Understanding Control Test Results
Recalling Test Results
Control Solution Range
4.9 ~ 6.7 mmol/L
13.8 ~ 18.8 mmol/L
L2 = Level 2 (Normal) L4 = Level 4 (High)
Your control solution test results should fall within the control solution range. If the results are within the range, the
GE200 Blood Glucose Monitoring System is working correctly.
Possible reasons your Control Solution results are out of the range :
- Your GE200 Control solution is expired or was first opened more than 3 months ago.
- Your GE200 Blood Glucose Test Strip has expired.
- You left the cap of the Blood Glucose Test Strip vial or the control solution off for a period of time.
- You did not perform the test procedure correctly.
- The GE200 Blood Glucose Meter or GE200 Blood Glucose Test Strip have malfunctioned.
If GE200 Control Solution results are out of range, your GE200 Blood Glucose Monitoring System
may not be working properly. Repeat the Quality Control Test. If your control solution results are still out of range, do
not use GE200 Blood Glucose Meter to test your blood glucose. Please contact the
toll free at
.
Auto Control Medical Distribution
and Customer Service Center
1-800 964-8434
The GE200 Blood Glucose Meter is able to automatically store a maximum of 1000 test results with time and date. If
your meter has stored 1000 results, the newest test result will replace the oldest one.
To recall your test memory, start the meter without a test strip inserted.
1. Press the Main button to switch from the Time Mode to Memory Mode. You will see the “mem”
icon in the upper left corner of the display. The latest test result will appear with the number “1”.
2. Use either the Left or Right button on the side to review all previous results with date and time.
Results display from the most recent (Sequence no. “1”) to the oldest (Sequence no. “1000”) in
the lower right hand corner of the screen.
3. The Quality Control Test result can be recalled from the saved data. When " " appears next to
the data, the test was performed using the GE200 Control Solution. The control solution test
result will not be used when recalling test average results.