Control Console Upper Cover (Rear) Installation Procedure
1) Raise the control console to its
maximum
height.
2) Return the control console upper cover (rear section) to the Vivid™ 4 control console, making sure
it is placed in the correct position (with the screw holes properly aligned and the aperture for the
monitor correctly placed).
3) Return and fasten the 4 screws (2 on each side as shown in
) to secure the upper rear
cover in place.
4.) Return the control console lower cover, as described in the
5) Remove the control console lower rear cover, as described in the
6) Working from the
left
side of the Vivid™ 4 unit, return the rectangular cover to the roof of the side
compartment and fasten in position with the 4 screws (previously removed) - refer to
.
7) Return the 17" Monitor to the control console, as described in the
Vivid™ 4 17" Monitor Installation
CAUTION: The Vivid™ 4 17" monitor is heavy. At least two people must be available to lift the
monitor. Attempts to move this unit by one person alone, could result in personal injury, and/or
damage to the system!