 8 132 Run PMCbootloader exe A window is displayed showing the current PMC version and the path of the PMCJPS2 hex file 10 Click Burn The burning process will take approximately 30 seconds When done a brief message will appear Burning process failed see Figure 8 133 below Note The failed message is normal and should be anticipated Figure 8 133 Burning Process Failed Message ...

Page 665: ...4 Wait approximately 5 seconds then boot up the system again Allow the system to boot up into Normal Scanning mode 15 Press Config The system will prompt you to login 16 Login as ADM The default password is ulsadm 17 Press the About button and activate the FwVersion tab verify that PMC version is now 23 as shown in the example in Figure 8 134 below Figure 8 134 Checking the PMC Version 18 In the e...

Page 666: ... be repeated Other indications of failure are that the system does not shut down automatically step 11 above or there is a strange pattern on the display In the event that failure has occurred perform a full shut down if necessary remove the AC adapter if the system seems to be frozen Re boot the system in Maintenance mode using the AC adapter only Continue from step 5 above 19 Proceed to perform ...

Page 667: ...e DVD and the PCMCIA 2 Press the Eject button of the DVD CD RW drive and remove any media if present The DVD CD RW drive tray ejects 3 Insert the CD provided with the software kit and press the Eject button again The DVD CD RW drive tray will load the software CD 4 Hold down the ON OFF button for 5 seconds to shutdown the system The shutdown options are displayed on screen 5 Select Full Shutdown T...

Page 668: ...ins the software installation process and a progress indicator is displayed as shown below This stage of the process takes approximately 5 mins A message is then displayed on screen prompting you to press CTRL ALT DELETE to reboot the system 10 Press Eject and remove the CD Press CTRL ALT DELETE The system re boots 11 Proceed to perform the following functionality tests Software Configuration Chec...

Page 669: ...oll down menu 2 At the command prompt type Echoloader service and click OK The default scanning screen opens 8 6 1 2 2 Running Disk Management NOTE See Disk Management in the Vivid i n or Vivid q N User s Manual on your document CD 1 Insert media e g CD DVD or connect to a network repository 2 From the keyboard press the Config button The Operator Logon screen opens see Figure 8 135 NOTICE IMPORTA...

Page 670: ...assword NOTE The Operator Logon screen only appears after the first boot up of the system 4 Insert media e g CD DVD or connect to a network repository 5 Click the Admin tab at the bottom of the screen 6 Click the Disk Management tab at the top of the screen see Figure 8 136 Figure 8 136 Admin Screen 7 Under Operation select Copy see Figure 8 137 Figure 8 135 Operator Logon Screen ...

Page 671: ...and follow the on screen wizard instructions The system stores the images 8 6 1 3 Running the Export Procedure 1 Insert the required storage media into the system 2 From the keyboard press the Patient button The patient menu opens see Figure 8 138 3 From the screen select the Patients List The Search Create Patient screen is displayed see Figure 8 139 Figure 8 137 Copy Images for Export Figure 8 1...

Page 672: ...soft key at the bottom of the screen see Figure 8 140 NOTE If the Export button is not displayed click More consecutively until the Export button is displayed Figure 8 140 Export and More Soft Keys The following warning Figure 8 141 is displayed to indicate that only the database is exported to the target media 5 Click OK The Export screen opens see Figure 8 142 Figure 8 141 Database Export Warnin...

Page 673: ...he required destination media Figure 8 142 and click OK The Export Patient List screen opens Figure 8 143 Figure 8 143 Export Patient Archive 7 Click the soft keys at the bottom of the screen consecutively in the following order as the system processes the actions Select All Copy and Done see Figure 8 144 Figure 8 144 Soft Keys Figure 8 142 Export Screen ...

Page 674: ... 4 Click the Import soft key at the bottom of the screen see Figure 8 145 NOTE If the Import button is not displayed click More consecutively until the Import button is displayed NOTE A warning is displayed Figure 8 147 to indicate that only the database is imported to the target device Figure 8 146 Database Export Import Warning 5 Click OK The Import screen appears see Figure 8 147 6 Under From i...

Page 675: ... order as the system processes the actions see Figure 8 148 NOTE Wait while the system processes each action before pressing the next soft key 8 Press Alt E to eject the CD 9 Press 2D on the keyboard to return to the default scanner screen 8 6 2 How to Restore User defined Presets For details refer to Archiving and Loading Presets on page 4 17 Figure 8 148 Select All Copy and Done Buttons ...

Page 676: ...n in Figure 8 149 below 8 7 1 Connecting and Removing Peripherals For detailed instructions on connecting peripherals to the Vivid i n Vivid q N portable ultrasound scanner refer to the following procedure as appropriate USB Global Modem on page 3 34 Wireless USB Network Adaptor on page 3 35 DVD CD RW Drive on page 3 36 MOD Drive on page 3 43 External USB Hard Disk on page 3 44 For detailed instal...

Page 677: ...d removal of peripherals perform the following functionality tests SafeLock Cart Grounding Continuity on page 10 23 only if the Vivid i n Vivid q N system is currently mounted on the SafeLock cart SafeLock Cart Chassis Current Leakage Test on page 10 25 NOTE For details on connecting and configuring printers refer to Connecting Printers on page 3 21 ...

Page 678: ...tup 8 7 2 4 External USB Hard Disk Removal Procedure NOTE Ensure that there is no activity on the drive before you turn it off If power is turned off while files are being written to the drive the files will be corrupted 1 Switch OFF the power to the unit The power switch is located at the rear of the unit see Figure 8 151 on page 8 151 2 Disconnect the AC Power cable connected between the power s...

Page 679: ...t at the rear of the disk unit 2 Connect the cable from the power supply to the power connector on the rear of the disk unit 3 Verify that the volume configuration is set to RAID 1 The External USB Hard Disk is configured as a mirrored volume RAID 1 This ensures optimal data integrity and immediate access to the data even if one disk fails 4 Verify that both MODE switches are set to the ON positio...

Page 680: ...ternal USB Hard Disk to the Vivid i n Vivid q N scanner at a time Follow these steps to set the Volume Name 1 Connect the USB cable from the External USB Hard Disk to a laptop or PC 2 Power up the External USB Hard Disk 3 Under My Computer after a while the disk will show as a new drive named Removable Disk refer to Figure 8 153 on page 8 153 It will be allocated the next available drive letter 1 ...

Page 681: ...ent Procedures 8 153 Figure 8 153 Removable Disk Shown under My Computer 4 Right click the drive name Removable Disk and select Rename from the menu Figure 8 154 Select Rename from the Drop down Menu 5 Enter the name you have decided to use for the drive Figure 8 155 Changing the Name Removable Disk F ...

Page 682: ...the Safely Remove Hardware tool in the Task Bar before continuing with the next step 9 Disconnect the USB cable from the laptop or PC 8 7 2 7 Connecting the External USB Hard Disk to the Vivid i n or Vivid q N Scanner 1 Connect the USB cable to a USB port on the Vivid i n or Vivid q N scanner 2 Turn ON power to the Vivid i n Vivid q N ultrasound scanner 3 Under My Computer refer to Figure 8 153 on...

Page 683: ...view of replacement parts for the Vivid i n and Vivid q N Table 9 1 Contents in Chapter 9 Section Description Page Number 9 1 Overview 9 1 9 2 List of Abbreviations 9 2 9 3 Mechanical Hardware Parts 9 3 9 4 Electronic Parts 9 4 9 5 Cables 9 7 9 6 Software 9 8 9 7 Probes 9 9 9 8 Peripherals 9 11 9 9 SafeLock Cart Parts 9 12 9 10 Optional Modo Cart 9 14 9 11 Vivid i n Spare Part Kits 9 15 9 12 Produ...

Page 684: ... 2 List of Abbreviations Assy Assembly BEP Back End Processor Ctrl Control FEP Front End Processor FRU 1 Replacement part available in parts hub FRU 2 Replacement part available from the manufacturer lead time involved LCD Liquid Crystal Display Int Internal I O Input Output PWA Printed Wire Assembly Recv Receive XFRMR Transformer ...

Page 685: ...06 2410799 BT 06 1 Y 4 HINGES concealed 45 MM 52 MM 2381805 2 BT 06 and BT 09 1 Y 5 Rubber inserts kit 2410804 BT 06 and BT 09 1 Y 6 Screw Kit FL200035 BT 06 and BT 09 1 Y 7 Internal Battery Pack 2378963 2 BT 06 Replaces FL200027 and FL200068 1 Y 8 Internal Battery Pack Black S2423267 Vivid i n BT 09 and Vivid q N 1 Y 9 Vivid q cover kit S2423362 1 Y 10 VI COVER KIT BT 09 S2423361 1 Y 11 VIVID I S...

Page 686: ...yboard Assy 5376323 1 Y 2 Vi Operator Panel w o ECG Keyboard Assy BT 09 S2423357 1 Y 3 RoHS Keyboard Assy Vivid I BT 06 R2409723 1 Y 4 Trackball without bracket 2405164 2 BT 06 1 Y 5 Trackball S2423360 BT 09 1 Y 6 Backend assy S2424596 SATA BEP SERVICE KIT BT 06 and BT 09 Replaces R2380485 1 Y 7 DC DC MODULE DC Power Supply LV and HV S2403248 4 BT 09 Replaces FL200056 and FL200074 1 Y 8 DC DC MODU...

Page 687: ...SSEMBLY R2403248 4 1 Y 10 AC adaptor wo cable S2422706 BT 06 and BT 09 1 Y 11 Hard Disk Module 2376515 2 BT 06 and BT 09 Excludes software 1 Y 12 HARD DISK ASSY VI Q NEW SATA BB R2424374 Hard Disk Support 1 Y 13 Vi q SATA BEP SERVICE KIT S2424596 1 Y Table 9 3 Electronic parts cont d 2 of 3 ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU ...

Page 688: ...SPLAY MODULE ASSY Vivid q R2423283 1 Y 16 ECG Module Assy NR2421861 For Vivid q N 1 Y 17 TR32 v12 Board NS2423369 For Vivid q N 1 Y 18 Probe and MUX Assy N5420885 For Vivid q N 1 Y 19 RFI Board N2418295 2 For Vivid q N 1 Y 20 Hard Disk Duplication Jig 2414958 2 BT 06 and BT 09 1 Y Table 9 3 Electronic parts cont d 3 of 3 ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU ...

Page 689: ...ation Box Cable Kit 2409727 BT 06 and BT 09 1 Y 2 LEAD WIRE KIT EUROPE YELLOW RED GREEN S2424558 For Vivid q N 1 Y 3 Cable ECG Cable Europe 2418832 2 BT 06 and BT 09 Replaces FL200039 1 Y 4 Cable External ECG 2413451 BT 06 and BT 09 Replaces FL200054 1 Y 5 Cable Vi Power Cable Continental Europe 2 5A 250V 2412094 2 BT 06 and BT 09 1 Y 6 RFI TO BB V2 FLEX VI Q R2361860 2 RFI TO BASE BOARD FLEX V2 1...

Page 690: ...ftware Table 9 5 Software ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU 1 Vivid q N BT11 System SW 6 1 9 and Application SW 11 2 0 build 40 FQ200002 Software for Vivid q N BT 11 1 Y 2 Service Software Kit BT 09 FL200112 Software for Vivid i n BT 09 1 Y 3 Service Software Kit BT 06 FL200075 Software for Vivid i n BT 06 1 Y ...

Page 691: ...BT 09 1 Y 11 Probe 10S RS 2377040 H4000PF BT 06 and BT 09 1 Y 12 Probe M4S RS H40452LH 1 5 3 6 MHz 1 Y 13 Probe 9L RS H40442LL 4 0 13 0 MHz 1 Y 14 Probe 6T RS KN100093 H45531MZ BT 06 and BT 09 Must have TEE Probe Interface Module H45021D 1 Y 15 Probe 6Tc RS KN100104 H45551ZE 2 9 MHz to 8 0 MHzBT 09 1 Y 16 Probe 9T RS KN100073 H45531YM BT 06 and BT 09 Must have TEE Probe Interface Module H45021D 1 ...

Page 692: ... On Bite Guard Adult H45511EE BT 06 and BT 09 1 Y 30 TEE Clip On Bite Guard Adult OR H45521CB BT 06 and BT 09 1 Y 31 TEE Conventional Bite Guard Pediatric H45521JG BT 06 and BT 09 1 Y 32 TEE Conventional Bite Guard Adult H45521JH BT 06 and BT 09 1 Y 33 Adult TEE Scanhead Protection Cover H45521CK BT 06 and BT 09 1 Y 34 Pediatric TEE Scanhead Protection Cover H45541RN BT 06 and BT 09 1 Y Table 9 6 ...

Page 693: ...deo Scan Converter H45531XE 2412280 BT 06 and BT 09 1 Y 7 Modem USB 5114058 BT 06 and BT 09 1 Y 8 USB Flash Card 1GB R2422220 BT 06 and BT 09 1 Y 9 USB Flash Card 512MB H45531XG 2411544 BT 06 and BT 09 1 Y 10 NetGear PCIMCIA Network Card 2417006 BT 06 and BT 09 1 Y 11 VGA adaptor 2420728 BT 06 and BT 09 1 Y 12 DVD RW Samsung T084M RSBN S2423870 BT 06 and BT 09 Replaces 2421141 2 and 2417993 2 Supp...

Page 694: ...ble 9 8 SafeLock Cart Parts 1 of 2 ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU 1 AC Distribution Assy 220V 240V S2423305 1 Y 2 Handrest S2423250 1 Y 3 Lower Shelf Package assy S2422977 1 Y 4 Main Column S2422975 1 Y 5 Top Shelf Kit S2422974 1 Y 6 Probe Shelf Assy Kit S2422972 1 Y 7 Gas Spring S2423307 1 Y 8 Lower Shelf Rear Cover Cart Kit S2422976 1 Y ...

Page 695: ...Y 12 Vivid i Cart Gel Holder Cup S2422982 1 Y 13 Probe Cable Hooks S2422983 1 Y 14 AC Cable Hook S2422984 1 Y 15 Pencil Probe Holder Kit S2422985 1 Y 16 Pheripherals Power Out Cable S2422986 1 Y 17 Vi Cart Screw Kit S2423344 1 Y 18 USB Network Cables Kit S2423489 1 Y 19 5V Med Grade AC Adapter PSE Kit S2423131 1 Y Table 9 8 SafeLock Cart Parts cont d 2 of 2 ITEM PART NAME PART NUMBER DESCRIPTION Q...

Page 696: ...robe and gel holder peripherals shelves hand rest and handles Y 2 Cable Clip M Cart KTZ220885 BT 06 and BT 09 Cable Clip M Cart 2 Y 3 Isolation Transformer M Cart KTZ220714 BT 06 and BT 09 Isolation Transformer Power Supply IMED300WR M Cart 1 Y 4 Probe Holder Kit M Cart 5183958 BT 06 and BT 09 Probe Holder Kit M Cart 1 Y 5 Front and Rear Wheels M Cart KTZ220887 BT 06 and BT 09 Front and Rear Wheel...

Page 697: ...includes Back end Assy FL200055 TR32 Board FL200069 1 Y 1 Spare Part Kit 1 for Vivid i n BT06 FL200087 BT 06 1 Y 2 Spare Part Kit 2 FL200115 BT 09 This kit includes Back end Assy FL200055 RFI Board FL200061 1 Y 3 Spare Part Kit 2 FL200088 BT 06 1 Y 4 RFI BOARD V2 NATURAL N2418295 2 1 Y 5 PROBE AND MUX ASSY NATURAL N5393629 1 Y 6 TR32 BOARD V12 NATURAL ROHS N2354258 12 1 Y 7 FRONTEND ASSY Vivid i q...

Page 698: ... 1 Y 9 FANS ASSY VIVID I R2376696 1 Y 10 ECG BOARD V3 Vivid i q N Natural N2419020 1 Y 11 Spare Part Kit 3 FL200111 This kit includes Keyboard assy FL200059 Fans and Speakers Kit FL200040 Cable and Flex kit FL200037 ECG Module Assy w Cables FL200036 1 Y 12 Spare Part Kit 3 FL200089 BT 06 Table 9 10 Spare part kits cont d 2 of 4 ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU ...

Page 699: ...Doc CD FL200028 SW CD FL200103 SW ver 1 36 19 Appl SW 6 1 0 build 122 FL200107 Application SW Patch for Vivid i n and Vivid q N 1 Y 17 VI COVER KIT BT09 S2423361 This kit includes BOTTOM BASIC ASSY VIVID I CARTON PIZA 40X32X5 CM DISPLAY RIGHT LATCH ASSY DISPLAY LEFT LATCH ASSY REAR DISPLAY COVER VIVID I DISPLAY FRONT FRAME ASSY VIVID I EXTERNAL GE LOGO LABEL IN GRAY DISPLAY SCREW CAP MAGNET PERMAN...

Page 700: ... BOTTOM BASIC ASSY VIVID I DISPLAY RIGHT LATCH ASSY VIVID I DISPLAY LEFT LATCH ASSY VIVID I REAR DISPLAY COVER VIVID I DISPLAY FRONT FRAME ASSY DISPLAY SCREW CAP VIVID I MAGNET PERMANENT 4X20MM DISPLAY FLEX COVER VIVID I INTERNAL BATTERY DUMMY CASE ASSY HARD DISK PANEL VIVID IBT 06 1 Y Table 9 10 Spare part kits cont d 4 of 4 ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU ...

Page 701: ...nual Turkish FQ092006 1 N 6 BT 11 Vivid q N Reference manual English FQ092007 1 N 7 Vivid i n SafeLock Cart User manual EN FR TR FL092077 English French and Turkish 1 N 8 Vivid i n Cart User manual EN FR FL092096 1 N Table 9 12 Product Manuals BT 09 ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU 1 Vivid i n Service Manual FL091021 This Manual 1 N 2 Vivid i n Unpack Pack Procedure FL094023 1 N 3 BT...

Page 702: ...ct Manuals BT 06 ITEM PART NAME PART NUMBER DESCRIPTION QTY FRU 1 Vivid i n Service Manual FL091021 This Manual 1 N 2 Vivid i n Unpack Pack Procedure FL094023 1 N 3 BT06 Vivid i n User manual English FL092083 1 N 4 BT06 Vivid i n User manual French FL092084 1 N 5 BT06 Vivid i n Reference manual English FL092094 1 N 6 Vivid i n Cart User manual EN FR FL092096 1 N ...

Page 703: ...sound scanner and its peripherals These procedures are designed to protect your investment in the system and are intended to maintain the quality of system performance After delivery and installation of the Vivid i n Vivid q N ultrasound unit your service representative will contact you to schedule the first of several routine service visits In addition to routine maintenance by trained Field Serv...

Page 704: ...eaker before you remove any parts Proceed with caution whenever power is ON and covers are removed DANGER Do not pull out or insert circuit boards while mains power to the system is ON CAUTION Practice good ESD prevention Wear an anti static strap when handling electronic parts and when disconnecting connecting cables CAUTION Do NOT operate the unit unless all board covers and frame panels are sec...

Page 705: ...ation from organizations such as the American College of Radiology USA it is the customer s responsibility to have a Quality Assurance program in place for each scanner The program must be directed by a medical physicist the supervising radiologist physician or appropriate designee Routine Quality Control testing of the system must be conducted regularly The same tests are performed regularly duri...

Page 706: ...evels of safety dependability and system performance Your GE Medical Systems Service Representative has an in depth knowledge of your Vivid i n Vivid q N ultrasound scanning system and can best provide competent efficient service Please contact us for details of your nearest GE Medical Systems Service Representative and or a quotation for services by our qualified professional FSEs The service pro...

Page 707: ...akage Checks Also after corrective maintenance or as required by your facility s QA program Surface Probe Current Leakage Checks Also after corrective maintenance or as required by your facility s QA program Endocavity Probe Current Leakage Checks As prescribed in the Probe Manual Transesphongeal TEE Probe Current Leakage Checks As prescribed in Probe Manual Surgical Probe Current Leakage Checks A...

Page 708: ... cuts or abrasions Equipment for loose or missing hardware The keyboard for defects Brakes for defective operation DANGER To avoid electrical shock hazard do not remove the panels or covers from the machine This servicing must be performed by qualified service personnel only Failure to comply could result in serious injury DANGER If any defects are observed or any system malfunctions occur DO NOT ...

Page 709: ...tatic Vacuum Cleaner 46 194427P278 46 194427P279 120V 230V Safety Analyzer 46 285652G1 DALE 600 KIT or equivalent for electrical tests SVHS VCR Cassette E7010GG E7010GF 60 minutes 120 minutes SVHS VCR Head Cleaner See VCR user manual for requirements 3 5 MOD MEDIA E8381AA E8381AB Blank 128 M disk Blank 230 M disk 5 25 MOD MEDIA 2277190 Blank 2 3 M disk Blank 5 2 M disk 5 25 MOD Disk Cleaning Kit V...

Page 710: ... N SERVICE MANUAL 10 8 Section 10 4 Tools Required System Service Code Required in order to enter system diagnostics User Manual Current revision Service Manual Current revision Table 10 3 Overview of Requirements for Care Maintenance cont d Tool Part Number Comments ...

Page 711: ...e customer if they have any problems or questions about the equipment 2 Paperwork Fill in the appropriate details at the top of the Vivid i n Vivid q N System Inspection Certificate see page 10 34 Note all probes and system options 3 Power up Turn the system power ON and verify that all fans and peripherals turn on Watch the displays during power up to verify that no warning or error messages are ...

Page 712: ...sic system controls that affect this mode of operation 5 Applicable Software Optionsa a Some software may be considered as standard depending upon system model configuration Verify the basic operation of all optional modes such as Multi Image 3D Contrast Harmonics Cine Stress Echo etc Check the basic system controls that affect each the operation of each option 6 Probe Elements Perform an Element ...

Page 713: ... 3 Color Printer Verify hardcopy output of the Color video page printer Clean heads and covers if necessary 4 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 5 InSite iLinq Verify that InSite is functioning properly Ensure two way remote communications Warranty Contract Customers only 7 ECG Verify basic operation with customer 8 DVD CDRW Verify that the DVD CDRW is ...

Page 714: ...ent CAUTION When performing Cleaning Procedures to prevent the risk of system damage always observe the following precautions Use only cleaning materials and solutions such as Isopropyl Alcohol IPA as recommended in the procedures described in Table 10 8 on page 10 13 Do not use any solutions or products not listed in the Vivid i n Vivid q N User Manual Do not spray any liquid directly onto the Vi...

Page 715: ...ot contain impurities Wring out as much of the liquid as possible then wipe the LCD Display again Do not let any liquid drip into the computer Be sure to dry the LCD Display before closing the cover 3 Control Panel and Keyboard Control Panel On a weekly basis moisten a soft non abrasive folded cloth or sponge with a 70 30 mixture of isopropyl alcohol and water that does not contain impurities Do n...

Page 716: ...yl alcohol and water Never allow liquid to flow into the keyboard Use a 70 30 mixture of isopropyl alcohol and water Staff users doctors Monitor display surface Weekly In event of contamination clean immediately with a 70 30 mixture of isopropyl alcohol and water Clean with dry soft lint free cloth When contaminated with blood or other stains secretions use a 70 30 mixture of isopropyl alcohol and...

Page 717: ...age 10 13 Inspect the Probe Holders for cracks or damage 7 SafeLock Cart Check all four SafeLock Cart wheels and castors for wear Check the brakes on each of the two front wheels Make sure that pressing the brake locks the wheel preventing it from swivelling or rolling and therefore stopping movement of the SafeLock Cart Verify that releasing the brakes allows free movement of the SafeLock Cart Ch...

Page 718: ...canning 13 Peripherals Check and clean the peripherals according to the manufacturer s directions To prevent EMI or system overheating dress the peripheral cables inside the peripheral cover 14 External I O Check all connectors for damage and verify that the labeling is good 15 Power and System Status Indicators Check for proper operation of all Power and System Status Indicators 16 Battery Check ...

Page 719: ...m a visual check of the probe pins and system sockets before plugging in a probe Avoid pulling pinching or kinking the probe cable since a damaged cable may compromise the electrical safety of the probe To avoid the risk of a probe accidentally falling do not allow the probe cables to become entangled with or to be caught in the SafeLock Cart wheels Protect the probe when moving the unit Use a sof...

Page 720: ...em sockets before plugging in a probe When handling probes always observe the precautions listed in Probe Handling on page 10 17 The TEE and Interoperative probes often have special usage considerations always refer to the individual probe manufacturers handling instructions user manual 10 6 4 Probe Cleaning 10 6 4 1 Basic Probe Cleaning NOTE For details on general probe cleaning refer to the info...

Page 721: ...brasive soap 3 Follow the manufacturer s cold sterilization instructions to disinfect the probe including concentration and soaking time 10 minutes is a typical disinfecting period 4 Follow step 3 for a much longer duration to sterilize the probe 10 hours is a typical cold chemical sterilization duration 5 When complete thoroughly rinse the washed lens in clear water Apply sterile distilled water ...

Page 722: ...s must be properly removed from any part or equipment prior to shipment GEMS employees as well as customers are responsible for ensuring that parts equipment have been properly decontaminated prior to shipment Under no circumstances should a part or equipment be shipped before being visibly clean and properly disinfected The purpose of the regulation is to protect employees in the transportation i...

Page 723: ...cleaning and disinfecting procedures before handling the equipment Test the system peripherals and probes for current leakage WARNING EXCESSIVE CURRENT LEAKAGE CAN CAUSE FATAL INJURY EXCESSIVE CURRENT LEAKAGE CAN INDICATE DEGRADATION OF INSULATION OR OTHER PART AND COULD POTENTIALLY CAUSE ELECTRICAL FAILURE DO NOT USE PROBES OR EQUIPMENT THAT HAVE EXCESSIVE CURRENT LEAKAGE Important When the SafeL...

Page 724: ...untry Normal Condition Open Ground Reverse Polarity Open Neutral Mains Applieda a Mains Applied is the Sink Leakage Test In this test mains supply voltage is supplied to the part or equipment to determine the amount of current that will pass sink to Ground if a patient comes into contact with the part USA 0 05 mA 0 05 mA 0 05 mA 0 05 mA N A Other 0 1 mA 0 5 mA 0 5 mA 0 5 mA 5 0 mA Table 10 16 Type...

Page 725: ...let tester can detect the condition in which the Neutral grounded supply conductor and the Grounding protective earth conductor have been reversed If later tests indicate high current leakages this should be suspected as a possible cause and the outlet wiring should be visually inspected 10 7 4 SafeLock Cart Grounding Continuity NOTE This test is applicable to the SafeLock Cart only It is not nece...

Page 726: ...power cable should be tested with a substitute Safety Tester 1 Connect one of the Ohmmeter probes to the tested AC wall outlet cable side Ground pin 2 Using the other Ohmmeter probe touch the exposed metal part of the SafeLock Cart 3 Set the meter Function switch to the Resistance position 4 Measure and record the Ground wire resistance This should be less than 0 2 Ohms NOTE Since very low resista...

Page 727: ...N Be sure to turn the unit power OFF before switching them using the Polarity switch and or the Neutral switch Otherwise the unit may be damaged 10 7 5 2 Generic Procedure The test verifies the isolation of the power line from the chassis The testing meter is connected from accessible metal parts of the case to Ground Measurements should be made with the unit ON and OFF with the power line polarit...

Page 728: ...own in Table 10 14 Record all data on the Vivid i n Vivid q N System Inspection Certificate see page 10 34 Table 10 17 Typical Data Sheet for Chassis Source Current Leakage Test Unit Power Tester Polarity Switch Tester Neutral or Ground Switch Test 1 Probe Connector Test 2 Caster Wheel Test 3 CRT Test 4 Optional Test 5 Optional Enter Name of tested peripheral here ON NORM OPEN ON NORM CLOSED ON RE...

Page 729: ...owered ON Be sure to turn the unit power OFF before switching them using the Polarity switch and or the Neutral switch 10 7 6 2 Generic Procedure Measurements should be made with the Ground Open and Closed with power line polarity Normal and Reversed and with the Vivid i n Vivid q N unit ON and OFF For each combination the operating controls such as the lead switch should be operated to find the w...

Page 730: ...Isolated Patient Lead Sink Leakage Isolation Test NOTE Refer to the procedure in the IEC 60601 1 CAUTION Line voltage is applied to the ECG leads during this test To avoid the risk of possible electric shock the system being tested must not be touched by anyone patient or operator while the ISO TEST switch is depressed NOTE It is not necessary to test each lead individually or power condition comb...

Page 731: ... Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Current Leakage Test and Patient Lead to Lead Current Leakage Test 115V 10uA 10uA 220 240V 500uA 10uA Table 10 19 Maximum Allowance Limit for ECG Lead Isolation Sink Test AC Power Source Maximum Allowance Limit Patient Lead Isolation Current Test 115V 20uA 220 240V 5mA Table 10 20 Typical Data Sheet for ECG Lead to Le...

Page 732: ...hould be made with the Ground OPEN and CLOSED with power line polarity Normal and Reversed and with the unit OFF and ON For each combination the probe must be active to find the worst case condition NOTE Each probe will have a certain amount of current leakage depending on its design Small variations in probe current leakages are normal from probe to probe Other variations will result from differe...

Page 733: ...XTERNAL connector 4 Set the meter s FUNCTION switch to EXTERNAL position 5 Connect the probe for test with the connector of the console 6 Add the saline probe and the imaging area of the probe into the saline bath 7 Turn unit power ON for the first part turn it OFF for the second half 8 Depress the ISO TEST rocker switch and record the highest current reading 9 Follow the test conditions described...

Page 734: ...age 10 34 CAUTION When power to the unit is ON never switch the Polarity and the status of Neutral Be sure to turn OFF power to the unit before switching them using the POLARITY switch and or the NEUTRAL switch Failure to comply with this warning may cause damage to the unit Table 10 21 Typical Data Sheet for Transducer Source Current Leakage Test Transducer Tested Unit Power Tester Power POLARITY...

Page 735: ...with the probe or the scanner NOTE Each probe will have some amount of leakage depending on its design Small variations in probe current leakages are normal from probe to probe Other variations will result from differences in line voltage and test lead placement The maximum allowable current leakage for body surface contact probes differs from that of inter cavity probes Be sure to enter the corre...

Page 736: ...Frequency Scan Format Model Number Serial Number Probe 6 Frequency Scan Format Model Number Serial Number Probe 7 Frequency Scan Format Model Number Serial Number Probe 8 Frequency Scan Format Model Number Serial Number Probe 9 Frequency Scan Format Model Number Serial Number Scan Format Phased Array Linear Array Curved Array Mechanical Array or other Functional Checks if applicable OK or N A Phys...

Page 737: ... Caster Chassis Source Current Leakage CRT Patient Lead Source Leakage Lead to Ground Patient Lead Source Leakage Lead to Lead Patient Lead Source Leakage Isolation Peripheral 1 Current Leakage Peripheral 1 Ground Continuity Peripheral 2 Current Leakage Peripheral 2 Ground Continuity Peripheral 3 Current Leakage Peripheral 3 Ground Continuity PROBES Probe Number from page 10 34 Max Value Allowed M...

Page 738: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL 10 36 Section 10 9 Vivid i n Vivid q N Inspection Certificates This page was intentionally left blank ...

Page 739: ......

Page 740: ... 2011 General Electric Company GE Medical Systems a General Electric Company going to market as GE Healthcare GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...

Reviews: