7-42
Section 7-3 - Performing Front End (FE) Diagnostics
The
Report
dialog box includes the following:
•
Test Result Message:
Displays the test result.
•
Pencil Probe Relays:
Displays a graphic representation of the eight front board pencil probe
relays, as follows:
•
Each relay is represented by a button (highlighted in green) that displays the relay
number.
•
Each button is highlighted in yellow when selected.
•
Each button is highlighted in red if errors were detected for one or more relays.
•
Checkboxes:
Enable you to manipulate bad channels:
•
Show Only Bad Channels:
Select this checkbox to display only the bad channels in the
Display
area.
•
Mark bad Channel:
Select this checkbox to manually mark the currently selected channel
as bad.
5) Check the test result message and proceed as follows:
•
If one or more relays were identified as bad, trackball to
Save Report
and press
Select
, and
contact a GE Medical Systems service representative.
•
If no relays were identified as bad, mark the channels suspected as being bad as follows:
•
Trackball to the channel in the Display area and press
Select
.
•
Trackball to the
Mark bad channel
checkbox and press
Select
. Repeat as required to
mark additional channels.
•
Trackball to
Save Report
and press
Select
, and contact a GE Medical Systems service
representative.
6) Trackball to the
Start Close
test
and press
Select
. The pencil probe test (close) is performed and
the test results are displayed in the
Report
dialog box, as shown in