Chapter 3 - Installation
3-5
3-3-1-2
Unpacking and Removing the Unit from the Wooden Crate
Before unpacking the unit, inspect the wooden crate for damage. Inspect the
Drop and Tilt
indicators
(on the
Shock-watch and Tilt-watch labels, respectively) for evidence of accidental shock or tilting of the
crate during transit - see
NOTE:
If the crate is damaged, or if either the Drop or Tilt indicators have turned red (showing failure),
please inform the GE Medical Systems
sales representative immediately. In addition, mark on
the shipping consignment note or packing slip/post-delivery checklist (in the “Package” column)
that the Tilt and/or Drop indicators show failure.
It is recommended to keep and store the crate and all other packing materials (including the Clip-lok
TM
clips, support foams, anti-static plastic cover, etc.), in the event that transport or shipment of the unit to
a different location will be required in the future.
Figure 3-2 Drop and Tilt Indicators
Tilt
indicators turn
Red
Drop
indicators turn
Red